| If more than one tissue is contained in a voxel, the CT number is determined by a _____________ of the densities of the contents of the voxel. | weighted average |
| What artifact can occur when more than one tissue type is contained in a voxel? | partial volume averaging |
| What artifact results from the mathematical averaging of various attenuation coefficients from several millimeters of different types of tissue? | partial volume averaging |
| How can we minimize the partial voluming effect? | using thinner slices |
| The edge gradient effect is one of the many causes of ____ artifacts in CT images. | streak |
| Out-of-field artifacts may look like ____ or ____ on CT images. | streaks or shading |
| The severity of the out-of-field artifact depends on what? | amount and density of the anatomy outside the SFOV |
| Which artifact results from a faulty or miscalibrated detector element? | ring artifact |
| If a detector has an offset difference of as little as ___% with neighboring detectors, a ring artifact may result on the images. | 0.1% |
| When might tube arcing occur within the tube? | when the tube vacuum is compromised |
| When might tube arcing occur outside of the tube? | when impurities cause the oil to become electrically conductive |
| Each time the tube arcs, the ____ data for that view is distorted. | projection |
| The spreading of the x-ray photons along the z-direction is responsible for an artifact called what? | cone beam artifact |