Save
Upgrade to remove ads
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.
Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't Know
Remaining cards (0)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

RAD107_Week 5-8

QuestionAnswer
An increase in mAs causes ______________ in beam quality and ___________ in beam intensity. no change; an increase
Patient dose increases are directly related to increases in mAs
Both beam quality and beam quantity are increased by ______________ increasing kVp and using a 3-phase generator
There is a direct relationship between the quantity of x-rays and the ___________ milliamperage
the 5% rule An increase of 5% in kVp may be accompanied by a 30% reduction in mAs to produce the same image receptor response at a slightly reduced contrast scale =
The small focal spot will provide a _______ finer detail of image
What are the prime exposure factors? kVp, mAs, and SID
Beam penetrability is increased if ________ is/are increased. kVp
Patient thickness should be measured with the ______ caliper
Added filtration has the effect of __________ the beam quality and ________ patient dose. increasing; reducing
Foreshortening is caused by a(n) ____________________ angled body part when the beam is perpendicular to the image receptor
Changes in kVp do not affect the number of number of electrons boiled off the filament
When only the intensity needs to be changed, only the _______ should be adjusted mAs
Beam quantity is improved when the __________ is increased, but __________ has no effect on beam quantity. kilovoltage; milliamperage
Sharpness of detail can be improved by increasing _______ SID
If mAs is increased from 20 mAs to 40 mAs, the patient dose will _______ e doubled
With automatic exposure techniques the exposure is terminated when the optimum _____ is reached mAs
List the following tissues in order from the most radiolucent to the most radiopaque. Lung, fat, muscle, bone
At least a _____% change in mAs is necessary to produce a visible change in the image 30
Radiographic image quality is improved when the __________ is increased source image distance
The visibility of image detail is best measured by_____ Contrast resolution
A high quality radiographic image should always demonstrate ________ Structures and Tissues
What is the formula for the magnification factor? MF = SID/SOD
Distortion can be reduced by _________________ placing the object plane perpendicular to the image plane
Which of these body parts has the highest subject contrast? Chest
Sharpness of image detail is best measured by _____________ spatial resolution
Patient thickness affects radiographic technique by affecting _________ mAs and kVp settings
Spatial resolution is limited to ____________ pixel size
Which body habitus indicates an average patient? Sthenic
Radiographic technique charts embedded in the operating console are based on hypersthenic patients (T/F) False
The best way to minimize magnification is to use a __________. long SID and small OID
Subject contrast is affected by ________ Patient Thickness
The technologist primarily controls radiographic contrast by varying the ________ kilovoltage
_________ is defined as the ability to image two separate objects and visually detect one from the other. Resolution
Patient factors include ___________ anatomical thickness and body composition
_______of the image is when the anatomical part appears smaller than normal. Foreshortening
Which body habitus indicates a thin, but healthy appearing patient? Hyposthenic
The ability to image small objects that have high subject contrast is known as ______ spatial resolution
Which group of factors includes spatial resolution and distortion? Image-quality factors
____may be described by identifying three groups of factors: patient factors, image-quality factors, and exposure technique factors Radiographic technique
Which group of factors includes grids, focal-spot size, and filtration? Exposure technique factors
The ability to distinguish between and image similar tissues is known as ______ contrast resolution
To obtain a magnified radiograph, the _______ is increased while the ______ is held constan OID; SID
Distortion is reduced by positioning the anatomical part of interest in a plane _________ to that of the image receptor. Parallel
Which group of factors includes kVp, mAs, and SID? Exposure technique factors
Which group of factors includes anatomical thickness and body composition? Patient factors
The purpose of image subtraction is to enhance Contrast
What is the basic photometric unit? Lumen
_______ improves with the use of higher megapixel digital display devices. Spatial resolution
Reregistering an image to correct for patient motion is done with ____________ pixel shift
A liquid crystal is a material in a ______________________ state. state between a liquid and a solid
Preprocessing the digital image is done _____________, and post-processing is done __________ automatically; manually
The SI unit for luminance is the nit
Almost all digital images in medical imaging are viewed and interpreted on a __________ digital display device
Signal interpolation is performed during preprocessing
Almost all medical flat panel digital display devices are _____________ monochrome liquid crystal displays (LCDs)
Medical flat panel display devices are identified by the _____________ number of pixels
Annotation is performed during post-processing
A digital display monitor is best viewed ________ straight on
The cosine law is used to describe ________________ luminous intensity of a digital display device
__________________ is corrected by the application of an offset voltage Image lag
Which medical information organization is working to make patient radiation dose ALARA? ACR
File size is the matrix multiplied by the _______ byte depth
______ implements and speeds image processing, viewing, interpretation, storage, and recall PACS
Which three organizations created the document, “Technical Standard for Electronic Practice of Medical Imaging”? SIIM, ACR, AAPM
The national standard for image transmission in teleradiology is the ____________ format. DICOM
What does DICOM stand for? Digital Imaging and Communications in Medicine
PACS stands for ________________ Picture Archiving and Communication Systems
The _________ deals with schedules, protocol descriptors, diagnostic conclusions, and billing. Radiology Inforamtion Systems (RIS)
The process of remote transmission and viewing of medical images is known as __________. Teleradiology
What does EMR stand for Electronic Medical Record
What system became the standard for PACS in order to receive/read images? DICOM and HL7
The term network is used to describe the ___________ manner many computers can be connected
Text data and email are generated at a _____________________________ workstation in PACS. Radiology Information System (RIS)
The process of matching any worker to the work environment in order to maximize efficiency is Ergonomics
When light arrives at the retina, it is detected by the _____________ rods and cones
When the ROC analysis is completed, the data is evaluated using the metric, _________ area under the curve (AUC)
Digital radiography is superior to analog radiography because of ______________ Image brightness
________ is a term related to changes we make voluntarily and involuntarily in order to improve our image perception. Visual accommodation
Light incident on the eye must first pass through the ____________ cornea
The special demands of digital imaging are ________________ and _____________. postprocessing; numerical analysis
The ambient light levels in the reading area should be ______________. near darkness
Photopic vision is used for _____________ daylight vision
Receiver operating characteristic curves is a decision matrix used for __________ decision-making by the radiologist
The decreasing luminous intensity with increasing distance from the source of light follows _______________ the inverse square law
All of the following are components of receiver operating characteristic curves, except ______ sensitivity index
Image perception is a scientific term for what we call visual _________ sensitivity
The Task Group Report 18 is a procedure manual for _______________ testing digital display devices
A telescope photometer is used at __________. 1 meter
Most of the quality control testing of digital display devices is performed with the use of _______________ test patterns. AAPM TG18
Daily quality control testing is done by the radiologic technologist
SMPTE stands for the _____________________ Society of Motion Picture and Television Engineers
Digital display reflections and ambient light conditions are assessed with a(n) ___________ illuminance meter
There are _______ types of photometers commonly used. two
A standard pattern made to measure the resolution of digital display systems is called the ______________ SMPTE pattern
The TG18-LPV/LPH test is used to evaluate ______________ geometric distortion
Medical images are transferred using the ____________ standard. DICOM
The quality control standard for flat panel displays was released by _______ in 1998. VESA
The qualitative evaluation of luminance response is measured with the ________ test tool. TG18-CT
Specular reflection is reduced with the use of ___________ improved antireflective coating
The consistent appearance of DICOM images on any compatible digital display device is achieved through _____________ perceptual linearization
Which of the following is the principal contributor to image noise in radiographic and fluoroscopic imaging procedures? Quantum mottle
Radiographic __________ is random fluctuation of x-ray interaction on the image receptor. noise
Patient thickness affects image quality by affecting ____________. magnification, radiographic contrast and focal spot blur
The best way to minimize magnification is to use a __________. long SID and small OID
Focal-spot blur is small on ______________ side and large on the ___________ side of the image receptor anode; cathode
_________ is defined as the ability to distinguish anatomical structures of similar subject contrast and visually detect one from the other. Contrast resolution
What is the formula for Radiographic contrast? Image receptor contrast × subject contrast
Distortion depends on _______ object shape
Spatial resolution improves with decreased focal spot size, motion and pixel size.
A high-quality medical image should always demonstrate ___________ structures and tissues
The technologist primarily controls radiographic contrast by varying the __________. kilovoltage
Higher speed image receptors generally produce images with ____________________. increased noise
Which of the following will reduce quantum mottle? High mAs, low kVp, and slower image receptors
The three primary geometric factors affecting image quality are ___________ magnification, distortion, and focal spot blur
Distortion depends on ______ position object
What is the formula for the magnification factor? MF = SID/SOD
Focal spot blur can be reduced by using __________. a small focal spot
Radiographic image quality is improved when the __________ is increased. source image distance
Magnification factor = image size/object size
The x-rays that are transmitted through the patient without interaction contribute to _________ useful information
Photoelectric interactions increase when ___________ is decreased. kVp
Lowering kVp ________ patient dose and _________ image contrast. increases; increases
Federal regulations passed in _____ required that positive beam limiting devices be on all manufactured x-ray collimators. 1974
The positive beam limiting device (PBL) assures that the x-ray beam is collimated to _____________ the image receptor size
Which kVp selection would result in the most scattered x-rays in the image-forming beam? 90kvp
Approximately _______% of the x-ray beam incident on the patient is transmitted through the patient without interaction to become part of the image-forming beam. 1%
Compton scatter contributes to __________ image noise
Contrast resolution is improved by _____________. tight collimation, lowering kVp and patient compression
The use of a compression device will increase __________. image contrast
The three primary factors influencing the intensity of scatter in the image-forming beam are ___________________. kVp, field size, and patient thickness
A fixed aperture diaphragm should leave a _____ cm unexposed border on the film. 1cm
The use of __________ improves contrast and reduces patient dose. collimation
Beam restriction with a(n) ____________ is only accurate at a fixed SID. aperture diaphragm
The most commonly used beam restricting device is the ____________. variable collimator
How can you improve image contrast with a heavy patient without increasing patient dose? Use tight collimation.
Decreasing the kVp will increase ___________. patient dose
The x-ray interaction that contributes to the clear parts of the image is _____________. photoelectric interaction
Image-forming x-rays include those which have been _________________ transmitted without interaction and scattered through Compton interaction
Scatter radiation increases as ___________________ increases. field size
Computed radiography cassettes can sit for ____________ between exposures. a few days
Histogram selection is done _________________ by the radiographer before each exam
The size and number of digital images have to be compressed to permit _______ transmitting and archiving
Lossy compression reduces data files greater than _____ 10:1
During preprocessing, the digital output of the image receptor may need to be manipulated to correct for ______________ dead pixels
The three classifications of digital imaging artifacts are _______________________. image receptor, software, and object
The software correction that equalizes the response of each pixel to a uniform exposure of x-rays is called ___________ flatfielding
Each anatomical part has an image histogram with a(n) _______________ shape. characteristic
What size is the average CR or DR image file per image? 20 MB
Backscatter radiation artifacts can occur more readily in digital imaging because of the _____________ highly sensitive receptors
A graph of the frequency of occurrence versus digital value intervals is called a(n) ______________ histogram
Which modality produces the largest image files? Digital mammography
Object artifacts in digital imaging can occur from errors in ________________ x-ray beam collimation, histogram selection, and patient positioning
Computer aided diagnosis (CAD) requires the use of ______________________. uncompressed images
The ability to image small objects that have high subject contrast is known as ____________. spatial resolution
Which statement is true regarding medical image quality? Spatial resolution is limited to pixel size.
Of all of the imaging modalities, _______________ is considered to have the best contrast resolution. MRI
As collimation increases, what is the effect on image contrast? Increases
As the x-ray beam field size increases, what is the effect on scatter radiation? Increases
The height of the grid divided by the interspace width is known as the __________________. grid ratio
Bucky factor = incident remnant x-rays/transmitted image-forming x-rays
Contrast improvement factor = image contrast with a grid/image contrast without a grid.
Which type of compression is up to 3:1 and is generally considered acceptable and helpful in digital radiographic image management? Lossless compression
Digital radiographic image receptors have unique artifacts associated with Pixel failure
What is a preprocessing software correction that is performed to equalize the response of each pixel to a uniform x-ray beam? Flatfielding
Which type of compression is greater than 10:1 and supports teleradiology but not CAD? Lossy compression
Created by: jetsetjes
 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards