Intro to Rad Chp 8 Word Scramble
|
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.
Normal Size Small Size show me how
Normal Size Small Size show me how
| Question | Answer |
| What is the primary role for a radiographer? | The manipulation of expensive, high-technology x-ray equipment |
| Generic components of radiographic systems include: | X-ray tube, collimator, x-ray table, control console, Tube stand |
| What part of the radiographic system produces the x-rays and radiation? | X-ray tube |
| What is the x-ray tube made of? | Glass or aluminum and encased inside a sturdy metal housing. |
| the two primary components of the x-ray tube | Anode & Cathode |
| Positive electrode or terminal; where the x-rays are emitted | Anode |
| negative-electrode or terminal; where the electrons are liberated | cathode |
| How are x-rays created? | The tube will convert electrical energy (electricity) to x-rays |
| - controls the quantity or number of x-rays | mAs |
| -controls the strength or penetrating power of the x-rays | kVp |
| x-ray beam-limiting device attached directly below the x-ray tube | collimator |
| -controls the size and shape of the x-ray field coming out of the tube | collimator |
| diaphragm or system of diaphragms made of an absorbing material | collimator |
| designed to define the dimensions and direction of a beam of radiation device consisting of four rectilinear radiopaque blades that are adjustable to control xray field size and shape | collimator |
| -act of adjusting the collimator to increase or decrease the field size is referred to as | collimating or coning |
| increase collimation | making a smaller field size |
| The __________ decreases or limits the field size and reduces unwanted scatter radiation from reaching the IR or cassette. | Collimator |
| - automatic collimation systems | Positive Beam Limitation ( PBL) |
| allows the x-ray unit to detect the size of the IR and automatically limit the x-ray field size to that size to minimize patient dose | PBL - Positive beam limitation |
| 6 types of x-ray tables: | tilting, nontitling, free-floating, stationary top, adjustable, nonadjustable |
| - move in two directions -may be buttons for table movement or foot switches | Free-Floating or moveable table tops |
| 2 types of titling tables are: | 90-90 (tilt from horizontal to vertical in either direction.) 90-30 (tilt vertical one way and 30 deg another) |
| -located below the table top and above the cassette tray | bucky mechanism |
| -works as an oscillating grid device to decrease the amt of scatter radiation reaching the IR by moving or vibrating during the exposure | bucky mechanism |
| using the bucky system will do what to contrast? | increase it |
| another name for bucky tray | cassette tray |
| -holds IR in place under the x-ray table & keeps it center under the table | bucky tray |
| the device that gives the operator command of the x-ray machine | control console |
| 5 generic controls on the control console: | main power, kVp, mA, time, rotor-exposure switch |
| what is technique selection? | manipulation of mA, kVp, and time |
| What two components can be formed to create mAs? | mA and time |
| Main power switch is usually located? | on the wall above the control console |
| Machine power is usually located? | on the control console |
| 1kVp = _____ V | 1000 |
| kVp will range from ___ to ___ on most machines | 30, 150 |
| major kVp will change in steps of? | 10 |
| minor kVp will change in steps of? | 2 |
| 1000 mA = ____ A | 1 |
| control consoles will range from ___ mA to ___ mA | 10, 1200 |
| mA is usually in increments of... | 100 ( 100, 200, 300, etc) |
| most routine diagnostic radiography is done: | btwn 100 and 400 mA |
| the device that begins the exposure | Rotor-exposure switch |
| -contains two differnt exposure switches that are mechanically interlocked so one must be activated before the other | roto-exposure switch |
| -causes the anode rotor to rotate and prepares the anode for the exposure factors/ takes 1-2 sec | rotor switch |
| begins the exposure | exposure switch |
| -automatically terminates the exposure -termination of the exposure must be indicated both audibly and visually | timer |
| How do you avoid ending the exposure too soon? | Get in the habit of not releasing the exposure switch until the audible tone or light message indicates |
| 4 types of tube movements: | longitudinal, transverse, vertical, tube angulation |
| -travel is moving the x-ray tube lengthwise to the technologist's left or right or toward the patients head or feet | longitudinal |
| -travel is moving the tube at right angels across the table/towards or away from the technologist | transverse |
| -travel is moving the tube up toward the ceiling or down toward the floor | vertical |
| -pivoting the tube at a point where it is attached to its support | tube angulation |
| two types of tube locks: | electrical or Mechanical ( sometimes both) |
| The student must make sure the central ray (tube) is _____ to the center of the bucky tray when the tray is being used. | centered |
| -the presentation of continuous or dynamic radiographic image | fluoroscopy |
| What is added during an fluoroscopy exam? | additl fluoroscopy x-ray tube, placed under the table; an image intensifier |
| -allows the radiologist to obtain static radiographs during dynamic fluoroscopic exams | spot film device |