Save
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.
Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't Know
Remaining cards (0)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

Imaging Equipment

Final Review-Radiologic Technology

QuestionAnswer
Keeping the last image displayed on the monitor Last image hold
The time it takes for the x-ray tube to be switched off extinction time
Grouping sets of 4 smaller pixels together to form a larger effective pixel when taking spot films binning
Actual radiographic technique factors area adjusted to compesate for fluctuations ABS
Amplification of the electronic signal (current) is increased or decreased as needed by the circuits AGC
A series of short visual checks, rather than continuous fluoro, dramatically reduces cumulative exposure to the patient intermittent fluoro
Scattered random light from the output phosphor veiling glare
Lower frame rates than realtime fluoro/ automatically spreads out the fluoro on time to reduce patient dose pulsed fluoro
Indicates the emission of light scintillation
Reduces the FOV on the input phosphor magnification mode
Slight magnification and increased blur around the periphery of the image pin cushion distortion
Loss of brightness toward the periphery of the image vignetting
The ratio of input area to the output area of the image intensifier minification
converts the xray image into an electron beam that will be accelerated and focused for intesification onto a TV Image intesnsifier
The time it takes for the xray tube to be switched on interrogation time
T/F Grid cut-off is not an issue in fluoroscopy TRUE
Primary source of operator radiation exposure during fluoroscopy Patient
Image intensifiers at a basic technological level Energy conversion tubes
What negatively impacts spatial resolution in fluoroscopy Binnning
Concentration of electrons onto an output phosphor much smaller than the input phosphor Minification
How long the x-ray is off for pulse interval
smaller anatomical area is displayed with the same number of pixels available on the monitor multifield intensification
number of photons generated at the output phosphor for every photon generated at the input phosphor flux gain
Increases by the mAs setting as more photon flux is needed pulse height
Duty Cycle Pulse width
Advantages of the flat panel detector over the conventional image intesifier. (3 things) High SNR More FOV Wide dynamic Range
Who invented the fluoroscope Thomas Edison
II concave shape main purpose distance between photocathode and output phosphor is exactly equal
Fluoro grids match Input phosphor (circular)
Key advantage of spot filming as compared to the dynamic image improved spatial resolution
Stopping and starting fluoroscopy using a dead-man's switch and under manual control of exposures intermittent fluoroscopy
In order to shift the electron stream focal point to the correct position in a multi-field image intensifier a stronger negative charge is placed on the electrostatic focusing lens
Minimum source to table distance of a mobile fluoroscopic unit within QC measures 30 cm (12 inches)
To maintain the same brightness gain, a decrease in flux gain would require Increase in minification gain
Longer widths of pulsed fluoro results in Increased patient exposure (and more blurring)
Conventional fluoroscopic mA is usually about 0.5-5 mA
FPD fluoroscopy (3 things) cover more anatomy than i.i. Rectangular shape Capable of radiographic exposures
Major distinction between digital and flat panel fluoroscopy system designs is the position of fluoroscopic xray tube
When utilizing Mag mode 2, what should you except to happen when switching to mode 1 Decreased spatial resolution
What two things happen with mag mode will increase patient exposure, Resolution
Describe the term "gain" in I.I. Amplification of electrical signals
Product of minification gain and flux gain is calculated to determine Total brightness gain
two types of electrical devices that will change the level of brightness to compensate for differences in part thickness and tissue density automatic stabilization control Automatic gain control
placing the tube beneath the patient beneficial more intense radiation levels below the knees of the operator
Volt potential difference needed to accelerate electrons rom the cathode side to the anode side of I.I. 25,000-30,000
Electronically, functional unit of FPD in fluoro DEL
Operating the image intensifier tube in the____ mode can help reduce pincushion distortion magnified
Photocathode is an electrical device which emits electrons
T/F Placing a fluoroscopy room too close to an MRI department, could cause a distortion artifact to an image intensifier system True
The output phosphor of an I.I. changes _______ energy into ________energy electron;light
CCD video imaging uses what type of technology Solid-state computer chip
T/F Longer frame rates allow less smearing if a fluoroscopic procedure requires panning or sweeping False
The product of minification gain and flux gain is calculated to determine Total brightness gain
For image intensifiers, there are two types of electrical devices that will change the level of brightness to compensate for differences in part thickness and tissue density, the automatic stabilization control Automatic gain control
Placing the x-ray tube beneath the patient.. Beneficial because there is more intense radiation levels below the knees of the operator
What is the volt potential difference needed to accelerate electrons from the cathode side to the anode side of an image intensifier 25,000-30,000
Electronically, the functional unit of FPD in fluoroscopy is the Detector element (DEL)
T/F Signal to Noise (SNR) ratios should be kept as high as possible True
What innovative software is strongly recommended for all mobile and trauma x-ray procedures, allows to work as the patient lies. No grid cutoff from angling, alignment and positioning Virtual grid software
T/F High DQE, like 0.8, reflects a low transfer of information from the detector to the output image False
If the dynamic range or bit depth of a digital processing system is too limited, it is possible for ________ to occur when either brightness or contrast are adjusted Data clipping
Ghost images can be caused by all of these except Under exposure of the IR
T/F Sudden deviations in focal spot size indicate damage to the cathode False
T/F Quality Assurance is for monitoring quality patient care, and QC is about monitoring the quality of our equipment True
This describes the power of the light flow (flux) traveling through space Lumens
For light emitted isotropically in all directions from a source, one candela generates one lumen of light intensity per meter Steradian
Spatial Frequency is measured in (three things) LP/mm, Lp/cm, Cycles/pixels
"Black Bone" is another way to say the image is Positive
CR QC requires beam angulation cannot be off by more than 2 degrees
The kVp of any x-ray beam should not deviate by more than ___% from the read-out from the kVp 5
Post processing technique that is key in displaying C7-T1 junctions is Targeted area brightness
To prevent aliasing artifacts, a CR image should be processed with the laser scanning lines and the short dimension of the grid lines in perpendicular
The post processing procedure called stitching means 2 or more images are combined together
Size of the projected light field (collimator) must be within plus or minus the x-ray beam 2% of SID
Halo Effect occurs as a result of too much Edge enhancement
On SMPTE test pattern and other test patterns for electronic image display monitors, there are sets of adjacent squares with JNDs in density between them Contrast resolution/gray scale
What QC calibration should fall within plus or minus 5% exposure timer and exposure reproducibility
Ever radiographic image must include Date of exam RT LT marker Patient DOB
Are Ei or S number required to be ON the image No
The use of repeat analysis in a radiographic imaging department provides pertinent direction Inserve topics management decisions cost-efficiency data
T/F DQE is better in computed Radiography vs. Indirect Radiography False
Which filter application uses waves in relations to space to change/enhance the image Frequency
Stationary Fluoro 38cm (15 inc)
X= output signal amplitude divided by input signal amplitude MTF
What QC checks can be reasonably performed visually by a radiographer Field uniformity Erasure thoroughness or ghosting Uneven spatial resolution
Repeat rate to strive for 3-5%
mA station be linear to two adjacent stations 10%
X= SNR squared of output divided by SNR squared of input DQE
<30% from average: consistency of a single brightness level displayed across the area of the display screen luminance uniformity
+/- 5% The ability to repeat the same overall technique settings and obtain the same results iin actual exposure exposure reproducibility
Assesses dexel reactions to an exposure for evenness of exposure, utilize a long SID to decrease the anode-heel effect Field uniformity
The thickness of absorber material required to reduce the intensity of the x-ray beam to one-half its original value-- 2.5 mm Al or more Half-value layer
+/- 10% exposure accuracy linearity that requires a back up time check as a part of its QC AEC reproducibility
Utilizes LP/mm on a lead template to assess focal spot accuracy Slit camera
+/- 2% SID can use plexiglass cylinder or 8 pennies to measure accuracy collimator check
+/- 5%; High voltage diaode or oscilloscope is used to measure accuracy kVp calibration
+/- 5% monitors the time allowed for the electrons to flow from the cathode to the anode for a specific period of time exposure timer
an increase in FSS at increased tube current and/or decreased tube potential; effects spatial resolution focal spot blooming
Within 10% of each other; the accuracy of a particular mA station relative to the 2 adjacent mA stations mA linierty
Uses a SMPTE pattern to evaluate monitor quality image display system
+/- 2% SID; use a tape measure to measure from the anode (red+) to the imaging plate distance
10% differences in adjacent squares on SMPTE pattern within 10% accuracy grayscale standard display funciton
Artifact occurs during CR processing due to grid lines overlapping laser readouts Miore and Aliasing
The minimum filtration required on an x-ray unit 2.5 mm Al equivalency
T/F low MTF values like 0.2, meants the image is displaying high spatial resolution/frequency True
The post process procedure called translation means an image is flipped right to left or vice versa
T/F HVL improves beam quality True
with QC program in place, what is the majority of repeat exposure Positioning
T/F CR image can avoid alignment issues if 1/4th of the image receptor is exposed False-- Must be 1/3 of the IR
Single CR phosphor plate is erased and then processed without exposing it to xrays. This is a test for intrinsic (dark) noise
Device designed to directly measure luminance, or the light intensity emitted from the display monitor photometer
Clss 1 LCD must have _____ or fewer bad pixels across the entire area of the monitor 15
Dynamic range compression Supports tissue equalization
How many pixels does it take for a monitor to fix a dead pixel 8
Represents the point in the histogram where pixels must start to be eliminated from brightness correction calculation Smax
Histogram, what does the x-axis represent Specific exposure values captured by the receptor
Which of the following modalities generally produces the sharpest images, due to large image matrix size and very small pixels DR
Selection of a narrow window width will increase contrast in the image
determined by the computer by finding the mid-point between SMIN and SMAX in the main lobe of the generated histogram for the image exposure indicator
A marker for exposure quality that porvides the radiographer with an indication of accuracy of their technique Exposure index
Ratio or percentage of original xray beam intensity absorbed by a particular tissue area in the patient Attenuation coefficient
A post processing funciton that changes raw data into a favorable image Rescaling
Quanitfies the difference between the actual EI and target EI Deviation index
Range of pixel values actually present in a displayed image at 1 moment Gray scale
Gradual acceptance over time to use higher radiographic exposures, and hence doses to the patient, for the same xray exam and projection dose creep
Maximum shades of gray a system store for you to use per pixel bit depth
Frequency of recorded exposure values recorded by the image receptor histogram
Range of exposures that produce quality images at an appropriate patient dose exposure latitude
Process of the ADC taking electronic signals and changing them to brightness levels displayed in pixel values Quantization
Range of exposures that can be captured by the detector Dynamic Range
Detector elements are responsible for capturing the aerial image and converting x-rays into electric signals
T/F the more zoom applied to an image the better the spatial resolution is represented False
The higher the fill factor in the dexel the _____ the spatial resolution better
T/F using incorrect collimation can create a histogram analysis error True
Saturation of an image occurs when Exposing the patient to 8 times too much radiation
What step in digital image processing identifies the useful pixel values in an image histogram analysis
Which device can be used in place of a photomultiplier tube in CR reader CCDs
Histogram analysis failure is less common with DR systems because the DR algorithms use the data set for processing only exposed dexels
T/F the tail lobe on a histogram represent barium or a prosthesis False
xray exposure above the level of detectors dynamic range Saturation
What would result in a displayed digital image which is brighter decreased window level
Element for scintillation Gadolinium oxysulfide
Element for Direct photoconductor Amorphous Selenium
Element for the phosphor layer Barium Fluorohalide
Element for Indirect Photodiode Amorphous Silicon
Element for CR processor Helium Neon
CR or DR Scinitllator DR
CR or DR TFT DR
CR or DR Digitizer CR
CR or DR Translation CR
CR or DR Double exposure artifacts CR
CR or DR F centers CR
CR or DR Radiation fog artifacts CR
CR or DR Dexels DR
CR or DR photostimulable phophors CR
CR or DR Latent image CR
What is the bit depth of a 6 bit system 64
Steep S curve through a histogram represents a ______ image short gray scaled
Fill factor of a detector element xray sensitive portion of the detector
Range of the image gray scale is controlled window width
For a large abdomen which completely covers the image receptor place the expected shape of the original histogram with no spikes, only the main lobe
What is the role of the PMT-Photomultiplier tube collect and multiply the light photons
What aspect of an image receptor system creates bursts of light Scintillation layer
T/F Cesium Iodide is an element utilized in scintillation system for a direct capture DR imaging False-- scintillation is not used in Direct capture, just indirect
CR, the image information that is stored on the imaging plate prior to processing latent image
Rad must view images within the steridian
Device that acts as a switch to release the electric charge collected by the detector element TFT
T/F Lossless compression is adequate to sustain high spatial resolution in digital imaging True
Which device in the image receptor converts light into electrons Photodiode
The laser light in the CR exposes the plate in a rapid back and forth motion Raster pattern
Luminescence refers to a release of light photons
What is the initial response of CR phosphors when the phosphor layer of the imaging plate is stimulated by xrays Release of light photons
Imaging and archiving systems in radiology communicate using what digital format DICOM
A reference histogram that evolves by averaging it with the previous 50 procedures neural
What is the bit depth capacity of the human eye 32
Created by: rad123
Popular Radiology sets

 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards