Merrill Chpt 2 Word Scramble
|
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.
Normal Size Small Size show me how
Normal Size Small Size show me how
| Question | Answer |
| What structures have varied tissue of thickness & density? | AP Thoracic spine, Lateral Hip & Lateral Cervicothoracid spine (swimmer's) |
| Certain filters also decrease skin entrance exposure by? | Absorption of low-energy photons (decreasing dose to patient) |
| What are specially designed attenuating devices that produce images with even density on anatomic structures that vary significantly in tissue thickness? | Compensating filters |
| T/F: Without the use of a compensating filer, images of anatomy with varying tissue thickness would require 2 exposures or be of suboptimal quality? | True |
| What is the simplest and most common filter shape? | Wedge |
| The wedge is made of what material? | Clear leaded plastic (Clear-Pb) |
| Where are the most compensating filers placed? A. Under the IR; B. Between grid & IR; C. Between the tube and the patient; D. Between patient and the IR | C. Between the tube and the patient |
| The trough is made of what material? | Aluminum |
| Is the boomerang A. collimator mounted or B. contact? | B. Contact |
| Blocks the field of light when placed between tube and patient, makes positioning more challenging, & must complete positioning before mounting filter to collimator. These are all? | Disadvantage of aluminum filters |
| Which compensating filter is used for AP projections of the thoracic spine? | Trough |
| Which compensating filer is used for AP projections of the hips, knees, ankles, and long axis of T-spine, foot, & upright abdomen? | Wedge |
| The boomerang filer is used in AP projections of? | Shoulder and facial bones |
| The _________filter has 2 models for lateral and oblique projection use | Ferlic |
| Which filter material can become objectionably heavy, if made for certain exams? | Leaded plastic |
| What is the unique composition of the boomerang filer? | Silicon rubber |
| The physical principles of filters is in? | Shape, Composition & placement |
| The shape and composition of a filter depends on _______to be imaged? | Body part |
| What filter is used on body parts that vary in thickness along the long axis? | Wedge |
| Numerous special shapes have been designed for more technically challenging exams, which filters are these? | Trough, Ferlic, Boomerang |
| What are the 2 categories of filer placement? | Collimator-mounted & contact |
| What 2 places can contact filters be placed? | On patient & between patient and IR |
| T/F: When placed on patient, the filter decreases skin entrance exposure? | True |
| What filter is used on parts that are thin (less dense) on the edges & thick (more dense) in the center? | Trough |
| What filter improves quality of lateral projections? | Ferlic |
| Digital fluoroscopy uses specifically shaped filters to compensate for round image intensifier, they are? | Convex and concave conical-shaped filters |
| CT uses what shape filter to compensate for the rounded shape of head? | Bow-tie |
| What 2 ways is exposure reduced? | Eliminate need for 2 separate images and many absorb low-energy radiation |
| Compensating filters that are placed between the primary beam and the patient have the additional advantage of: A. Not blocking the light; B. Being lightweight; C. Reducing patient dose | C. reducing patient dose |
| Filters placed under the patient may cause ______, which can be objectionable to the radiologist? | Artifacts |
| Which category of filters has no effect on patient exposure, A. Collimator-mounted, B. Contact? | B. Contact |
| T/F: Digital radiology does not require the use of compensating filters | False |
| T/F: Placement of a compensating filter between the anatomy of interest and the IR can produce objectionable artifacts | True |
| T/F: Filters composed of aluminum block the field of light | True |
| T/F: Filters composed of clear leaded plastic block the field of light | False |
| T/F: Filters are specifically designed to be used with only one body part or radiographic exam | False |
| T/F: Pediatric patients rarely require the use of compensating filters | True |
| T/F: Collimator-mounted filters reduce patient exposure by hardening the primary beam | True |
| T/F: The use of improvised filters, such as saline filled bags, is a recommended practice | False |
| T/F: A wedge filter would be used to improve the image quality of the AP projection of the thoracic spine | True |
| T/F: A "bow-tie" filter is a special filter for use in digital fluoroscopy | False |