click below
click below
Normal Size Small Size show me how
Study Stack Exam Rev
Study Stack Exam Review
| Question | Answer |
|---|---|
| An artifact | any false visual feature on a medical image that simulates tissue or obscures tissue |
| Preprocessing | Interpolation is the process of assigning a value to a dead pixel based on the recorded values of adjacent pixels Flatfielding is a preprocessing software correction that is performed to equalize the response of each pixel to a uniform x- ray beam. |
| Image compression | Lossless compression up to 3:1 generally is considered acceptable and helpful in digital radiographic image management. Lossy compression is greater than 10:1 compressio |
| Image histogram | Graph of frequency of occurrence versus digital value intervals |
| Collimation and partition | Automatic radiation field recognition is essential for artifact-free images. Proper collimation and centering prevent histogram errors that can lead to artifacts. |
| X-rays that exit from the patient are remnant x-rays and those that exit and interact with the image receptor are called | image-forming x-rays |
| Collimation | reduces patient radiation dose and improves image contrast |
| Patient Thickness | Thicker parts produce more scatter than thin parts Compression of anatomy improves spatial resolution and image contrast and lowers the patient radiation dos |
| Effect of scatter radiation on image contrast | Reduced image contrast results from Compton scattered x-rays. |
| Beam restrictors | Aperture diaphragm Cones and cylinders Variable-aperture collimator Collimation reduces the patient radiation dose and improves image contrast. Positive-beam-limiting (PBL) devices |
| Grid ratio | h/D |
| Grid frequency | 10,000 μm/cm/(T+D)μm/line pair |
| Contrast improvement factor | image contrast with a grid/image contrast without a grid The contrast improvement factor is higher for high-ratio grids. |
| Bucky factor | incident remnant x-rays/transmitted image-forming x-rays As the Bucky factor increases, radiographic technique and patient radiation dose increase proportionately. |
| Grid Types | Parallel Grid cutoff Distance to Grid Cutoff = SID/Grid Ratio Crossed Grid cutoff Focused High-ratio grids have less positioning latitude than low- ratio grids. Moving Grid lines Virtual |
| Grid Problems | Off-level grid Off-center grid Off-focus grid Upside-down grid |
| Grid selection factors | Patient radiation dose increases with increasing grid ratio. High-ratio grids are used for high-kVp examinations. The patient radiation dose at high kVp is less than that at low kVp. |
| Air-gap technique | Disadvantage of the air-gap technique is image magnification with associated focal-spot b |