RadQA Word Scramble
|
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.
Normal Size Small Size show me how
Normal Size Small Size show me how
| Question | Answer |
| Two areas os activity are designed to ensure the best possible diagnosis at an acceptable radiation dose and with minimum cost. These areas are ______________ and _____________. | Quality assurance (QA) and quality control (QC) |
| Quality assurance deals with ___________. | People |
| A program of QA monitors proper patient __________, ___________, and _____________. | Scheduling, reception, and preparation |
| QA also involves image interpretation. Did the patient's ultimate disease or condition agree with the radiologist's diagnosis? This is called _______________. | Outcome analysis |
| All of these QA activities require attention from the imaging team, but they are principally the responsibility of the _____________. | Radiologist |
| A program of ________ is designed to ensure that the radiologist is provided with an optimal image produced through good equipment performance and resulting in minimal patient radiation exposure. | Quality Control (QC) |
| Quality Control deals with _____________ and ____________. | Instrumentation and equipment |
| ____ begins with the x-ray imaging systems used to produce the image and continues with the routine evaluation of image-processing facilities. | QC |
| Each new Rad eqpt must be acceptance tested b4 it is applied clinically. The test must be done by someone other than the _________ cause it is des. to show that the eqpt is working within the manfac’s spec. and is producing and acceptable pt rad dose. | Manufacturer's representative |
| An acceptable QC program consists of three steps: ____________, ________________, and ______________. | Acceptance testing, routine performance monitoring, and maintenance |
| As with QA, QC requires a team effort, but QC is principally the responsibility of the _________________. | Medical Physicist |
| Perphaps the most important patient protection characteristic of a radiographic imaging system is _______________ of the x-ray beam. | Filtration |
| State statutes require that general purpose radiographic units have a minimum total filtration of ___________. | 2.5 mm Al |
| Misalignment must not exceed _____ of the source-to-image receptor distance (SID). | 2% |
| With a PBL collimator, the x-ray beam must not be larger than the image receptor except in the _________________. | Override mode |
| The spatial resolution of a radiographic imaging system is determined principally by the ____________ of the x-ray tube. | Focal-Spot size |
| Three tools are used for measurement of foacl-spot size: ___________, ____________, and ____________. | Pinhole camera, star pattern, and slit camera |
| ____________ should be evaluated annually or whenever an x-ray tube is replaced. | Focal-spot size |
| The ____________ should be evalauated anually or whenever high-voltage generator component have changed significantly. | kVp calibration |
| A variation in kVp of approximately _____ is necessary to affect image optical density and radiographic contrast. | 4% |
| The measured kVp should be within ____ of the indicated kVp. | 10% |
| Exposure timer accuracy should be within ___ of the indicated time for exposure times greater than ___. | 5%; 10ms |
| Accuracy of ____ is acceptable for exposure times of ______. | 20%; 10 ms or less |
| Regardless of the material thickness and the absolute exposure time, the optical density of the processed image should be __________. | Constant |
| If the phototimer fails, the backup timer should terminate the exposure at ____ or _____, whichever occurs first. | 6 s or 600 mAs |
| Exposure linearity must be within ____ for adjacent mA stations. | 10% |
| Sequential radiation exposures should be reproducible to within ______. | +/- 5% |
| Viewbox illumination should be analyzed photometrically on an annual basis. This is done with an instrument called a _______, which measures light intensity at several areas of the illuminator. | Photometer |
| Intensity should be at least _________ and should not vary by more than _____ over the surface of the illuminator. | 1500 cd/m^2; +/- 10% |
| The entrance skin exposure (ESE) for an adult averages ___ to ______ during fluoroscopy; this can result in a skin dose of _____ for many fluoroscopic examinations. | 3 to 5 R/min; 10 rad |
| For interventional procedures, a skin dose of ________ is not uncommon but should be avoided if possible. | 100 rad |
| Federal law and most state statutes require that under noraml operation, the ESE shall not exceed ___________. | 10 R/min |
| For interventional procedures the fluoroscope may be equipped with a high-level control which allows an ESE up to ________. | 20 R/min |
| Measurements for exposure rates are made with a __________ that these levels are not exceeded. | Calibrated radiation dosimeter |
| _________, _______, _______, and _______ filters are used to determine the adequacy of any automatic brightness stabilization (ABS) system. | Lucite, aluminum, copper, lead filters |
| Proper exposure of the cassette spot film depends on the _____, ________, and _____________________________. | kVp, mAs value, and sensitivity characteristics of the screen-film combination |
| An ESE of approximately _______ may be assumed for a cassette spot film. | 200 mR |
| Photofluorospot images uses less _________, require les _____________, and are produced with a lower ____________. | film, personnel interaction, and patient dose |
| Photofluorospot images are recorded on film from the ____________ of an image-intesifier tube. | output phosphor |
| An ESE of approximately _____ may be assumed for a photofluorospot. | 100 mR |
| Each system functions in the manner of the phototimer of a radiographic imaging system, producing constant image brightness on the video or flat panel monitor, regardless of the ___________ or ___________ of the antaomy. | thickness; composition |
| Fluoroscopic AEC should be evaluated __________. | annually |
| With incrementing from one tomographic section to the next, the section level should be accurate to within ________. | +/- 2mm |
| Constancy of ______ from one QC evaluation to the next should be achieved. | +/- 1mm |
| Digital image receptors have unique artifacts associated with ______________. | Pixel failure |
| Environmental radiation can contribute to _____________. | Ghost artifacts |
| The variation in x-ray intensity should not exceed _____. | 5% |
| __________ is a process that monitors and evaluates the quality of the care and services provided to patients within a health care facility. | Quality management |
| The terms ______________ and _____________ are used to encompass all activities related to quality management, including utilization and risk management, infection control, surgical case review, medication usage evaluation, and heatlh record. | quality assurance; quality assesment |
| _____________ activites include work performed by various hospital committees and the medical staff, as well as other professional staff from various hospital departments. | Quality management |
Created by:
o3iiheartmmg04