chapter 15 fluoro

Quiz yourself by thinking what should be in each of the black spaces below before clicking on it to display the answer.
        Help!  

Question
Answer
mA for fluoroscopy   0.5-5 mA  
🗑
length and diameter of image intensifier tube   50cm long, 15-58cm diameter  
🗑
5 parts of image intensifier   1. input phosphor 2. photocathode 3. focusing lenses 4. accelerating anode 5. output phosphor  
🗑
material of input phosphor   cesium iodide  
🗑
material of output phosphor   silver-activated zinc cadmium sulfide  
🗑
brightness gain   expression of ability of image intensifier tube to increase image brightness  
🗑
flux gain   ratio of number of light photons at output phosphor to number of light photons emitted in input phosphor + represents tube's conversion efficiency  
🗑
conversion factor   expression of luminance at output phosphor divided by input exposure rate  
🗑
Automatic Brightness Control (ABC)   function of fluoro unit adjusting kVp, mA, or both  
🗑
magnification mode   increases voltage to electrostatic focusing lenses  
🗑
in magnification mode, the focal point is shifted _______ from output phosphor   farther  
🗑
trifocus image intensifier ________________ can be operated in any of these 3 modes   30/23/15  
🗑
distortion:   misrepresentation of true size or shape of object  
🗑
2 types of distortion from fluoro:   unequal magnification (pincushion) and loss of brightness around periphery (vignetting)  
🗑
image noise:   results when insufficient info is present to create image (not enough xrays expose input phosphor)  
🗑
image noise is fixed by ______ __   increasing mA  
🗑
2 tv cameras:   camera tube and Charge-Coupled Device (CCD)  
🗑
image intensifier can resolve ______lp/mm   5  
🗑
monitor can resolve ___to___lp/mm   1 to 2  
🗑
cassette spot filming   static imaging process using radiographic cassette to obtain an image  
🗑
Flat-Panel Detectors are used in ________ fluoroscopy   digital  
🗑
Flat-panel detectors:   cesium iodide amorphous silicon indirect capture detector  
🗑
Application-Specific Integrated Circuits (ASIC):   minimize noise, maximize readout speed, allow switching from low to high dose  
🗑
Backlighting:   erase detector between frames, preventing ghosting  
🗑
advantages of flat panel detectors in place of image intensifiers:   reduce bulk+weight don't degrade w/age more durable less artifacts reduce dose 50% better contrast resolution postprocessing options  
🗑
fluoro features:   Last Image Hold (LIH) for virtual collimation frame averaging, reducing pt dose option to include grid selection of added filtration  
🗑
frame averaging __________spatial resolution   decreases  
🗑


   

Review the information in the table. When you are ready to quiz yourself you can hide individual columns or the entire table. Then you can click on the empty cells to reveal the answer. Try to recall what will be displayed before clicking the empty cell.
 
To hide a column, click on the column name.
 
To hide the entire table, click on the "Hide All" button.
 
You may also shuffle the rows of the table by clicking on the "Shuffle" button.
 
Or sort by any of the columns using the down arrow next to any column heading.
If you know all the data on any row, you can temporarily remove it by tapping the trash can to the right of the row.

 
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how
Created by: JuliaW2003