Generator 156
Part 2
| Question | Answer |
|---|---|
| The higher mA station number, the ________ the resistance | Lower |
| _________ Modified power is required to heat tube filament | Modified Power |
| The Step-Down Transformer _______ voltage and adjusts amperage | Decreases |
| Primary Coil of Step-Down Transformer = | Approximately 150 V & 1 A |
| Secondary Coil of Step-Down Transformer = | 12 volts and 5-8 A |
| ________ Determines the temperature of the filament and the number of electrons produced | The Step-Down Transformer |
| ________ correspond to the dual sized filaments | Small and Large focal spots |
| Filaments are Thin tungsten wires located on the _______ side of the x-ray tube | Cathode |
| _________ results when filaments are heated | Thermionic Emission |
| _________ Receive properly modified current from the step-down transformer | Filaments |
| Filaments operate at currents of | 3 to 6 A |
| The ______ is located in the low voltage section | Timer Circuit |
| The Timer Circuit controls exposure time by | Terminating Exposure |
| Mechanical Timers- | A preset exposure time is set by turning a knob that winds a spring. When the exposure button is depressed, the spring is released |
| Synchronous Timers- | Driven by a synchronous motor which operates at 60 revolutions per second |
| Electronic Timers- | The most expensive, most complicated and most accurate of all the types of x-ray timers |
| mAs Timers- | special kind of electronic timer • Provides the highest safe mA for the shortest exposure time |
| Automatic Exposure Control- | Terminates exposure when required amount of radiation has reached the image receptor |
| Automatic Exposure Control Determines the ______ of radiation reaching IR | Quantity |
| What timers are useful in Angiography? | Electronic Timers |
| What timers are used mostly in older dental units? | Mechanical Timers |
| Rectification occurs after _______ in an x-ray circuit | The step-up transformer |
| Rectification occurs before ______ in an x-ray circuit | The x-ray tube |
| AC is changed to______ in a Rectification Circuit | Pulsating Direct Current (DC) |
| Electric Current is _____ the direction of the arrow of a diode symbol | With |
| Electron Flow is _____ the direction of the arrow of a diode symbol | Against |
| Ripple- | Decrease in voltage because the potential drops to zero with every change in direction |
| Ripple results in a | Pulsating X-Ray Beam |
| Generators & Rectifiers- | Are used to manipulate the waveform in order to reduce ripple |