2brt settings
2brt initial vent settings
| Question | Answer |
|---|---|
| FIO2 to acive a result of what | PaO2 of 60 and SAT of 90% |
| Tidal vol. set at | 8 to 10 kgs of IBW |
| Rate (Fequency) | 12 to 16 |
| Pleautau presure | <30 |
| I:E ratio | 1:2 higher E value for COPD patients |
| PEEP set at | 3 to 5 |
| Flow set at | 60 to 80 |
| high preasure alarm | 10 above PIP that pt. does |
| low presure alarm | 10 below PIP that pt. does |
| high tidal vol. alarm | 100 above what tech. sets |
| low tidal vol. alarm | 100 below what tech. sets |
| apnea alarms intervals | 20 to 30 seconds |
| PEEP alarm | 2 to 3 below set PEEP |
| High respiratory rate alarm | double what pt. is doing |