| Question | Answer |
| What type of x-ray tube is designed to turn on and off quickly, providing multiple short, precise exposures? | Grid-controlled Lange Q&A pg 273 |
| What principle do these circuit devices operate on: autotransformer and choke coil? | Self-induction Lange Q&A pg 273 |
| The image is magnified when using the small or large field in a dual field image intensifier? | Small Lange Q&A pg 274 |
| The minimum response time of an automatic exposure control (AEC)uses its ______ exposure time | shortest Lange Q&A pg 274 |
| The electron cloud within the x-ray tube is the product of a process called: | thermionic emission Lange Q&A pg 274 |
| What does repeated, frequent overloading do to the anode's focal track? | Causes pitting Lange Q&A pg 274 |
| A parallel-plate inoization chamber receives a particular charge as x-ray photons travel through it. This is the operating principle of what device? | AEC Lange Q&A pg 275 |
| Do kV and HVL have a direct or indirect relationship? | direct Lange Q&A pg 276 |
| What converts electrical energy into mechanical energy? | Motor Lange Q&A pg 276 |
| To terminate the x-ray exposure once the IR is correctly exposed is the job of the _______. | Phototimer Lange Q&A pg 276 |
| An electron approaching a positive nuclear charge changes direction and loses energy in which production of radiation? | Bremsstrahlung Lange Q&A pg 277 |
| What type of device is used to measure focal spot size and spatial resolution? | slit camera Lange Q&A pg 276 |
| What is the voltage ripple for a three phase-six pulse rectified generator? | 13% Lange Q&A pg 277 |
| Which device receives the remnant beam, converts it into light, and then increases the brightness? | Image Intensifier Lange Q&A pg 277 |
| Give three components of the image intensifier. | Photocathode, focusing lens, accelerating anode Lange Q&A pg 278 |
| In digital fluoroscopy, what has taken place of the image intensifier's televesion camera tube? | The charge-coupled device Lange Q&A pg 278 |
| ________ involves reciprocal motion of the x-ray tube and film. | Tomography Lange Q&A pg 278 |
| In fluoroscopy, the ABC is used to adjust the kVp and _____. | mA Lange Q&A pg 278 |
| When kV is selected on the control panel, ________ is adjusted. | the autotransformer Lange Q&A pg 278 |
| What is the device that directs the light emitted from the image intensifier to various viewing and imaging apparatus? | the beam splitter Lange Q&A pg 279 |
| The x-ray tube in CT is most likely to be associated with what? | a pulsed x-ray beam Lange Q&A pg 280 |
| A device used to ensure reproducibility radiographs, regardless of tissue density variations, is the _____. | Phototimer Lange Q&A pg 281 |
| The device used to change alternating current into unidirectional current is the _____. | Solid state diode Lange Q&A pg 281 |
| Which part if an induction motor is located outside the x-ray tube glass envelope? | Stator Lange Q&A pg 283 |
| Which part if an induction motor is located within the x-ray tube glass envelope? | Rotor Lange Q&A pg 283 |
| What mode of a trifield image intensifier will result in the highest patient does; 25 in., 17 in., or 12 in.? | The 12 in. mode Lange Q&A pg 284 |
| Fractional focus tubes, with a 0.3-mm focal spot or smaller, have special application is which of the following? magnification radiography, fluoroscopy, tomography, or image intensification | Magnification radiology Lange Q&A pg 284 |
| An increase in the heat of the filament will cause the mA to what? | increase Lange Q&A pg 284 |
| The filtering effect of the x-ray tube's glass envelope and its oil coolant are referred to collectively as what? | inherent filtration Lange Q&A pg 288 |
| What device is best used to determine the accuracy of the x-ray timer? | the spinning top Lange Q&A pg 288 |