Save
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.
Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't Know
Remaining cards (0)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

Boards; memorize

Radiography need to memorize QC, Landmarks, Rad Safety

TermDefinition
Exposure Timer (QC) +/- 5%
Centering (QC) 1% of SID must align to centering indicator
Thickness Aluminum above 70 kVp 2.5 mm Al
Thickness aluminum 50-70 kVp 1.5 mm Al
Distance (SID) (QC) +/- 2% SID
Collimation (QC) +/- 2% SID (?)
Thickness aluminum below 50 kVp 0.5 mmm Al
coccyx Symphysis Pubis and greater trochanter
S1-S2 ASIS
L4 iliac crest
L3-L4 umbilicus
L3 inferior costal margin
L1 Transpyloric Plane
T12-L3 kidneys
T9-T10 xiphoid process
T7-T8 inferior angle of scapula
T4-T5 sternal angle
T2-T3 Suprasternal notch (jugular notch)
Secondary barrier (Pb equivalent) 1/32 inch 0.08 cm 1/2 inch overlap of primary
C7 vertebra prominens
C5 Thyroid cartilage
Brightness display (QC) Photometer Brightness/display contrast betwen 2 areas
C1 Mastoid Process
Monitor Calibration (QC) each monitor calibrated to DICOM GSDF (gray scale standard display function)
Spatial Resolution (QC) line pair tools more lines-higher resolution lp/mm
Reader Calibrations (QC) Produce same response to samem exposure levels
Exposure linearity (QC) 10% between mA stations
Exposure reproducibility (QC) +/- 5% exposure values for same technical selection (measuring mGy)
kVp calibration (QC) +/- 5% of assigned
AEC testing (QC) Same receptor exposure despite any changes in thickness or density and despite change in technical factors
Mobile C-arm 30 cm 12 inch
Leakage radiation allowed 1 mGya at a meter
Thickness aluminum typical inherent 0.5-1mm
mA range in DR fluoro 2-5 mA
fixed fluoro 38 cm 15 inches
Primary barrier Pb equiv 7 ft 1/16 in .16 cm, 1.6mm
Drape/curtain (Pb/equivelent) 0.25 mm 0.5 preferred
Bucky slot cover (Pb equivalent) 0.25 mm 0.5 mm preferred
Glasses (Pb eq) 0.35 mm
Thyroid Sheild (Pb eq) 0.5 mm
Gloves (Pb equivelent) 0.25 mm
Apron (Pb eq) 0.5 mm
Neurological/CNS >50 Gy Death in hours or days
Gastointestinal 10-50 Gy Death in one week
Hematologic 1-10 Gy Death in 3-6 weeks
Embryo Microencephaly 0.2-0.4 Gy
Embryo per month Dose limit 0.5 mSv
Embryo entire period Dose limit 5 mSv
Uncontrolled area occupancy factor 0.1 mGy/week
Controlled areas occupancy factor 1 mGy/week
Public Exposure under 18 100mrem/year (1 mSv)
Public exposure Lens of eye 50 mSv
Public Exposure Annual whole body 5 mSv 10% of occupational
Occupational cummulative effective dose limit 10 mSv x age
Occupational extremity 500 mSv
Occupational Bone marrow 500 mSv
Occupational Breast 500 mSv
Occupational Lens of eye 150 mSv (15 rem)
Occupational Whole body dose 50 mGy
Decreased sperm count (after a few weeks) 100 mGy
Suppression of Menstration 100 mGy
Gonadal dose, temperary infertility 2Gy
Gonadal dose, Permanent infertility 5 Gy
CNS Syndrom occurs at what whole body dose >50 Gy
Grid absorption of the useful beam Cutoff
8:1 Grid Conversion Factor 4
5:1 Grid Conversion Factor 2
6:1 Grid Conversion Factor 3
10:1 Grid Conversion Factor 5
12:1 Grid Conversion Factor 5
16:1 Grid Conversion Factor 6
DAP meter Dose area Product Used for patient Dosimetry
Portion of the collimator that reduces off-focus radiation Shutter closest to the tube port
Unit of measurement for absorbed dose Gray
Average lifespan of an erythrocyte 120 Days
Contrast resolution of a CR/DR system is deteremined by bit depth
mA is selected at Rheostat
dose-response model best represents the basis for radiation protection guidelines linear-non threshold stochastic
Skeletal abnormalities embryo 3 weeks to 20 weeks
Acute diverticulitis contrast medium preferred, especially with a fever water-soluble iodinated
Sum of the individual doses receibed in a given period by a specific population from expsosure to a specified source of radiation Collective Effective Dose
Created by: rad123
Popular Radiology sets

 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards