operation, QA, acquisition, and evaulation

        Help!  

Question
Answer
Fixed spatial resolution characteristic of direct digital systems is determined by   TFT DEL size  
🗑
What is foreshortening?   an object is placed at an angle on the detector (object not flat)  
🗑
grid absorption of useful beam   grid cutoff  
🗑
what is bit depth   number of bits per pixel which has an effect on the gray scale  
🗑
a technique of 85kV and 10mAs with no grid. what mAs is needed with a 8:1 grid   40 mAs  
🗑
what gird ratio is similar to the air gap technique   6:1 ratio  
🗑
formula for magnification   SID/SOD  
🗑
formula for heat units   # of exposures*kVp*mA*seconds*generator  
🗑
multiplier for a three phase 6 pulse generator   1.35  
🗑
multiplier for three phase 12 pulse   1.41  
🗑
find the new intensity for an AP chest x-ray at 55 inches, with an intensity of 50mR. The new distance is 72 inches.   29.2  
🗑
old mAs is 10 with a 6:1 grid. find new mAs with a 16:1 grid   20 mAs  
🗑
purpose of fuses on the xray circuit   safety to protect equipment. will stop exposure if there is too much voltage  
🗑
what is an autotransformer   single coil transformer, operates on self induction, minor adjustments in kVp  
🗑
anode is positive or negative   positive  
🗑
cathode is positive or negative   negative  
🗑
kvp is quality or quantity   quantity  
🗑
mAs is quality or quantity   quality  
🗑
what type of histogram is used for extremities   type 1  
🗑
what histogram is used for KUB's   type 2  
🗑
what histrogram type is used for contrasted exams   type 3  
🗑
primary coil of the high voltage transformer receives what current   AC  
🗑
energy needed to eject k-shell electrons from a tungsten atom   70 kVp  
🗑
three causes for tube failure   1. vaporized tungsten 2. pitted anode 3. cracked anode  
🗑
the x-ray tube is a   vaccuum  
🗑
what is needed to find the total filtration   additive and inherent filtration  
🗑
less than 1% of the energy from electrons are turned into this   x-rays  
🗑
what relationship does the line focus principle show   actual and effective focal spot  
🗑
characteristics of tungsten include   1. heat dissipation 2. high melting point 3. high atomic number  
🗑
what is thermionic emission   electron cloud within the x-ray tube is the product  
🗑
what is aec used for   ensure reproducible radiographs regardless of tissue differences  
🗑
circuit device that permits electrons to flow in one direction   resistors  
🗑
what voltage must the primary side of the x-ray circuit be at   220 volts  
🗑
what percent of voltage ripple happens in a 3 phase 12 pulse have   4%  
🗑
how much voltage ripple occurs in a high frequency generator   less than 1%  
🗑
3 parts of the x-ray circuit   primary, secondary, and filament circuit  
🗑
what is located between the secondary coil of the high voltage transformer and the x-ray tube   filament circuit  
🗑
how is pixel size determined   dividing FOV by matrix  
🗑
interrogation time is   the time needed for the x-ray tube to turn off  
🗑
what are the two types of AEC   ionizing chamber and phototimer  
🗑
what does bit depth identify   gray scale values  
🗑
what effects spatial resolution   OID, SID, focal spot size, and motion  
🗑
magification is effected by   OID and SID  
🗑
elongation and foreshortening are effected by   tube alignment of x-ray tube, anatomic part, and IR  
🗑
visibility factors are   brightness, contrast, and noise  
🗑
what describes DQE   percentage of detected incoming x-ray photons  
🗑
terms used to describe distortion   magnification, elongation, and foreshortening  
🗑
percent of incoming x-ray photons that are identified and absorbed by the detector describes   DQE  
🗑
focal spot blur is greatest   toward the cathode end of the x-ray beam  
🗑
true or false: SID has no effect on contrast   true  
🗑


   

Review the information in the table. When you are ready to quiz yourself you can hide individual columns or the entire table. Then you can click on the empty cells to reveal the answer. Try to recall what will be displayed before clicking the empty cell.
 
To hide a column, click on the column name.
 
To hide the entire table, click on the "Hide All" button.
 
You may also shuffle the rows of the table by clicking on the "Shuffle" button.
 
Or sort by any of the columns using the down arrow next to any column heading.
If you know all the data on any row, you can temporarily remove it by tapping the trash can to the right of the row.

 
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how
Created by: smshelton2
Popular Radiology sets