Pt protection Word Scramble
|
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.
Normal Size Small Size show me how
Normal Size Small Size show me how
| Question | Answer |
| characteristics of radiation risks | invisible, long-term, cumulative |
| Biologic Effects of Ionizing Radiation (BEIR) Committe VII risk model for exposure to low-level radiation | approx. 1 in 100 people are likely to develop solid cancer or leukemia from an exposure of 100 mSv above background dose |
| Important for patient safety | Good communication and quality assurance programs |
| Beam restriction | single most important factor in keeping patient dose to a minumum, only tissues of diagnostic interest should be iradiated -reduces pt dose, reduces production of scattered radiation, improves quality of image |
| Beam restrictor types | aperture diaphragm, cone/cylinder, collimator |
| Collimator | most efficient, attached to tube head, first set of shutters placed close to x-ray tube port window to control amt of image degrading "off-focus" radiation leaving tube, lead shutters (two adjustable)for length and width |
| Aperture diaphragm | elementary -dedicated chest, dental unit, trauma imaging equipment -flat piece of lead having a central opening with a size and shape that determines the size and shape of x-ray beam |
| Cone/cylinder -used for relatively small field size (paranasal sinuses) | circular, lead-lined, slide into place in the tube head or onto the collimator housing Straight-prox and dist diameters identical Flare-dist diameter greater than prox diameter -limited use, more efficient bcs beam restriction is closer to anatomy |
| Disadvantages of aperture and cone beam restrictors | have a fixed opening size, only one field size at a given distance |
| mAs | controls quantity, no affect on quality |
| kVp | controls quality, affects quantity |
| NCRP guideline for manual collimation must be within _____ of the SID. | 2% |
| Positive beam limitation-PBL | sensors located in Bucky tray or other IR holder signal the collimator to open or close accordingly to the IR size being used, will provide a small unexposed border on all sides of the finished image -required by the NCRP to be withi |
| PBL requirements by NCRP | -required by the NCRP to be within 3% of SID for a single side and within 4% of SID for all 4 sides |
| NCRP guidelines for manual collimation | within 2% of the SID |
| Collimator assembly-light localization apparatus | small light bulb to illuminate the field and a 45-degree angle mirror to deflect the light. |
| T/F For the x-ray light field and x-ray field to correspond accurately, the x-ray tube focal spot and the light bulb must be exactly the same distance from teh center of the mirror. | TRUE |
| exposure factors for patient safety | high kVp, low mAs -also good bcs heat delivered to x-ray tube is lower and tube life is extended |
| Three-phase or high frequency generators | stationary radiologic equipment, constant potential waveform, reduced patient dose |
| Inherent filtration | built in, permanent -window of glass envelope (.5 mm Al) -thin layer of oil coolant/insulation surrounding the x-ray tube -tends to increase as tube ages, tungsten deposits |
| Filtration summary | less than 50 kV = .5 mm Al 50-70 kV = 1.5 mm Al more tha 70 kV = 2.5 mm Al |
| Radiosensitive organs | gones, lens, blood-froming organs |
| Whose gonads to shield | all women under 55, all men under 65 |
| Gonad shielding | easier and more effective on males than on females |
| When to shield the gonads | -when diagnostic objectives permit -when pt is of reasonable reproductive age -when gonads lie in or within 5 cm of the collimated field |
| Contour shaped shields | effective for gonads, shaped to enclose male reproductive organs, held in place by disposable briefs, effective for oblique, erect, and fluro exams |
| Shadow shields | leaded material casts a shadow within the illuminated field that corresponds to shielded area, inially expensive but one time expense, can be used without contacting sterile fields, NOT TO BE USED IN FLUORO. |
| Mammography equipment with a molybdenum target has how much filtration? | .025-.03 mm molybdenum filtration |
| Glass window in mammo is made of what? | beryllium-low atomic number (4) |
| Added filtration | thin sheets of aluminum that are added, collimator and its mirror |
| Why use PA instead of AP chest or spine? | decrease dose to breast tissue to .1% of that received in AP, magnification considerations are minimal |
| T/F Entrance skin dose is significantly greater than exit dose. | True -place radiosensitive organs away from the entrance beam |
| Perform these exams in PA projection when possible. | Skull, chest, abdomen, scoliosis series, spine |
| good communication | helps gain patient's trust and confidence with a thorough explanation |
| lead aprons during fluoro | placed under patient |
| Minimum response/reaction time | length of the shortest exposure possible with a particular AEC |
| If the minimum response time is too high, how is this fixed. | needs fixed or will have excess density -preferable to decrease ma, but may decrease kV |
| Rare earth phosphors | film/screen radiography -four times faster than calcium tungstate phosphors |
| Advantage of digital fluoro | lower patient dose due to x-ray beams are pulsed not continuous |
| grids | function to remove a large percentage of scattered radiation mostly from Compton scatter, improve radiographic contrast |
| air-gap technique | limited to imaging fairly thin parts (thick and dense tissues would require excessive and impractical radiation exposures) |
| equipment must be properly calibrated to give predictable results. What two values need to be checked? | linearity and reproducibility |
| linearity | adjacent mA stations should be consistent 200 mA should be 2x exposure rate as 100 mA, any variation in output intensity must not exceed 10% |
| Reproducibility | consistency in exposure output during repeated exposures at a particular setting, any variation in output intensity must not exceed 5% |
| High ratio grids with low kV | BAD, discouraged because of unnecessary exposure required |
| high ratio grids affect on positioning | decreases positioning latitude |
| low ratio grids | used in mobile radiography, more positioning latitude |
| 8:1 grid good for what kV range? | up to 90 kV |
| 16:1 grid good for what kV range? | 100 kV or more |
| 10:1 or 12:1 grid used in what type of equipment? | general fixed equipment |
| T/F Moving grids call for less exposure than stationary grids. | False. They require more exposure. |
| Ways to decrease patient fluoro dose | -decrease length of exposure -use last image hold -keep pt as close to II as possible -use automatic brightness control setting with high kV, low mA -minimize "boost" and "mag" modes, low pulse rate -collimate to smallest field of view |
| Ways to decrease patient fluoro dose cont. | -use lowest practical pulse rate -change tube angle or patient position to spread dose over larger area |
| NCRP guidelines for x-ray tube housing leakage | less than 100 mR/hr 1 meter from tube |
| Source-Skin distance for all radiographic procedures (NCRP) | no less than 12 in -12 inches in mobile radiography -15 inches in stationary radiography |
| (NCRP) Fluoro tabletop intensity | fewer than 10 R/min |
| Where should the focal spot location be indicated? (NCRP) | outside of tube housing |
| How often should intensifying screens be cleaned and checked? (NCRP) | every 6 months |
| (NCRP)What type of exposure switches are used? | dead-man switch |
| (NCRP) How visible should the patient be to the radiographer? | RT should be able to see and communicate with the patient at all times |
| NCRP | national council on radiation protection and measurements |
| Principal function of x-ray tube filtration | reduce patient skin dose |