Pure Tone Audiometry Word Scramble
|
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.
Normal Size Small Size show me how
Normal Size Small Size show me how
Question | Answer |
What level should ambient noise not exceed? | 35dB |
What should communication strategy take into account? | - Age - Hearing level - Language Skills - Attention/Concentration - Other communication difficulties |
Audiometry is preceded by... | Otoscopy |
Insert earphones are used where... | ...there is a possibility of ear canal collapse with supra-aural headphones |
History taking (audiological) | Noise exposure in the last 24 hours - Tinnitus - Recent ear infections / Pain in ears - Surgery to ears at any stage - Which ear is more acute |
What two important points should be covered at the start of the consultation? | - Verifying patient - Introducing the appointment |
What is considered loud noise exposure? | Noise in which you need to shout at a distance of 1 meter |
How long should the test take? | The test should not exceed 20 minutes |
Dealing with tinnitus | -Subjects asked to ignore tinnitus as much as possible - Inform the tester if experience difficulty (note made) |
Instructing the patient | - Testing the quietest sounds you can hear -As soon as hear sound, press button - Release when no longer heard. |
Press button irrespective of... | - Ear it is hear in - How quiet it is |
Other important instruction points | - Sit still and quietly - Interrupt is uncomfortable or if the need a break |
Where should insert earphones not be used? | - Excessive wax - Infection - Obstructions - Abnormalities |
PTA should test the ......... ear first | Most acute (better) |
Test order | 1000, 2000, 4000, 8000, 500, 250,Retest 1000 if first ear. |
If the retest value is less than 5dB different then... | The most sensitive value is taken |
If the retest value is more than 10dB different then... | Reason should be investigated (reinstruction and repeat) |
Duration of test tone | Varied between 1 - 3 seconds |
Interval between test tones | Varied between 1 - 3 seconds |
Initial familiarisation level | 1000Hz at 40dB HL or 30dB above previous audiogram. If no response increase in 20dB steps until response (5dB steps above 80dB). |
Example of audiometry: Satisfactory familiarisation | Decrease in 10 dB until response |
No response | Increase in 5dB until response |
Response | Decrease by 10dB |
No response | Increase in 5db until response at consistent level >50% of presentations |
Which is better: 3 accurate frequencies or 6 non accurate frequencies? | 3 accurate frequencies |
I am ommitting frequencies or modifying the test | I must note on the audiogram/journal |
The bone conductor is placed on the... | WORST hearing ear (noted on audiogram or journal) |
Points on position of bone conductor | As close to the pinna without touching it and with no hair obstruction |
Test order | 1000, 2000, 4000, 500 |
Instructions | Same as A/C but with emphasis on irrespective of side |
At ......Hz the signal presented by the bone conductor may conduct by air conduction, what should be done? | 4000Hz and a plug should be placed in NTE |
What is the other major source of error in bone conduction? | Vibrotactile threshold |
Normal | -10 to 20dB HL |
Mild | 20 to 40dB HL |
Moderate | 40 to 70dB HL |
Severe | 70 to 90dB HL |
Profound | Above 90dB HL |
Created by: