Save
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.
Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't Know
Remaining cards (0)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

CT Registry Prep 1

QuestionAnswer
_______ blood pressure indicates the pressure within the arteries during cardiac contraction, should be less than ____ systolic 120
_______ blood pressure is measured during relaxation of the heart and should be less than _____ diastolic 80
_____ is a device used to measure pulse and respiratory status. can be placed ____, ____, or ____. Should be within ________ pulse oximeter, finger, toe, or earlobe, 95% to 100%
Normal respiration for child 20-30 resp
__________ refers to the series of blood flow related events that occur from the beginning of one heartbeat to that of the next. cardiac cycle
It is the frequency of the _________ that determines the patients heart rate. cardiac cycle
__________ is a graphic representation of the electrical activity of the heart. Used during cardiac CT to evaluate heart rhythm and cycle electrocardiogram (ECG or EKG)
Atrial systole is contractions of left and right atria, corresponds with the ______ p wave
Ventricular systole is contractions of the left and right ventricle, corresponds with the beginning of the ______ QRS complex
Complete cardiac diastole is a period of ______ after heart contracts, corresponds with the T wave Relaxation,
Cardiac CT images are typically reconstructed from the data acquired during ________ diastolic phase
Patients with _____ heart rates exhibit longer diastolic phases, which yield higher quality cardiac CT exams slower
________ may be used to reduce the patients heart rate for cardiac CT. 65 beats is prefered for optimal imaging. Beta-adrenergic (b-blockers)
If not contraindicated, sublingual ________ may be used before cardiac CT to cause dilation of coronary vessels improving the vision nitoglycerine
Normal adult patient temperature 97.7 F- 99.5 F (36.5 C- 37.5 C)
Normal respiration for adult 12-20 resp
Normal adult patient pulse 60-100
Normal patient blood urea nitrogen (BUN) 7-25 mg/dL
Normal patient creatinine (Cr) .5-1.5 mg/dL
Normal patient GFR (Glomerular Filtration Rate) Adults 90-120ml/min can decrease with age 70+/- 14 mL/min/m men & 60+/- 10mL/min/m women
Normal patient PT (prothrombin time) 12-15 sec
Normal patient PTT (partial thromboplastin time) 25-35 sec
Normal patient INR (international normalized ratio) .8-1.2
Normal patient platelet count 140,000-440,000 uL of blood
Is the approximation of creatinine clearnace or the rate by which creatinine is filtered from the blood GFR (glomerular filtration rate)
Is a measure of blood coagulation PT (prothrombin time)
Is utilized for diagnosing deep vein thrombosis and pulmonary embolism d-dimer
The presence of elevated _____ in the blood may indicate recently degraded blood clots d-dimer
If d-dimer is elevated, testing such as CT _____ of the pulmonary arteries may be indicated angiography
Injection of medication or contrast into blood stream is a type of _____ administration parenteral
The volume of contrast is administered at a slow rate over a long period Is called drip infusion
The volume of contrast is pushed into the blood at a rapid rate Is called bolus injection
power injectors are capable of consistently injecting large volumes of contrast agent at a flow rate of up to ________ 5-6 mL/sec
For flow rates up to 3mL/sec, a _______angiocath is used 22 gauge
For flow rated exceeding 3mL/sec, a ________or larger is used 20 gauge (Possible 18 for trauma)
Flushing the tubing with _____ saline immediately after contrast bolus allows for reduction of contrast agent and a streaking artifact 30-50ml
In the event of extravasation, remove catheter, apply __________ compress warm and moist
______ contrast agents belong to a class of substances known as radiopaque contrast media (RCM) positive
Examples of 2 radiopaque contrast media (RCM ) barium and iodine
_________ describes the contrast agents propensity (natural tendency) to cause fluid from outside the vessel to move into blood stream osmolarity
_______ contrast media have greater osmolarity (double of non ionic) due to more iodine (salt) atoms per molecule. Water soluble and mostly harmless to body but have more side affects. ionic
______ contrast media have low osmolality and less toxicity fewer side affects are non salt chemical compounds. non ionic
______ contrast media have the same osmolality as blood and can be offered to improve pt comfort and reduced potential of side effects iso-osmolar (IOCM)
an example of non-ionic iso-osmolar contrast Iodixanol-Visipaque
Enteral Rectal Contrast Media (RCM) are administered orally or rectally
Barium sulfate is used to opacify the GI tract, and typical transit time is______ 30-90 min
3 Reasons barium is contraindicated (not used) suspected perferation, future surgery, aspiration
3 negative contrastS used for CT are water, air, and gas
The range of serum iodine concentration for adequate opacification during CT is ______ 2-8mg/ml
Flow rate for central venous catheter 2mL/sec
3 examples of central venous catheters subclavian lines, implanted access ports, and peripherially inserted central cath (PICC)
For studies of distal large bowel, contrast agent may need to be administered ______prior 4-6hrs
A ____ enema is used to opacify the rectum, the sigmoid, and the distal large bowel 150-300ml
Myelograms uses what method to inject contrast intrathecally
Injecting contrast into joint space is what type of administration intra-articular
2 pretreatments for patients that maybe allergic to contrast that are taken 24 hrs prior to exam are______ diphenhydramine HCL (benadryl) and a corticosteroid
People that are at increased risk for adverse reaction asthma, allergies, renal disease, mult myeloma, diabetes mellitus, pheochromocytoma, sickle cell disease, hyperthyroidism, cardiac disease
contraindications for IV dye allergy to iodine, allergy to iodinated contrast, renal insufficiency/failure
nausea and vomiting Is a mild reaction
vasovagal response (Drop in heart rate and B/P making patient faint) is a moderate reaction
tachycardia from hypotension (Low B/P) is a moderate reaction
profound (severe) hypotension is a severe reaction
laryngeal edema (throat swelling) is a severe reaction
for a bronchospasm reaction to iv contrast what is used to treat bronchodialator-inhaler
for a urticaria (skin rash) reaction to iv contrast what is used to treat diphenhydramine-benadryl
for a hypotension reaction to iv contrast what is used to treat elevate legs and iv fluids
cutaneous (skin) reaction to iv contrast are most common within ________ after administration 3hrs to 7 days
________ is a potentially serious delayed effect of contrast agent administration contrast-induced nephrotoxicity (CIN)
4 Groups of patients at risk for contrast induced nephrotoxicity (CIN) include __________ diabetes, myeloma, advanced age, cardiovascular disease
patients taking metformin are at increased chance of _______ when given IV dye lactic acidosis
what dilates the vessels for cardiac studies sublingual nitroglycerine
What phase of kidneys best show transitional cell carcinoma excretory
A rim of viable ischemic (severely damaged) cerebral tissue that surrounds the infarct core in a patient with an acute stroke is termed _________ ischemic penumbra
arterial phase of liver time range of bolus tracking is ______ 25-35 sec
portal hepatic phase of liver time the range of bolus tracking is _______ 60-80 sec
equilibrium, wash out, or delivery phase of liver the range of bolus tracking is _______ 3-4 min
arterial phase of pancreas the range of bolus tracking Is _______ 20-30 sec
Delayed/Late arterial phase of pancreas the range of bolus tracking is _______ 35-45 sec
venous phase of pancreas the range of bolus tracking is _______ 60-80 sec
corticomedullary phase of kidney the range of bolus tracking is ______ 30-40 sec
nephrographic phase of kidney the range of bolus tracking is _______ 70-90 sec
excretory phase of kidney the range of bolus tracking is _______ 2-3 min
A math/statistical method that allows for motion free images from helically acquired images interpolation
________ is multiple detectors that are the same length uniform matrix array
________ thinnest at center, surrounded by detectors of incrementally increasin widths of z axis adaptive matrix array
________ narrow detectors are positioned midline flanked by wider detectors hybrid matrix array
_________reduces image unsharpness & removes blurring convolution
__________optimal level that is necessary for 3d recon on a ct scan to define region of interest (ROI) threshold setting
2 forms of analytic image recon fourier recon & filtered back recon
________ is the product of the X-ray tube current (in milliamperes) and exposure time per rotation (in seconds). Tube current time product
the full width at half max (FWHM) of scanner is used to describe spatial resolution
________noise- result of an insufficient xray photon flux per voxel quantum
What causes a star like streak, Metal artifacts
What can be done about streaking or what software can take the streaking out reconstruction algorithms can be applied to smooth the streaks caused by the metal, interpolation
What cause streaks and blurring motion artifact
Artifact that results from beam hardening and appears on image as a vague area of increased density in a somewhat concentric shape around the periphery of an image similar to the shape of a cup Cupping or Beam Artifacts
Bands or streaks, reduced by thinner slices and computer algorithms Partial volume artifact
Equipment miscalibration, poor gantry rigidity, tube wobble or a defective detector element causing a bright or dark ring Ring artifact
What noise- signal maybe lost and noise introduced to the recon process by the various electronic components of ct electronic
What noise- artifacts is viewed as noise artifactual
What artifact occurs between contiguos sections due to inconsistant breathing misregistration
areas of ct image that contain abrupt changes in tissue density are electricaly represented by ______ high spatial freq
the dimensions of a voxel are determined by multiplying the ________ pixel dimensions x slice thickness
the rate at which a quantity of xrays emitted from the tube pass through a unit area over a unit of time is called photon flux
Iothalamate Meglumine (Conrad) and Diatrizoate Sodium (Gastroview and Hypaque) which have high osmolarity are what kind of contrast ionic and developed first
Iopamidol (Isovue), Iohexol (Omni), Lopromide (Ultravist) are what kind of contrast nonionic
iodixanol or visi is what kind of contrast nonionic iso osmolar
the radiation dose index calculation that takes into accound the variations in absorption across the vield of view due to beam hardening is called ________ CTDIw (CT dose index weighted)
A math process that allows MDCT images to be recon at any point along the acquired volume is called___________ z-filtering
_________ is the measure of total radiation exposure for entire series of ct images CTDI vol (CT dose index volume)
dose length product is calculated by multiple what CTDI vol and total scan length
_________used to approximate the radiation dose for the ct section acquired during a helical scan CTDI vol
quantity of xrays passing through a unit area. emitted from tube toward pt photon fluence
Limited resolution on modern ct scanner 25 lp/cm
A high resolution comb is utilized by an MSCT detector array is used in an effort to reduce ______ Scatter
What is the distance between centers of 2 reconstructed ct images called section interval
clinical indication of coronary artery calcium (CAC) artherosclerosis
which are tissues other than the structure excluded by setting the threshold value and eliminating pixels above and below 3d shaded surface
when a sharp or high pass recon is used, it does so at the expense of losing Contrast resolution
convolution enhances important characteristics of attenuation profiles thus reduces what streak and star artifact
all mdct scanners that have 4 or more rows of detectors use ____________ not filtered back projection cone beam reconstruction
spatial resolution is dependent on what two things reconstruction field of view (fov) and image matrix
increasing matrix size will ______ voxel size decrease
decreasing reconstruction fov will _____ voxel size decrease
increase slice thickness by _______ voxel size increasing
decreasing reconstruction fov will ________ pixel size reduce
decreasing reconstruction fov will __________ spatial resolution increase
decreasing reconstruction fov will ______ noise increase
decreasing reconstruction fov will _______ size anatomy increase
on mdct penumbra appears on first and last slice
on sdct, penumbra appears around every slice
5mGy= ___ .05 rad
1mGy=___ .01 rad
hypotension, gray tone to skin, tachycardia, cold clammy blue color, confusion Are symptoms of Shock
osmolarity of plasma 285
a bone window utilizes a _____ ww (window width) and _____ wl (window level) Wide High
soft tissues use a ww of ___________ and window level of ___________ 300-400 0-50
cone beam artifact looks like _____ coming from structures around the _____ of the image stars edge
Measure of total pt radiation dose over a given scan acquisition length and is proportional to scan length dose length product (DLP)
units to express total pt dose milligray (mGy) per cm
the 3d ct technique that includes all the voxel info in the recon model with adjustments to opacity volume rendering
reformatted ct that lie perpendicular to the original plane orthogonal
controlling factors for temporal resolution gantry rotation speed and recon method
daily air and water calibrations are used to evaluate linearity
3 Phases of tissue enhancement classifications are________ Bolus and Non-equilibrium and Equilibrium
linear attenuation coefficient for dense bone .528
linear attenuation coefficient fo muscle .237
linear attenuation coefficient for white matter .213
linear attenuation coefficient for gray matter .212
linear attenuation coefficient for blood .208
linear attenuation coefficient CSF (cerebral-spinal fluid) .207
linear attenuation coefficient for water .206
linear attenuation coefficient for fat .185
linear attenuation coefficient for lungs .093
linear attenuation coefficient for air .0004
ct # dense bone +1500-2000
ct # muscle +50 +65
ct # white matter +40 +46
ct # gray matter +35 +40
ct # blood +30 +35
ct # csf +15
ct # fat -50 -100
ct # lungs -550 -950
window level and window width of abdomen +50 L +400 W
window level and window width of brain +40 L +80 W
window level and window width of subdural +50 -100 L +130-300 W
window level and window width of stroke +35 L +10-30 W
window level and window width of lung -600 L +1500 W
Window level and window width of liver +40 L +150 W
Window level and window width of spinal bone +400 L +1800 W
A patient that has severe vagal reaction to IV contrast that includes bradycardia. what do you give them? Atropine
Test bolus and bolus triggering are used for _______ Ensuring imaging during peak enhancement regardless of patient age,disease, or cardiac output
in msct, the beam pitch is equal to _____ table feed per rotation divided by total collimation
calculation of the avg cumulative dose to each section within acquisition of multiple sections msad (multi scan average dose)
if a person is left dominate, the posterior descending artery branches off of_______ left circumflex artery
if a person is right dominate, the posterior descending artery branches off of_______ right coronary artery
what quantifies effective dose sieverts
test phantom of water is scanned at 5 ROI (region of interest) this is done to evaluate cross field uniformity
Mild agatston score for calcium seen in heart 11-100
Moderate agatston score for calcium seen in heart 101-400
Extensive agatston score for calcium seen in heart >400
normal range of cerebral blood volume contained within 100g brain tissue 4-5
Due to it inherent retrospective reconstruction capability the helical CT allows substantially better what? longitudinal spatial resolution
smaller filament size _______ improves spatial resolution
What controls the volume of transmission data acquired for each gantry rotation sampling freq or views per rotation
An increase in sampling frequency increases spatial resolution
A technique that allows the user to select the pixels for a ct 3d reformat is called ________ thresholding
a decrease in apperature size will _____ noise increase
International system SI unit of radiation dose expressed as absorbed dose and can be used for any type of radiation (alph,beta,neutron,gamma) and does not describe biological effects of radiation gray (rad)
Biological effects of radiation are measured in units of Sieverts (older version rem)
ability of a ct system to maintain constant Hounsfield units across homogenous (same) image uniformity
isotrophic resolution is approached by using _____ _____ and ____ _____ thin slices, small dfov (display field of view)
dimensions of a voxel is found by using pixel size and section width
high osmolar contrast has avg osmolarity of 1000-2400
a decrease in tube total rotation = ______ and _________ decrease in time decrease pt dose
decrease in total tube rotation from 360 to 180 is known as half scan or segmenting
Gantry Control to amplifier to software & hardware (s/h) to Analog to Digital Converter (adc) to array processor to host computer to storage or console CT system components left side
console to host computer to scan controller to digital to analog converter (dac) & high volt generator to gantry control then Gantry CT system components right side
this is where instructional commands are entered, info about patient, image parameters, post process technique console
Primary link between tech and other components of imaging system. It translates techs commands entered at the console into the computer language and passes it onto the scan controller or storage device is__________ host computer
host computer passes commands to this which is responsible for timing and operation correct performance of each CT system component once the scanner has been started controls high-voltage generator, gantry and patient tablet scan control
The components that convert the digital commands coming out of the scan controller into analog commands that can be executed by the power producing and mechanical components dac (digital to analog converter)
To save the digital data in case it is needed at any time in the future images of kept the long term storage by Archiving
A pre-defined set of parameters that is used to run a CT scan for specific clinical application Protocol
The negative terminal of the CT x-ray tube in the part of the tube with the negatively charged electrons flow into the tube from the high bolted generator cathode
The positive terminal of the CT x-ray tube in the part of the tube where the target material is located anode
A structure on the cathode terminal of the CT x-ray tube from which the electron beam is released filament
The area on the anode target bombarded by the electron beam and a factor in determining the image resolution focal spot
As the mA increases the number of photons Increase
The perimeter that determines the energy level of x-ray photons produced by the x-ray tube tube voltage kV
As the kV increases the penetrability of the photons Increase
This restricts x-rays to only the selected cross-sectional region Collimation
Has the collimation increases scatter production which results in patient does decreases
What are the two types of detectors used in CT Gas and solid state detectors
When a photon strikes an xenon atom the atom splits into an ion and an electron and can be measured as an electrical impulse
Xenon gas detectors can only be used in what generation scanner which is giving a chamber aligned with the CT x-ray tube at all times because of the narrow chambers Third
What is the most prevalent type of detectors used in today’s CT scanners Solid state
Specialized computer component used to reconstruct a CT images Array processor
A component that aids in the distribution viewing and storage facility for digital images across an imaging facilities computer network picture archiving & communication system (PACS)
The standard used for communicating digital medical information which was established to improve workflow efficiency between imaging systems in healthcare environments worldwide digital imaging & communications in medicine (DICOM)
low window level is brighter
high window level is darker
intrathecal injection goes into subarachnoid space
patients vessels dilate to compensate for the increased volume of fluid. causes the flushed nausea feeling and is directly contributed to the total voume, rate of injection viscosity (Thickness) is what type of reaction chemotoxic
What reaction- result of histamine being release idiosycratic
The methods and tools used to protect patients in personnel from ionizing radiation exposure Radiation safety
Radiologist or medical physicists who are approved by the NRC and state organizations to ensure that institutions comply with guidelines for radiation safety Radiation safety officer
What stage in which the radiation can cause the most harm to the fetus due to the embryo being more sensitive to harmful effects of radiation than an adult First trimester
medical wrong doing or failure to do something a professional negligence most common charge against healthcare workers malpractice
breach or failure to fulfill expected standard of care and an unintentional arming of a person by acting in an improper way negligence
the negligent act speaks for itself res ipsa loquitor
Because in which a person is harmed because of another persons actions or failure to act. Examples are assault and battery and false imprisonment and malpractice tort
Intentionally threatening or attempting to harm a person. And example is telling a person that he is going to do something that he does not want to assault
Intentionally touching another person without a persons permission even if contact is not causing harm. Example is doing something to the patient that the patient has said they don’t want done battery
Intentionally violating a persons freedom. Example is using restraints false imprisonment
Created by: Reeni1995
Popular Radiology sets

 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards