Save
Upgrade to remove ads
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.
Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't Know
Remaining cards (0)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

QC

Red Boxes

TermDefinition
Ion Chamber 100-300V Least sensitive
Proportional counter 300-900V
Geiger Muller 900-1200V Most sensitive
1R 2.58e4 C/kg 1000mR
kVp accuracy HARP : +/- 8% Semi-annual SC35: +/- 10% Annual (Average kVp - set kVp)/set kVp
HVL HARP: 2.5 mmAl SC35: 1.5 mmAl
Reproducibility of Exposure HARP: CV<0.08, Exposure within 20% from mean - Semi-annual SC35: CV<0.05, Exposure within 15% from mean - Annual 1. Find Average 2. Find SD 3. Find CV 4. Variation from mean: (x-X)/X*100
Linearity HARP (Semi-Annual) |x1-x| < 0.10(x1+x2) SC35 (Annual)
Time HARP : 1/30 s or 5mAs - Semi-annual
AEC variation Variation = (max EXI - min EXI)/min EXI kVp --> SC35 --> 0.15 variation OD part thickness --> SC35 --> 0.20 variation OD
AEC testing back-up timer HARP: at 50 kVp <2000mAs - Semi-annual SC35: N/S: Annual
AEC min exposure time HARP: 1/60 s or 5mAs - Semi-annual SC35: N/S: Annual
AEC consistency with varying mA HARP: N/S - Semi-annual SC35: N/S: Annual
AEC consistency with varying kVp HARP: N/S - Semi-annual SC35: 0.15 variation in OD - Annual
AEC consistency with varying part thickness HARP: N/S - Semi-annual SC35: 0.20 variation in OD - Annual
AEC consistency with varying field sizes HARP: N/S - Semi-annual SC35: N/S: Annual
AEC consistency of detectors HARP: (max mR - min mR)/min mR <0.20- Semi-annual SC35: N/S: Annual
AEC reproducibility HARP: N/S - Semi-annual SC35: 0.10 variation in OD - Annual
AEC Density control HARP: N/S - Semi-annual SC35: N/S: Annual
Collimation Stepless Min field size > 5cm x 5cm @ 100 cm SID
Collimation - HARP Misalignment of Length and width +/2% of SID Size of x-ray field within 3% of SID
Collimation SC35 Misalignment of Length and width +/2% of SID PBL: Size of x-ray field within 3% of SID PBL: total misalignment should not exceed 4% of SID
Light-field /Radiation Field Alignment HARP +/- 2% of SID - S-A SC35 +/- 2% of SID - A
IR/Radiation Field Alignment HARP +/- 3% of SID - S-A SC35 +/- 3% of SID - A
Beam perpendicularity HARP Alignment +/- 2% of SID - S-A SC35 Alignment +/- 2% of SID - A
Focal Spot size HARP: N/S - Semi-annual SC35: N/S: Annual
Created by: user-1987883
 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards