Question
- position within the tissue where the values of acoustic impedance change
click below
click below
Question
Normal Size Small Size show me how
ULTRASOUND 3
Question | Answer |
---|---|
"tissue interface" - position within the tissue where the values of acoustic impedance change | ACOUSTIC BOUNDARIES |
weakening of the sound as it propagates through a medium | ATTENUATION |
Reduction in amplitude and intensity as sound travels | ATTENUATION |
unit of attenuation | DECIBEL |
Factors affecting attenuation | Medium and Frequency |
______will not only occur in the beam of sound produced by the transducer as it produces propagates through tissue, but also in returning ______ as the travel back to the transducer. | Attenuation, echoes |
______ frequencies are more attenuated than _____ frequencies. | HIGHER, LOWER |
____ frequency penetrates deeper than _____ frequency because they are attenuated to a "lesser" degree | LOWER, HIGHER |
Where the molecules of the tissue are densely packed, attenuation will be much g____ than in l____ densely packed tissue | GREATER, LESS |
If path length "increases", attenuation ____ | INCREASES |
attenuation that occurs with each centimeter the sound wave travels | ATTENUATION COEFFICIENT |
unit of attenuation coefficient | dB/cm |
If the attenuation coefficient increases, attenuation _____. | INCREASE |
For soft tissues, the typical value for attenuation coefficient is __. | 0.5 dB)cm. |
term used to describe the average tissue that makes up the soft tissues of the human body (e.g., liver, kidney, spleen). | SOFT TISSUE |
Process by which energy in the ultrasound beam is transferred to the propagating medium, where it is transformed into a different form of energy, mostly heat. The medium is said to absorb energy from the beam | ABSORPTION |
The rate of absorption is directly ________ to the frequency. Higher frequency results in increased ultrasound absorption. | PROPORTIONAL |
two large structures of significantly different acoustic impedance form an interface, the interface becomes a reflector and some of the wave energy is reflected back to the transducer | REFLECTION |
the major interaction of interest for diagnostic ultrasound | INCIDENT ENERGY AND ECHO |
the sound that hits an acoustic interface | INCIDENT ENERGY |
reflected beam | ECHO |
As we have seen, the percentage of incident energy that is reflected depends upon the _____ | ACOUSTIC IMPEDANCE MISMATCH |
The "values of Z for the soft tissues" are quite similar to one another. We conclude that reflections at boundaries between soft tissue will give rise generally s__ e__. | SMALL ECHOES |
The "Z value for bone" is several times higher than the soft tissue average.We conclude that reflection from gas/soft tissue interface gives rise to a l___e___. | LARGE ECHOES |
The "value of Z for air (and other gaseous material)" is much lower than the soft tissue. We conclude that reflection from gas/soft tissue interface gives rise to a ____. | VERY LARGE ECHOES |
ANOTHER FACTOR THAT AFFECTS THE STRENGTH OF THE REFLECTION IS THE SIZE OF THE REFLECTOR. Reflectors can either be ___ or ____. | SPECULAR OR DIFFUSE |
- boundary is smooth and larger than the beam. - Angle of incidence = angle of reflection | SPECULAR REFLECTOR |
- reflecting interface is irregular in shape and its dimensions are smaller than the diameter of the ultrasound beam. - Incident beam is reflected in many different directions. | DIFFUSE REFLECTOR |
describe when the beam is perpendicular to the interface | NORMAL INCIDENCE |
Denotes a direction of travel of the incident ultrasound that is not perpendicular to the boundary between two media | OBLIQUE INCIDENCE |
- when an ultrasound waves strikes a boundary or interface between 2 small structures and the wave is scattered in different directions - responsible for providing the internal texture of organs in the image | SCATTERING |
when the scatter in equal in all direction | RAYLEIGH SCATTERING |
- change in direction of a sound beam as it enters the medium. - Transmission with a bend | REFRACTION |
If the angle of incidence is 90 degrees, no refraction will occur. The physics of refraction are described by | SNELL'S LAW |
two requirements for REFRACTION to ovcur are: | OBLIQUE INCIDENCE AND DIFF. PROPAGRATION SPEED ON EITHER SIDE OF THE BOUNDARY |
As a beam of ultrasound travels through it will diverge. This divergence will result in the same power spread over larger area. The intensity of the beam will therefore be reduced | DIVERGENCE |
convert one form of energy to another | TRANSDUCER |
G E N E R A L C O M P O S I T I O N O F A N UL T R A S O U N D T R A N S D U C E ivL | PHYSICAL HOUSING , ELECTRODES, PIEZOELECTRIC ELEMENTS BACKING MATERIAL , IMPENANCE MATCHING LAYER |
- contains individual components - Provides structural support and acts as an electrical and acoustic insulator | PHYSICAL HOUSING |
- "Grounded electrode - " Protects patients from electric shock. | OUTSIDE ELECTRODE |
- "Live Electrode - " Abuts against a thick backing block. | INSIDE ELECTRODE |
- " transducer element" - most important component - Approx. 6-19 mm in diameter and 0.2-2 mm in thickness - Crystalline materials - dipolar molecules | PIEZOELECTRIC ELEMENT |
naturally occurring material with piezoelectric properties | QUARTZ |
- Commonly used materials - man made ceramic - more efficient, better sensitivity and can be easily shaped | LEAD ZIRCONATE TITANATE(PZT) |
Greek term "piezo" means___ and Elektron ____ | To press and ember |
- Was described 1880 by Pierre and Jacques Curies - some materials produced a voltage when deformed by an applied pressure. | PIEZOELECTRICITY |
Eliminate the vibrations from the "back" face and to control the length of vibrations from the face of the crystals. | BACKING/DAMPING MATERRIAL |
- Sandwiched between the piezoelectric crystal and the patient. - Has acoustic impedance value halfway between that of the crystal and soft tissue. | MATCHING LAYER |
Area through which the sound energy emitted from the transducer travels | ULTRASOUND BEAM |
2 regions of Ultrasound beam | NEAR FIELD AND FAR FIELD/FRESNEL ZONE OR FRAUNHOFER ZONE |
Region "nearest" the transducer face, characterized by a highly collimated beam with more uniform intensity. | NEAR FIELD(FRESNEL ZONE) |
Region "farthest" from the transducer and characterized by the divergence of the beam with great variation in intensity. | FAR FIELD(FRAUNHOFER ZONE) |
location where the beam reaches its minimum diameter | FOCUS OR FOCAL POINT |
energy from the transducer that radiates at various amgles from the transducer face | SIDE LOBES |
arrangement of crystals within the transducer | ARRAY |
types of electronic array transducer | LINEAR ARRAY, CURVILINEAR(or SECTOR) ARRAY TRANSDUCER,PHASE ARRAY TRANSDUCER |
physical size of yhe part of the ultrasound that contacts the patient | FOOTPRINT |
width of the image that is seen on the screen | FIELD OF VIEW |
produces parallel scan lines and has a rectangular field view | LINEAR ARRAY |
- image superficial structures and vessels - operate at frequencies above 4MHz - used for vascular, small parts | LINEAR ARRAY |
- similar to the linear areay but the transducer face is formed curve(convex) - "sector arrays" | CURVILINEAR ARRAY TRANSDUCER |
- operate ar lower frequencies - 3.5 MHz - best suited to image deep lying strictures - application: abdominal and obstetric | CURVILINEAR ARRAY TRANSDUCERS |
PHASED ARRAY | |
transducer designed to enter the body via the vagina, rectum, esophagus or blood vessel(catheter-mounted type) | INVASIVE TRANSDUCERS |
1. ultrasound transducers should not be sterilized | TRANSDUCER CARE AND CLEANING |
temperature at which polarization in a crystal is lost | CURIE POINT |