Digital Radiography: An Introduction Chapter 5 (Euclid Seeram)

Quiz yourself by thinking what should be in each of the black spaces below before clicking on it to display the answer.
        Help!  

Term
Definition
Indirect Detector   A system that uses phosphors to convert x-ray energy into an electrical signal through an intermediate stage that utilizes light photons.  
🗑
Direct Detector   A system that uses photoconductors to convert x-ray energy directly into electrical charge without the intermediate stage.  
🗑
Amorphous Silicon   An Indirect detector has a layer of ______.  
🗑
Selenium   A direct detector converts x-ray to electrical signal with amorphous _____.  
🗑
CCD Chip   The CCD digital detector is based on an indirect conversion process and uses a ____ to convert light to electrical charge.  
🗑
Scintillation Screen   The ______ detects x-rays and converts them into light.  
🗑
Cesium Iodide, Gradolinium oxysulfide   The x-ray scintillation layer used in the direct flat-panel digital detector is usually either _____ ______ or _________ _______.  
🗑
Structured Phosphor/Scintillator   Cesium Iodide crystals are arranged in a needle-like fashion and run in the direction of the x-ray beam.  
🗑
Turbid Powdered Unstructured Phosphor   Gradolinium oxysulfide crystals are deposited as powdered particles in the photodiode array.  
🗑
Turbid Powdered Phosphor   Which type of phosphor or scintillation screen causes lateral spreading of light and decreases the spatial resolution of the image?  
🗑
Photodiode layer   The purpose of a a-Si ______ ______ is to convert the light from the x-ray detection scintillator into electrical charges.  
🗑
Photoconductor   The ______ in a direct digital detector system converts the x-ray to an electrical signal.  
🗑
Thin Film Transistor   An array for readout of electrical charges generated by the photodiode layer in an indirect detector system, or by the storage capacitor from the photoconductor of the direct detector system.  
🗑
Photodiode   The sensing/storage element of an indirect flat-panel TFT detector is the _______.  
🗑
Storage Capacitor   The sensing/storage element of a direct flat-panel TFT detector is the _____ ________ in the photoconductor.  
🗑
Fill Factor   The ratio of sensing area of the pixel to the area of the pixel itself.  
🗑
high   Detectors with ____ fill factors will provide better spatial and contrast resolutions.  
🗑
light optics   The _____ ______ in a CCD digital detector system reduces the output image from the scintillator phosphor to the size of the CCD array thereby reducing the image quality.  
🗑
Signal noise ratio   The light optics reduces the image quality by reducing the ______to_____ _____.  
🗑
photodiode   Indirect detectors use a light-sensitive __________ to capture light and then produce electrical charges.  
🗑
photoconductor   The direct detectors use a a-Se ____________ to convert x-ray energy directly into electrical charges.  
🗑
amorphous selenium   X-ray scintillators are CsI and Gradolinium oxysulfide for indirect flat-panel TFT detectors and a _______ for direct flat-panel TFT detectors.  
🗑
Exposure latitude   The response of the image receptor to the radiation falling upon it.  
🗑
wider   The exposure latitude for digital detectors is much _____ compared to the film-screen receptors.  
🗑
Flat fielding   A pre-processing technique properly referred to as system calibration that ensures detector performance integrity by removing artifacts.  
🗑
size   The spatial resolution depends on the ____ of the pixels in the matrix.  
🗑
Spatial Resolution   The ability of an imaging system to resolve fine details present in an object. It's also referred to as the sharpness of the image.  
🗑
Modulation Transfer Function   A mathematical function that measures the ability of a detector to transfer its spatial resolution characteristics to the image.  
🗑
one   An MTF of __ represents a perfect detector.  
🗑
decreases   As spatial frequency increases, the MTF _______.  
🗑
one   A perfect digital detector would have a DQE of __.  
🗑
Image lag   also known as memory effect;The persistence of an image in a flat-panel detector due to charge still being produced after the x-ray radiation beam has been turned off  
🗑
dynamic range   The response of a detector to different levels of radiation exposure.  
🗑
Pre-processing   ____________ operations deal with applying corrections to the raw data.  
🗑
Post-processing   _____________ operations address the appearance of the image displayed on a monitor for viewing and interpretation by a radiologist.  
🗑
detector elements   The flat-panel TFT digital detector is designed as a matrix of ______ _______.  
🗑
______ ________ involves the digital subtraction of images in time. A pre-contrast image(mask image) and post contrast images are digitally subtracted from the mask image.   Temporal subtraction  
🗑
The _____ _______ operation is based on subtraction of images taken at different kVps.   energy subtraction  
🗑


   

Review the information in the table. When you are ready to quiz yourself you can hide individual columns or the entire table. Then you can click on the empty cells to reveal the answer. Try to recall what will be displayed before clicking the empty cell.
 
To hide a column, click on the column name.
 
To hide the entire table, click on the "Hide All" button.
 
You may also shuffle the rows of the table by clicking on the "Shuffle" button.
 
Or sort by any of the columns using the down arrow next to any column heading.
If you know all the data on any row, you can temporarily remove it by tapping the trash can to the right of the row.

 
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how
Created by: frenchfrielover
Popular Radiology sets