| mag mode places the focal point ____, while increasing voltage to the ___. | farther from the output phosphor; focusing lenses |
| mag mode increases size ____ | 1.5-4x |
| magnification formula= | normal input diameter/input diameter during magnification |
| FPD keep the same resolution except if ____ is used. | binning |
| to give the smallest pt dose, we use | largest fov |
| pros and cons of mag mode | pros: greater detail and spatial resolution. cons: higher pt dose. dimmer image |
| intensifier artifact S-distortion is caused by | external magnetic field. equipment or earth's magnetic field. |
| intensifier artifact pincushion distortion is caused by | the input is curved |
| QC testing is done by | medical physicists |
| the reproducibility variance btwn exposures must be less than | 5%/0.05 |
| blooming in fluoroscopy | when bright areas appear larger. bone, metal. means too much radiation is hitting that area |