Save
Upgrade to remove ads
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.
Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't Know
Remaining cards (0)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

X Ray Tube 147

Part 2

QuestionAnswer
Rotor- Ferromagnetic metal cylinder (Copper & iron) attached to target
Stator- Motor with electromagnetic windings
The _______ Activates a magnetic field when the prep button (rotor switch) is depressed Stator
The _______ Rotates to align with changing magnetic poles of the stator when prep button depressed Rotor
What is the Area where electrons strike the anode and X-Rays are produced? Focal Spot
_______ is used in Dental Radiography Stationary Target
_________ have a Greater Area exposed to electrons and Withstand Higher Heat Loads Rotating Targets
Large Focal Spot- Large or Higher exposures
Small Focal Spot- Better image detail (resolution)
Physical area of the focal track that is impacted Size projected is the same size as the filament selected Actual Focal Spot
Measured directly under anode; Size projected on pt and IR Stated focal spot size Effective Focal Spot
________ (5-20 degrees) Determines effective focal spot size Anode Angle
The Anode Angle Allows the effective focal spot to be ______ than the actual focal spot Smaller
________ Provide better resolution and Endures great heat loads Smaller Effective Focal Spots
Radiation intensity is Less on the Anode side because x-rays are _______ Attenuated
Increase in focal spot size caused by electrostatic repulsion Blooming
X-rays produced away from focal spot (X-rays go outside of EFS size and expose unnecessary areas of the patient) Off-Focus Radiation
Primary interaction of x-ray production Heat
What is the purpose of the X-Ray tube warm up procedures? Extends the life of the x-ray tube by preventing damage such as anode cracking
X-Ray tube warm up procedures are performed at the beginning of the day and/or when the tube has not been used in_______ Two Hours
Typical X-Ray Tube warm-up procedure- – Two exposures – Technique: 70 kVp; low mA; 0.5 – 2 seconds
Radiographic Tube Rating Charts Purpose- Determines the maximum technical factor combination allowed without overloading the tube
Radiographic Tube Rating Chart X-Axis- Time (in seconds)
Radiographic Tube Rating Chart Y-Axis- kVp
Radiographic Tube Rating Chart Curve- mA station
Any technique _______is safe Under The Curve
Anode & Housing Cooling Curves Used to determine the amount of time needed for the anode or housing to completely cool down May also be used to determine how much cooling time is needed prior to making additional exposures
Anode & Housing Cooling Curves X-Axis Time (in minutes)
Anode & Housing Cooling Curves Y-Axis Heat Units (as calculated)
Heat Units- Used to measure the amount of heat generated by specific technique factors
Single Phase Formulae: mA × time × kVp
Three Phase, Six Pulse Formulae: mA × time × kVp × 1.35
Three Phase, Twelve Pulse Formulae: mA × time × kVp × 1.41
High Frequency Formulae: Is the same as Three Phase, Twelve Pulse (Used to be x 1.45)
Created by: user-2016295
 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards