X-ray ARRT QM
Radiology Quality Management
| Type | Requirements |
|---|---|
| X-ray Collimator/PBL | Must be accurate within 2% of SID. |
| X-ray kVp selector | Must be accurate within 10%. ±3-5 kVp. |
| Uncontrolled Area | Must be kept under 20 μSv/week. |
| Secondary Barriers | 2ndary must overlap the 1st by 1/2 inch and touch the ceiling. |
| Device used for quality control testing | Digital dosimeter - several tests can be done with one device. |
| Fluoro dose table-top | Must be limited to no more than 100 mGy/min |
| X-ray mAs adjacent stations/Linearity | Must be accurate within 10% of each other. |
| Fluoro image intensified | Primary beam must never go beyond the limits of the IR. |
| X-ray tube leakage | Must not exceed 1 mGy/hr @ a distance of 1 m from the housing. |
| X-ray bean perpendicularity | Must be accurate within 1 degree.. |
| X-ray kVp linearity | Must be accurate within 10%. |
| X-ray tube speed | Generally most operate @ 3,000 & 10,000 rpm. |
| Lead apron | Must contain at least .25 mm Pb. |
| Thyroid Shield | Must contain at least .5 mm Pb. |
| Controlled Area | Must be kept under 1 mSv/week. |
| Fluoro Curtain | Must contain .25 mm Pb |
| Fluoro Bucky Slot Shield | Must contain .25 mm Pb |
| X-ray Timer | Should be within 5% of the time chosen for exposures over 10ms |
| Exposure reproducibility | Radiation intensity of sequential exposures should not vary more than 5% |
| Fluoro exposure rate | A digital dosimeter, dose should not be more than 100 mGy/min. |
| Fluoro total brightness gain | Ranges from 5,000-20,000 |