RAD107 final review
| Question | Answer |
|---|---|
| The _______ in the retina are stimulated by _______ light; the _______ are stimulated by _______ light. | rods-low cones-bright |
| The milliamperage used during fluoroscopy is operated at less than? | less than 5 mA |
| Image intensified fluoroscopy is performed at illumination levels similar to ________. | radiograph viewing |
| The ability of the eye to detect differences in brightness levels is termed ___________. | contrast perception |
| Electrons hit the _______________ after exiting the anode. | output phosphor |
| The ratio of x-rays incident on the input phosphor to light photons exiting the output phosphor is called _______ gain. | flux |
| Light produced at the output phosphor of the image intensifier has been increased __________ times in intensity. | 50 to 75 |
| With image intensification the light level is raised to ___________ vision. | photopic |
| The image intensifier improved fluoroscopy by increasing image _________________. | brightness |
| The _______________ is the product of the minification gain and the flux gain. | brightness gain |
| A charge-coupled device used in digital fluoroscopy provides high _________________. | spatial resolution, signal to noise ratio, and detective quantum efficiency |
| Automatic brightness control (ABC) maintains the brightness of the image by varying ___________________. | kVp and mA |
| Energy subtraction technique takes advantage of the difference in ______________ during contrast injection. | K-edge absorption |
| Brightness gain is typically in the range of ____________. | 5000 to 30,000 |
| Fluoroscopy for an air contrast barium enema is generally done at _____ kVp. | 80-90 |
| The principal drawback of the first-generation CT scanner was the _______________. | long scanning time |
| Many CT x-ray tubes have the capacity for millions of _________ with each exam. | heat units |
| CT scanners operate on ________________ voltage generation. | high-frequency |
| Each CT projection records variations in _________________________. | mass density and effective atomic number |
| The time from the end of CT imaging to image appearance is called the ______ time. | reconstruction |
| The concentration of scintillation detectors affects the ___________ of the CT image. | spatial resolution |
| The second-generation CT scanners had the disadvantage of increased ____________. | beam scatter |
| Each pixel of information in the CT image contains numerical information in ___________ units. | Hounsfield |
| Each CT projection is ______________________. | stored on the computer |
| The relationship between patient couch movement and the x-ray beam is known as ______________. | pitch |
| CT imaging has excellent ______________ because of the narrow beam collimation. | contrast resolution |
| A CT image is a ___________ or __________ image that is perpendicular to the long axis of the body. | transaxial, transverse |
| The x-ray source and the detector move ______________ for computed tomography. | synchronously |
| The CT scanner must be calibrated so that ____________ is at CT number zero. | water |
| A high variation in pixel values in a homogeneous phantom is a measure of ________. | image noise |
| The power rating for an interventional radiography tube should be at least _____ kW. | 100 |
| The x-ray tube used for interventional radiologic procedures has a _____________. | small target angle |
| The use of ______________ reduces the risk of a drug reaction during angiographic procedures. | nonionic contrast |
| The first digital radiographic tomosynthesis (DRT) appeared in which modality? | mammography |
| An advantage of DRT is to _______________. | reduce or eliminate tissue superimposition |
| What is the typical source to center of rotation distance in DRT? | 60-100cm |
| During digital radiographic tomosynthesis (DRT), what is the range of the sweep of the x-ray tube? | +/– 5 to +/– 25 degrees |