X Ray Tube 147
Part 2
| Question | Answer |
|---|---|
| Rotor- | Ferromagnetic metal cylinder (Copper & iron) attached to target |
| Stator- | Motor with electromagnetic windings |
| The _______ Activates a magnetic field when the prep button (rotor switch) is depressed | Stator |
| The _______ Rotates to align with changing magnetic poles of the stator when prep button depressed | Rotor |
| What is the Area where electrons strike the anode and X-Rays are produced? | Focal Spot |
| _______ is used in Dental Radiography | Stationary Target |
| _________ have a Greater Area exposed to electrons and Withstand Higher Heat Loads | Rotating Targets |
| Large Focal Spot- | Large or Higher exposures |
| Small Focal Spot- | Better image detail (resolution) |
| Physical area of the focal track that is impacted Size projected is the same size as the filament selected | Actual Focal Spot |
| Measured directly under anode; Size projected on pt and IR Stated focal spot size | Effective Focal Spot |
| ________ (5-20 degrees) Determines effective focal spot size | Anode Angle |
| The Anode Angle Allows the effective focal spot to be ______ than the actual focal spot | Smaller |
| ________ Provide better resolution and Endures great heat loads | Smaller Effective Focal Spots |
| Radiation intensity is Less on the Anode side because x-rays are _______ | Attenuated |
| Increase in focal spot size caused by electrostatic repulsion | Blooming |
| X-rays produced away from focal spot (X-rays go outside of EFS size and expose unnecessary areas of the patient) | Off-Focus Radiation |
| Primary interaction of x-ray production | Heat |
| What is the purpose of the X-Ray tube warm up procedures? | Extends the life of the x-ray tube by preventing damage such as anode cracking |
| X-Ray tube warm up procedures are performed at the beginning of the day and/or when the tube has not been used in_______ | Two Hours |
| Typical X-Ray Tube warm-up procedure- | – Two exposures – Technique: 70 kVp; low mA; 0.5 – 2 seconds |
| Radiographic Tube Rating Charts Purpose- | Determines the maximum technical factor combination allowed without overloading the tube |
| Radiographic Tube Rating Chart X-Axis- | Time (in seconds) |
| Radiographic Tube Rating Chart Y-Axis- | kVp |
| Radiographic Tube Rating Chart Curve- | mA station |
| Any technique _______is safe | Under The Curve |
| Anode & Housing Cooling Curves | Used to determine the amount of time needed for the anode or housing to completely cool down May also be used to determine how much cooling time is needed prior to making additional exposures |
| Anode & Housing Cooling Curves X-Axis | Time (in minutes) |
| Anode & Housing Cooling Curves Y-Axis | Heat Units (as calculated) |
| Heat Units- | Used to measure the amount of heat generated by specific technique factors |
| Single Phase Formulae: | mA × time × kVp |
| Three Phase, Six Pulse Formulae: | mA × time × kVp × 1.35 |
| Three Phase, Twelve Pulse Formulae: | mA × time × kVp × 1.41 |
| High Frequency Formulae: | Is the same as Three Phase, Twelve Pulse (Used to be x 1.45) |