| Question | Answer |
|---|---|
| single phase waveform (lowest power rating | 100% ripple |
| three phase, 6 pulse | 14% ripple |
| three phase, twelve pulse (higher efficiency and power) | 4% |
| operating console | allows adjustments for exposure factors |
| kVp | determines beam energy (penetration) quality |
| mA | determines tube current (quantity/ intensity) |
| timer | controls exposure duration |
| line voltage compensator | corrects fluctuations in incoming power(keeps voltage constant) |
| high voltage generator | converts low voltage (from autotransformer) into high voltage required for x-ray production |
| control console include | autotransformer, exposure timers, mA, and kVp |
| high voltage section | include step up transformer, and rectifiers |
| filament circuit | controls thermionic emission(heating of the filament) |
| autotransformer | function-select kVp, location- control console |
| step-up transformer | function- increases voltage to kilovoltage levels, location high voltage section |
| rectifiers | function-convert AC to DC, location- high voltage section |
| step down transformer | function- lowers voltage for filament heating, location- filament circuit |
| radiation quantity | intensity |
| radiation quality | penetrability |