Save
Upgrade to remove ads
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.
Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't Know
Remaining cards (0)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

RTT 213 - Ch. 18

Ch. 18 - (Egan's) Analysis and Monitoring of Gas Exchange

QuestionAnswer
what is the most common approach to analyzing gas exchange between the blood and tissues? measure O2 levels in the mixed venous (pulmonary artery) blood
what is the term analysis broadly defined as? study or interpretation
in clinical practice, what does laboratory analysis refer too? discrete measurements of fluids or tissue that must be removed from the body
what are these measurements made by? laboratory analyzer
what does monitoring mean? ongoing process by which clinicians obtain and evaluate dynamic physiologic processes in a timely manner
a ________ is a device that provides the important date to the clinician in real time, usually without the removal of samples from the body. monitor
_________ procedures require insertion of a sensor or collection device into the body, whereas ___________ monitoring is a means of gathering data externally. invasive; noninvasive
what does the analysis of gas exchange begin with? knowledge of the system inputs - inspired O2/CO2
what are the two common types of electrochemical O2 analyzers? 1. Clark electrode 2. galvanic fuel cell
what is the clark electrode similar to? those used in blood gas analyzers and transcutaneous monitors
what are the response times for Clark electrode O2 analyzer? galvanic fuel cell? 10-30 seconds; 60 secs
the clark electrode and galvanic fuel cell are suitable for basic _____ monitoring. FiO2
what should be selected when greater accuracy or faster response times are needed? paramagnetic, zirconium cell, Raman scattering, or mass spectroscopy
what must first be done to obtain accurate results with an O2 analyzer? calibration
although procedures differ according to the manufacturer, the basic steps are similar, what are these steps? exposure of the sensor to two gases w/ different O2 concentrations (100% O2/21% room air)
what should be adjusted in order for the analyzer to read 100% O2? calibration
after exposing to 100% O2, what should the clinician do? expose the sensor to room air and confirm a second reading of 21%
when should the clinician know that an analyzer is not working? fails to calibrate or gives an inconsistent reading
what are the most common causes of analyzer malfunction? low batteries (Clark), sensor depletion, electronic failure
what causes inaccurate readings with electrochemical analyzers? condensed water vapor or pressure fluctuations
what type of O2 analyzer is particularly sensitive to condensation? galvanic cells
how can this problem be avoided during continuous use in humidified circuits? place the analyzer sensor proximal to any humidification device
what are fuel cell and Clark electrode readings also affected by? ambient pressure changes
what conditions cause the devices to read lower than the actual O2 concentrations? low pressure (high altitude)
what causes the devices to read higher than the actual FiO2? high pressures (positive pressure ventilation)
what is the definition of respiratory failure still largely based on? blood gas measurements
how can blood gas samples be obtained? percutaneous puncture of a pulmonary artery, catheter (arterial, central venous, or PA), capillary sampling
results obtained from sampling ABG are the cornerstone in the diagnosis and management of wht? oxygenation and acid-base disturbances
what are ABGs considered? gold standard of gas exchange analysis
what does arterial puncture involve? drawing blood from a peripheral artery through a single percutaneous needle puncture
what is the preferred site for arterial blood sampling? radial artery
why is the radial artery most often used? 1. near surface, easy to stabilize 2. collateral circulation 3. no large veins near 4. pain free
what are the indications for ABG? evaluate ventilation, acid-base, oxygenation, ozygen-carrying capacity of blood; assess response to therapy/tests; monitor disease
what are the contraindications of ABG? negative Allen's test results; surgical shunt; no femoral outside of hospital; coagulopathy/anticoagulation (heparin, warfarin)
what are 5 precautions and/or possible complications of ABG? 1. hematoma 2. contamination 3. hemorrhage 4. trauma to the vessel 5. pain
what are examples of assessments for the need of an ABG? history/physical indicators, diagnostic tests, initiation, change, discontinuation of therapy, pulmonary rehabilitation program
what are 5 examples of monitoring during an ABG? 1. FiO2 2. O2 device 3. air bubbles 4. RR 5. temperature
what are the 2 purposes of a needle capping device? 1. isolates from air exposure 2. prevent needle stick injuries
what should be performed before any radial puncture? modified Allen test
when is the modified allen test positive? skin flushes pink within 10 secs
if other conditions make interpreting allen test results difficult, what should be used to assess the pulsatile flow of the thumb? Doppler pulse transducer
in most cases, a sample volume of __-__ mL of blood is adequate. 2-4
what does the actual sample volume needed depend on? 1. anticoagulant 2. requirements of specific analyzer 3. if other tests will be performed on sample
what are the rules for careful handling of the needle that will help avoid transmission of blood-borne diseases? never recap used needle without safety device, never handle it using both hands or point it toward body; never bend, break, or remove used needles from syringe by hand; always dispose of used syringes
what are the first 5 steps for radial artery puncture? 1. check chart 2. confirm steady-state conditions 3. equipment 4. wash hands 5. explain to pt
what are the next 5 steps for radial artery puncture? 6. position pt (extend wrist 30 degrees) 7. Allen test 8. cleanse site (70% alcohol) 9. heparinize the syringe 10. palpate/secure artery with one hand
what are the next 5 steps for radial artery puncture? 11. insert needle, bevel up, 45 degree angle 12. 2-4 mL of blood 13. apply pressure (5-10 mins) 14. expel bubbles, cap syringe 15. mix sample
what are the next 5 steps for radial artery puncture? 16. place sample in transport container 17. dispose waste 18. document 19. check site 20 mins later for hematoma
what are 5 clinical indications for ABG? 1. cyanosis 2. dyspnea 3. abnormal breath sounds 4. tachypnea 5. heavy use of accessory muscle
what are the 2 problem areas associated with arterial puncture? 1. difficulties in getting a good sample 2. preanalytical error
what are the problems with getting a good sample? inaccessible artery, absent pulse, deficient sample return, alteration of test results
when should the tip of the needle be redirected? after it is first withdrawn to the subcutaneous tissue
what can alter the blood gas results? excessive suction
___________ _______ are problems occurring before sample analysis, which can alter the accuracy of the blood gas results. preanalytical errors
how can clinicians avoid most preanalytical errors? ensuring the sample is obtained anaerobically, anticoagulated, analyzed within 15 mins
what is the traditional method used to avoid errors caused by blood cell metabolism? quickly chill the sample by placing it in ice slush
when is this needed? if the sample is not analyzed within 10-15 mins
when should chilled samples be disposed? after 60 mins
how long do pts with healthy lungs take to achieve a steady state? COPD pts? 5 mins after changes; as long as 20-30 mins
to document the pt's status, what should be recorded? 1. date, time, site of sampling 2. allen test results 3. body temp, position, activity level, RR 4. FiO2
what is the first step of interpretation of the results? ensure you're looking at the results of the correct pt
what are the 2 basic steps of interpretation of the results? 1. interpretation of the oxygenation status 2. interpretation of the acid-base status
what is the oxygenation status determined by examining? PaO2, SaO2, CaO2
what does SaO2 represent? the degree to which the hemoglobin is saturated with O2
what does CaO2 represent? the content of O2 in 100mL of arterial blood and is a function of SaO2
what is the normal SaO2? CaO2? 95%-100%; 18-20 mL of O2 per 100mL of arterial blood (16-20 vol%)
what is the acid-base status of the pt determined by examining? pH, PaCO2, HCO3
what do indwelling catheters provide? continuous monitoring of vascular pressures
whats more likely to happen with indwelling catheters than they are with intermittent punctures? infections and thrombosis
what are the most common routes for indwelling vascular lines? peripheral artery, central vein, pulmonary artery
in an indwelling catheter, what does the catheter connect to? a disposable continuous-flush device
how does this device keep the line open? by providing a continuous low flow (2-4 mL/hr) of heparinized IV fluid
what is connected to the flush device that provides an electrical signal to an amplifier or monitor, which displays the corresponding pressure waveform? strain-gauge pressure transducer
what provides access for sampling blood from most intravascular lines? three-way stopcock
what is the first step in the procedure of an indwelling catheter? ensure that the balloon is deflated and prepare to draw the sample directly from the catheter's distal port
what is the second step in the procedure of an indwelling catheter? clinician slowly withdraws the sample
what can dilute the blood sample and affect O2 content measures? rapid flow of IV fluid
what is capillary blood gas sampling sometimes used as? an alternative to direct arterial access in infants/small children
what can properly obtained capillary blood from a well-perfused pt provide rough estimates of? arterial pH and PaCO2 levels
the capillary PO2 is of no value in estimating _________ __________. arterial oxygenation
capillary blood values are meaningful only if what? the sample site is properly warmed
what is the most common technical error is capillary sampling? inadequate warming and squeezing of the puncture site
a sample obtained from a warmed capillary site is often referred to as ___________ ______. arterialized blood
what part of the body do capillary blood samples reflect arterial PCO2 and PO2 better than a finger stick? earlobe
what is the equipment required for a capillary blood sampling? lancet, preheparinized glass capillary tubes, metal "fleas," magnet, clay/wax sealant/caps, gauze, bandages, ice, gloves, skin antiseptic, warming pads, sharps container, labeling
what is the most common site of capillary sampling? heel, specifically the lateral aspect of the plantar surface
what do analyzers use the measurements of pH, PaCO2, and PO2 to compute? plasma bicarbonate, base excess, hemoglobin saturation
if actual measurement of Hb saturation, methemoglobinm and HbCO is required, the sample usually must be analyzed separately using ___________. hemoximetry
what are the key components that blood gas analyzers typically share? 1. operator interface 2. measuring chamber 3. calibrating gas tanks 4. reagent containers 5. waste container 6. results display, storage, transmittal system
to measure PO2, blood gas analyzers use...? the Clark polaragraphic electrode
what does the pH electrode actually consist of? two electrodes or half-cells
what is used to measure PaCO2? Severinghaus electrode
to provide accurate and clinically useful data, how must blood gas analysis be performed? 1. on a sample free of preanalytical errors 2. properly functioning analyzer 3. procedure that follows manufacturer's recommendations
what does quality pt care depend on? acurate blood gas results
what does the accuracy of blood gas testing depend on? rigorous quality control
what is the hallmark of a comprehensive quality control program? RECORDKEEPING and clearly written and comprehensive policies and procedures
__________ _________ is the process of testing a new instrument to confirm a manufacturer's claims. PERFORMANCE VALIDATION
_________ is examining the repeatability of the results. precision
what is the best way to avoid problems associated with analyzers, filters, membranes, or electrolyte solution deteriorating and failing? scheduling regular PREVENTIVE MAINTENCE
_________ is the only fully automated element of blood gas quality control. CALIBRATION
in most units, the media used to calibrate the gas electrodes are...? precision mixtures of O2 and CO2.
what is used for the pH electrode? standard pH buffer solutions
___________ ______ must meet the requirements set by nationally recognized standards organizations. calibration media
what is calibration performed to ensure? that the output of the analyzer is both accurate and linear across the range of measured values
what are the 2 steps of calibration? 1. adjusting the offset of the instrument so that low output = low input 2. the gain (slope) is adjusted to ensure that high output = high input
_____________ ___________ establishes and periodically confirms the validity of blood gas analyzer results. CALIBRATION VERIFICATION
what does calibration verification require? analysis of at least 3 materials with known values spanning the entire range of values expected for clinical samples ("controls")
_______ _______ _______ takes calibration verification a step further by applying statistical and rule-based procedures to help detect, repond to, and correct instrument error. INTERNAL QUALITY CONTROL
what are the 2 categories of analytical error? 1. random 2. systematic
random errors are errors of precision, or more precisely, __________. imprecision
what does bias plus impercision equal? total instrument error, or inaccuracy
_________ ________ requires analysis and reporting on externally provided control media with unknown values, usually three times per year, with five samples per test. PROFICIENCY TESTING
________ _______ is the ongoing process of applying appropriate measures to correct errors identified through the quality assurance cycle. REMEDIAL ACTION
____________ ______ takes blood gas analysis from the specialized laboratory to the pt's bedside. point-of-care testing
besides blood gas analysis, what can other devices measure during point-of-care testing? serum electrolytes, blood glucose levels, blood urea nitrogen, hematocrit, prothrombin, partial thromboplastin time
where does analysis take place? a disposable cartridge, which is inserted into a chamber in the body of the unit
a ______ ____ _______ is a bedside tool that can provide measurements either continuously or at appropriate intervals without permently removing blood from the pt. blood gas monitor
what are the 3 such systems in current clinical use? 1. transcutaneous blood gas monitor 2. intraarterial (in vivo) 3. on-demand (ex vivo)
_____________ blood gas monitoring provides continuous, noninvasive estimates of arterial PO2 and PCO2 through a surface skin sensor. transcutaneous
what is the comparison of transcutaneous blood gas monitoring and capillary sampling? the device arterializes the underlying blood by heating the skin
what are the two most important factors that influence the agreement between arterial blood and transcutaneous gas measurements? age and perfusion status
in terms of age, the _________ the pt, the better is the agreement between the PaO2 and PTCO2. younger
in perfusion status, when are the PaO2 and PTCO2 similar? in pts with normal cardiac output and fluid balance
agreement between PTCO2 and ______ is a little better. PaCO2
when is transcutaneous monitoring a reasonable choice? when there is a need for continuous, noninvasive, analysis of gas exchange in hemodynamically stable infants/children
transcutaneous monitoring also is useful for monitoring _________ in neonates. hyperoxia
instead of measuring gas tensions in a blood sample, transcutaneous electrodes measure PO2 and PCO2 in...? an electrolyte gel between the sensor and skin
what is the response time for these electrodes? 20-30 secs
what are the most common sites for electrode placement? abdomen, chest, lower back
what are the steps to take care to avoid thermal injury to the pt's skin? 1. careful monitoring of sensor temp (42 degrees C) 2. regularly rotating the sensor site
proper _____-________ contact is essential, as is proper application to the skin surface. sensor-electrolyte
with intraarterial (in vivo), rather than using electrochemistry, ______ measure blood gas parameters by photochemical reactions, which alter light transmission through optical fibers. optodes
light transmitted to this dye can be absorbed, reflected, or even re-emitted at a different wavelength called ___________. fluorescence
most photochemical blood gas systems used both ___________ and __________-based optodes. absorbance; fluorescence
because O2 "quenches" the dye's fluorescence, the intensity of this return signal is inversely proportional to the arterial _____. PO2
________ (__ ____) blood gas monitoring systems are a logical compromise between bench-top and in vivo blood gas analysis. on-demand (ex vivo)
what is the only major shortcoming of ex vivo systems? their inability to provide real-time continuous data
_________ is the measurement of blood hemoglobin saturations using ______________. oximetry; spectrophotometry
a substance's pattern of light absorption varies predictably with the amount of pressure; what law is this known as? Lambert-Beer law
the particular pattern of light absorption exhibited by a substance at different wavelengths is called its _________ ________. absorption spectrum
what are the two types of oximetry used in clinical practice? 1. hemoximetry (co-oximetry) 2. pulse oximetry
what is hemoximetry? laboratory analytical procedure requiring invasive sampling of arterial blood
how are the specific wavelengths needed for analysis yielded? light generated by a thallium cathode lamp passes through a series of lenses and filters
s beam splitter then divides the light into two portions, directing one through a reference solution and the other through a sample chamber, or _______. cuvette
because a laboratory hemoximeter used three different wavelengths of light, it can simultaneously compute the relative concentrations of multiple hemoglobin species, such as...? Hb, HbO2, HbCO, metHb
what is the first step in hemoximetry? blood is introduced into the sampling port of the analyzer, usually either by aspiration or injection
what is the next step? hemolysis (then, to cuvette for analysis)
what is a major assumption underlying hemoximetry? the measured changes in light absorbance result only from variations in the relative concentrations of various hemoglobins
a ______ ________ is an inexpensive and portable noninvasive monitoring device that provides estimates of arterial blood oxyhemoglobin saturation levels. pulse oximeter
the pulse oximeter combines the principle of spectrophotometry, as used by hemoximetry, with ______________. photoplethysmography
what does the pulse oximeter use? two wavelengths of light (one red, one infrared)
pulse oximeter actually measures transmission through...? living tissue, such as a finger or an earlobe
what does a baseline component represent? the stable absorbance pf the tissue bed, which mainly is the result of venous and capillary blood
the ______ the actual SaO2, the less accurate and reliable is the SpO2 measurement. lower
most clinicians consider pulse oximeter readings unreliable at saturations below ___%. 80
what are the two problem categories with pulse oximetry? 1. those inherent in the technology itself 2. those associated with clinical interpretation and use of data
what is the most common source of error and false alarms? motion artifact
what can minimize this problem? relocation of the sensor
how are the low alarms set for dark skin pigmentation? 3% - 5% higher
the pulse oximeter does not measure ____. PCO2
___________ is the measurement of CO2 in respiratory gases. capnometry
a __________ is the device that measures the CO2. capnometer
___________ is the graphic display CO2 levels as they change during breathing. capnography
what is the primary use of capnography? monitoring during general anesthesia or MV
what is the key component in a capnograph? rapidly responding CO2 analyzer
what is the most common rapid CO2 analyzer? infrared capnometer
what are the two different methods to sample respiratory gases that capnometers use? 1. mainstream sampling 2. sidestream sampling
what can interpretation of the capnogram be useful in assessing trends in? alveolar ventilation and detecting V/Q imbalance
what is gas sampled at the end of exhalation? end-tidal gas
what is the normal PETCO2 1-5 mmHg less than the PaCO2 or b/t 35-43 mmHg
what is the first step in assessing the capnogram? determine the actual PETCO2 and whether it has changed over time
what does a PETCO2 of zero usually indicate? system leak, esophageal intubation, cardiac arrest
once the capnogram as been assessed for changes in PETCO2, the _________ and its pattern should be analyzed. waveform
what does a normal capnogram have? starts with a sharp upstroke, followed by a plateau, then a rapid downstroke
what does an elevated baseline indicate? rebreathing
what is the most significant error with capnogram? assuming that the end-expired CO2 levels can substitute for actual PaCO2 measurements
what is the most common problem? contamination or obstruction of the sampling system or monitor by secretions or condensate
Created by: christa_2008
 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards