chapter 15 fluoro
Question | Answer |
---|---|
mA for fluoroscopy | 0.5-5 mA |
length and diameter of image intensifier tube | 50cm long, 15-58cm diameter |
5 parts of image intensifier | 1. input phosphor 2. photocathode 3. focusing lenses 4. accelerating anode 5. output phosphor |
material of input phosphor | cesium iodide |
material of output phosphor | silver-activated zinc cadmium sulfide |
brightness gain | expression of ability of image intensifier tube to increase image brightness |
flux gain | ratio of number of light photons at output phosphor to number of light photons emitted in input phosphor + represents tube's conversion efficiency |
conversion factor | expression of luminance at output phosphor divided by input exposure rate |
Automatic Brightness Control (ABC) | function of fluoro unit adjusting kVp, mA, or both |
magnification mode | increases voltage to electrostatic focusing lenses |
in magnification mode, the focal point is shifted _______ from output phosphor | farther |
trifocus image intensifier ________________ can be operated in any of these 3 modes | 30/23/15 |
distortion: | misrepresentation of true size or shape of object |
2 types of distortion from fluoro: | unequal magnification (pincushion) and loss of brightness around periphery (vignetting) |
image noise: | results when insufficient info is present to create image (not enough xrays expose input phosphor) |
image noise is fixed by ______ __ | increasing mA |
2 tv cameras: | camera tube and Charge-Coupled Device (CCD) |
image intensifier can resolve ______lp/mm | 5 |
monitor can resolve ___to___lp/mm | 1 to 2 |
cassette spot filming | static imaging process using radiographic cassette to obtain an image |
Flat-Panel Detectors are used in ________ fluoroscopy | digital |
Flat-panel detectors: | cesium iodide amorphous silicon indirect capture detector |
Application-Specific Integrated Circuits (ASIC): | minimize noise, maximize readout speed, allow switching from low to high dose |
Backlighting: | erase detector between frames, preventing ghosting |
advantages of flat panel detectors in place of image intensifiers: | reduce bulk+weight don't degrade w/age more durable less artifacts reduce dose 50% better contrast resolution postprocessing options |
fluoro features: | Last Image Hold (LIH) for virtual collimation frame averaging, reducing pt dose option to include grid selection of added filtration |
frame averaging __________spatial resolution | decreases |