RAD 107 EXAM 3
| Question | Answer |
|---|---|
| The ____________ in the image intensifier emits electrons when it is stimulated by light photons. | photocathode |
| With image intensification the light level is raised to ___________ vision. | photopic |
| The number of light photons emitted within the image intensifier is ____________ to the amount of x-ray photons exiting the patient. | directly proportional |
| The input phosphor converts _________ to _________. | xrays to light |
| Image intensified fluoroscopy is performed at illumination levels similar to ________. | radiograph viewing |
| Fluoroscopy was developed so that radiologists could view ______________ images. | dynamic |
| The capability of an image intensifier to increase the illumination level of the image is called its _____________________. | brightness gain |
| The ratio of x-rays incident on the input phosphor to light photons exiting the output phosphor is called _______ gain. | flux |
| Viewing the fluoroscopic image in magnification mode increases _____________. | contrast resolution and patient dose |
| The time-interval delay mode is principally used in ___________ imaging. | cardiac |
| In mask mode, the ________ image is shown on monitor B, and the ________ image is shown on monitor A. | mask and subtracted |
| Fluoroscopy for an air contrast barium enema is generally done at _____ kVp. | 80-90 |
| Brightness gain is typically in the range of ____________. | 5000-30,000 |
| Digital fluoroscopy energy subtraction has less ___________ than temporal subtraction. | motion artifact |
| Automatic brightness control (ABC) maintains the brightness of the image by varying ___________________. | kVp and mA |
| The combination of temporal and energy subtraction techniques is called _________. | hybrid subtraction |