Save
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password

Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.

Question

characteristics of radiation risks
click to flip
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't know

Question

Biologic Effects of Ionizing Radiation (BEIR) Committe VII risk model for exposure to low-level radiation
Remaining cards (61)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

Pt protection

Patient protection

QuestionAnswer
characteristics of radiation risks invisible, long-term, cumulative
Biologic Effects of Ionizing Radiation (BEIR) Committe VII risk model for exposure to low-level radiation approx. 1 in 100 people are likely to develop solid cancer or leukemia from an exposure of 100 mSv above background dose
Important for patient safety Good communication and quality assurance programs
Beam restriction single most important factor in keeping patient dose to a minumum, only tissues of diagnostic interest should be iradiated -reduces pt dose, reduces production of scattered radiation, improves quality of image
Beam restrictor types aperture diaphragm, cone/cylinder, collimator
Collimator most efficient, attached to tube head, first set of shutters placed close to x-ray tube port window to control amt of image degrading "off-focus" radiation leaving tube, lead shutters (two adjustable)for length and width
Aperture diaphragm elementary -dedicated chest, dental unit, trauma imaging equipment -flat piece of lead having a central opening with a size and shape that determines the size and shape of x-ray beam
Cone/cylinder -used for relatively small field size (paranasal sinuses) circular, lead-lined, slide into place in the tube head or onto the collimator housing Straight-prox and dist diameters identical Flare-dist diameter greater than prox diameter -limited use, more efficient bcs beam restriction is closer to anatomy
Disadvantages of aperture and cone beam restrictors have a fixed opening size, only one field size at a given distance
mAs controls quantity, no affect on quality
kVp controls quality, affects quantity
NCRP guideline for manual collimation must be within _____ of the SID. 2%
Positive beam limitation-PBL sensors located in Bucky tray or other IR holder signal the collimator to open or close accordingly to the IR size being used, will provide a small unexposed border on all sides of the finished image -required by the NCRP to be withi
PBL requirements by NCRP -required by the NCRP to be within 3% of SID for a single side and within 4% of SID for all 4 sides
NCRP guidelines for manual collimation within 2% of the SID
Collimator assembly-light localization apparatus small light bulb to illuminate the field and a 45-degree angle mirror to deflect the light.
T/F For the x-ray light field and x-ray field to correspond accurately, the x-ray tube focal spot and the light bulb must be exactly the same distance from teh center of the mirror. TRUE
exposure factors for patient safety high kVp, low mAs -also good bcs heat delivered to x-ray tube is lower and tube life is extended
Three-phase or high frequency generators stationary radiologic equipment, constant potential waveform, reduced patient dose
Inherent filtration built in, permanent -window of glass envelope (.5 mm Al) -thin layer of oil coolant/insulation surrounding the x-ray tube -tends to increase as tube ages, tungsten deposits
Filtration summary less than 50 kV = .5 mm Al 50-70 kV = 1.5 mm Al more tha 70 kV = 2.5 mm Al
Radiosensitive organs gones, lens, blood-froming organs
Whose gonads to shield all women under 55, all men under 65
Gonad shielding easier and more effective on males than on females
When to shield the gonads -when diagnostic objectives permit -when pt is of reasonable reproductive age -when gonads lie in or within 5 cm of the collimated field
Contour shaped shields effective for gonads, shaped to enclose male reproductive organs, held in place by disposable briefs, effective for oblique, erect, and fluro exams
Shadow shields leaded material casts a shadow within the illuminated field that corresponds to shielded area, inially expensive but one time expense, can be used without contacting sterile fields, NOT TO BE USED IN FLUORO.
Mammography equipment with a molybdenum target has how much filtration? .025-.03 mm molybdenum filtration
Glass window in mammo is made of what? beryllium-low atomic number (4)
Added filtration thin sheets of aluminum that are added, collimator and its mirror
Why use PA instead of AP chest or spine? decrease dose to breast tissue to .1% of that received in AP, magnification considerations are minimal
T/F Entrance skin dose is significantly greater than exit dose. True -place radiosensitive organs away from the entrance beam
Perform these exams in PA projection when possible. Skull, chest, abdomen, scoliosis series, spine
good communication helps gain patient's trust and confidence with a thorough explanation
lead aprons during fluoro placed under patient
Minimum response/reaction time length of the shortest exposure possible with a particular AEC
If the minimum response time is too high, how is this fixed. needs fixed or will have excess density -preferable to decrease ma, but may decrease kV
Rare earth phosphors film/screen radiography -four times faster than calcium tungstate phosphors
Advantage of digital fluoro lower patient dose due to x-ray beams are pulsed not continuous
grids function to remove a large percentage of scattered radiation mostly from Compton scatter, improve radiographic contrast
air-gap technique limited to imaging fairly thin parts (thick and dense tissues would require excessive and impractical radiation exposures)
equipment must be properly calibrated to give predictable results. What two values need to be checked? linearity and reproducibility
linearity adjacent mA stations should be consistent 200 mA should be 2x exposure rate as 100 mA, any variation in output intensity must not exceed 10%
Reproducibility consistency in exposure output during repeated exposures at a particular setting, any variation in output intensity must not exceed 5%
High ratio grids with low kV BAD, discouraged because of unnecessary exposure required
high ratio grids affect on positioning decreases positioning latitude
low ratio grids used in mobile radiography, more positioning latitude
8:1 grid good for what kV range? up to 90 kV
16:1 grid good for what kV range? 100 kV or more
10:1 or 12:1 grid used in what type of equipment? general fixed equipment
T/F Moving grids call for less exposure than stationary grids. False. They require more exposure.
Ways to decrease patient fluoro dose -decrease length of exposure -use last image hold -keep pt as close to II as possible -use automatic brightness control setting with high kV, low mA -minimize "boost" and "mag" modes, low pulse rate -collimate to smallest field of view
Ways to decrease patient fluoro dose cont. -use lowest practical pulse rate -change tube angle or patient position to spread dose over larger area
NCRP guidelines for x-ray tube housing leakage less than 100 mR/hr 1 meter from tube
Source-Skin distance for all radiographic procedures (NCRP) no less than 12 in -12 inches in mobile radiography -15 inches in stationary radiography
(NCRP) Fluoro tabletop intensity fewer than 10 R/min
Where should the focal spot location be indicated? (NCRP) outside of tube housing
How often should intensifying screens be cleaned and checked? (NCRP) every 6 months
(NCRP)What type of exposure switches are used? dead-man switch
(NCRP) How visible should the patient be to the radiographer? RT should be able to see and communicate with the patient at all times
NCRP national council on radiation protection and measurements
Principal function of x-ray tube filtration reduce patient skin dose
Created by: mmarnold2
 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards