MRI REGISTRY Word Scramble
|
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.
Normal Size Small Size show me how
Normal Size Small Size show me how
| Question | Answer |
| Nerve IV | Glossopharyngeal |
| Nerve X | Vagus |
| Nerve XI | spinal accessory |
| Nerve XII | hypoglossal |
| Nerve VIII | vestibulocochlear |
| Nerve VII | facial |
| Nerve VI | Abducens |
| Nerve V | Trigeminal |
| Nerve IV | Trochlear |
| Nerve III | oculomotor |
| Nerve II | Optic |
| Nerve I | Olfactory |
| QC test for low contrast spatial resolution | Circular image. Should be able to visualize 10 spokes |
| QC test for geometric accuracy ( Z) | Rectangular image showing triangles and rectangles |
| QC test for geometric accuracy (XY) | Circular image with the graph in the center |
| The sagittal dural sinuses drain into the | Internal jugular vein |
| A small FOV for a wrist exam would achieve | High spatial resolution |
| Bright signal in the prostate would indicate | Restricted diffusion |
| The prostate is posterior and inferior to the | Bladder |
| Hemosiderin Appears_______on T2 images | Hypointense |
| The two best planes for TOF of the subclavian artery‘s are | Axial and sagittal |
| Gradient echo fast breath hold dynamic contrast enhanced imaging of the abdomen utilize | Spoiling |
| Steady state imaging is | When there is leftover transverse magnetization (what occurs if the TR is less than T2/T2*) |
| _______ contrast is dependent on impairment of molecular mobility, sensitivity to Brownian motion, and direction of water diffusion | DWI |
| To compensate for FID | Increase NEX enable flow compensation |
| Tissues with a long T2 Time are bright on | T2 images |
| FDA approved oral contrast agent makes the bowel appear | T1 and T2 hypointense |
| DWI susceptibility in or around strong air tissue interfaces can be decreased by | Increasing the parallel imaging factor |
| ______Is caused by interface of aliased signals at different phases | Moire |
| Parallel imaging artifacts can be compensated for by | Decreasing the parallel imaging factor Increasing FOV |
| Acquiring half of the data in the frequency direction | Partial or fractional echo |
| Using gradient moment nulling | Increases the minimum TE |
| Tissues with short T1 times are bright on | T1 images |
| _______Has a long T1 time and a short T2 time | Fat |
| ______Has a long T1 time and a long T2 time | Water |
| High amplitude signal is found in The | Center of K space |
| Increasing the echo train length | Increases chances of blur |
| Increasing the number of excitations by 2 | Increases SNR by the square root of two |
| Acquiring half of data in the phase direction | Half forurier or zero fill |
| Active shielding | Superconducting windings in the MRI scanner ( reduce fringe Field) |
| Gradient slew rate | The rate of ascent or descent from zero to maximum |
| Echo planer imaging sequences | Are more at risk for TVMF effects |
| Transmit gain | Evaluate accuracy of flip angles |
| On a 1 T, the frequency difference between fat and water is | 147 HZ |
| The common iliac’s join to form the | Inferior vena cava |
| Pre-saturation bands for TOFMRA of the inferior vena cava would be placed | Superior |
| The vertebral arteries originate from the | Subclavian artery’s |
| Bloch | Discovered magnetic momentum |
| The first image of the human thorax was in the year | 1977 (By the Damadian and team) |
| Discovered to DNMR, Fourier transform, and phase and frequency gradients | Ernst |
| fMRI was available in the year | 1991 |
| The Fourier equation | Is for heat transfer |
| Discovered how fields are Altered by each other with charges or currents | Maxwell |
| Developed the first rotating magnetic field | Tesla |
| #Of slices X slice thickness X slice gap = | Anatomic coverage equation |
| Pixel size equation for phase and frequency matrix multiplied = | Pixel area (answer in MM squared) |
| FOV/phase & FOV/frequency = | Pixel size (answer in A x B format) |
| When imaging a patient with metal hardware | Use thin slices, add NEX, small pixels for large matrix, no parallel imaging, wide rBW (decreases chemical shift), use STIR vs fat sat, use fast spin echo over gradient echo |
| Coil location in magnet from closest to patient to furthest from patient | Radio frequency coils, gradient coils, shim coils, main magnet |
| Doubling the magnetic field strength | Doubles the SNR |
| Pneumonic for cranial nerves | Oh, oh, oh to take a family vacation. Go Vegas south hardees |
| Administering gadolinium | Increases T1 weighting Increases SNR |
| Tissues with a_____are bright on T1 and tissues with a______are bright on T2 | Short; long |
| Gadolinium only affects______protons | Water |
| 60,000 / HR minus delay and trigger window | Available imaging time |
| 60,000/HR | Effective TR |
| TR x Phase matrix x NEX x #of slices = | 3-D gradient echo scan time |
| TR x phase matrix x NEX / ETL = | Fast spin echo scan time |
| TR x phase matrix x NEX = | Spin echo scan time |
| Calculate pixel area and x by slice thickness | Voxel volume (answer in MM cubed) |
| Increasing the TE | Decreases SNR increases T2 weighting Increases susceptibility/distortion |
| Large FOV | Decreases spatial resolution, increases SNR, no affect on scan time |
| Small FOV | Decreases SNR, increases spatial resolution, no effect on skin time |
| High NEX value | Increases spatial resolution, increases scan time, increases SNR |
| Low NEX value | Decreases spatial resolution decreases scan time decreases SNR, increases motion artifact |
| Increasing flip angle | Increases T1 weighting |
| Decreasing phase encodings | Decreases scan time, decreases spatial resolution, increases SNR |
| Artifacts that occur in the frequency direction | Chemical shift, zipper |
| Artifacts that occur in the phase direction | Motion, ghosting, zipper, parallel imaging, alaising |
| Artifacts in the slice selection direction | Chemical shift, partial voluming |
| Increasing parallel imaging factor | Decreases scam time, decreases chemical shift artifact, increases sharpness, decreases SNR |
| Decreasing receiver bandwidth by 2 | Increases SNR by the square root of two |
| #Of shots and fast spin echo | # phase and coatings/ETL |
| Level two personnel | Those with extensive MR training |
| Level one personnel | Those with limited in MR training |
| Best plane for the ovaries is____and to the best plane for the uterus is_____ | Coronel; sagittal |
| Edema is bright on T2 images because it has a | Long T2 time |
| Edema is dark on T1 images because it has | Long T1 time |
| Removal of signal from vessels | Spatial pre-saturation |
| Can increase the strength of the magnet by increasing the turns of wire in a | Superconducting magnet |
| Uses solenoid Surface coils | Permanent magnet with a vertical field |
| Narrow bandwidth | Increases SNR, increases spatial resolution, increases scan time, increases chemical shift |
| Wide bandwidth | Decreases SNR, decreases special resolution, decreases scan time, decreases chemical shift, decreases TE, increases sharpness |
| Transmit bandwidth can affect | Slice thickness, spatial resolution, SNR |
| Receiver bandwidth can affect | SNR and chemical shift |
| Collecting low frequency/ high amplitude signal at the start of the scan | Elliptic centric K space filling |
| During dynamic contrast enhanced imaging for vasculature, Cay spaces filled_______to ensure well visualized contrast | Centrically |
| TOF MRA | Uses T1 gradient echoes with TR and flip angle selections to suppress signal from stationary tissue, visualizing flowing blood by flow enhancement |
| PC MRA | Uses T2 gradient echo with perimeter selections that suppress stationary tissues and rely on velocity induced phase shift |
| Increasing the flip angle by 2 | Increases RF absorption by four |
| Decreasing the field of view by two | Decreases voxel volume by four |
| Thin slices | Decrease SNR, increase spatial resolution |
| Thick slices | Increase SNR, decreased spatial resolution |
| Thin sliced gaps | Increase spatial resolution and SNR |
| Thick sliced gaps | Decrease special resolution and SNR |
| Decreasing the number of slices in volume imaging | Decreases SNR and decrease scan time |
| Small matrix | Decreased spatial resolution, increase SNR, decreased scan time |
| Large matrix | Increase spatial resolution, decrease SNR, increase scan time |
| Hi TR values | Increase Scan time, increase special resolution, increased number of slices, increase SNR, decrease T1 weighting |
| Low TR values | Decrease scan time, decrease spatial resolution, Decrease SNR, decrease the number of slices, increase T1 weighting |
| Active shimming | Adjustment of current within the shim coils for the purpose of achieving uniform fat suppression |
| Transverse magnetization | XY component at right angles to the main magnetic field |
| Net magnetization vector | Sum of the magnetization from a spin system |