click below
click below
Normal Size Small Size show me how
Bench Bio 240R 2026
2026 Bench 240R Statements of Fact
| 1. Use only exact replacement parts in the configuration as specified by the manufacturer. | |
| 2. The battery is to be changed in fresh air only. | |
| 3. MSHA approved for use with one of the following 9-Volt batteries only: Energizer 522, and Duracell MN 1604. | |
| 4. Never substitute, modify, add, or omit parts. | |
| 5. Prior to using the BioPak 240 Revolution it must be determined that the user is medically fit. | |
| 6. Always handle oxygen cylinders with care to prevent damage. | |
| 7. Do not open the cylinder valve in the presence of open flame, sparks, or high radiant heat. | |
| 8. Oxygen will enhance the combustion of other materials so that materials that normally will not burn in air may burn in O2-rich atmospheres; and materials that do burn in air will burn more vigorously and at a higher temperature in O2-rich atmospheres. | |
| 9. Oxygen will not cause materials to ignite without the presence of an ignition source. | |
| 10. The use of an SCBA will add to the workload and stress of the user. | |
| 11. The BioPak 240 Revolution is suitable for respiratory protection entry into and escape from oxygen deficient atmospheres with a temperature as low as -5 degrees F (-5F) (-20C). | |
| 12. The BioPak 240 Revolution is approved when the oxygen cylinder is fully charged with compressed medical or aviation grade oxygen at 3000 psi. | |
| 13. Allow the oxygen cylinder to cool after filling to determine the correct pressure. | |
| 14. A foreign gas may cause cylinder corrosion. | |
| 15. Always check for a current hydrostatic test date. | |
| 16. DOT requires carbon fiber wrapped aluminum cylinders be tested by an approved facility on a 5-year cycle from the date of manufacture. | |
| 17. An unapproved facepiece will compromise the protection provided to the user by the SCBA. | |
| 18. A good facepiece seal is important to achieving full protection and proper SCBA duration. | |
| 19. Users should conform to MSHA/NIOSH guidelines concerning facial hair and the use of facemasks. | |
| 20. Replace the battery when the low battery alarm has activated, after 200 hours of use or every 6 months whichever comes first. | |
| 21. The connectors of the monitoring device may only be connected to a Biomarine BioPak 240R breathing Apparatus oxygen regulator, manifold block and breathing chamber. | |
| 22. The fiber optic cable may only be connected to the BioPak 240R remote gauge assembly. | |
| 23. Turn-Around maintenance procedures should be performed as soon as possible after each use. | |
| 24. It is acceptable to leave the oxygen cylinder in place until after washing and disinfecting has been completed. | |
| 25. Immediately after completion of BioPak use, remove the used CO2 scrubber canisters. | |
| 26. DO NOT submerge the Alarm Module during turn-around maintenance. | |
| 27. Do not allow any fluids to contact the input port of the pressure regulator. | |
| 28. Use only cleaners and disinfectants that are approved by Biomarine. | |
| 29. If Cleaning is not immediately possible after each use, at a minimum remove and discard the CO2 scrubber and moisture pad. | |
| 30. Remove the oxygen cylinder making sure the seal washer or outlet tube O-ring remains in place and install the regulator cover. | |
| 31. Thoroughly rinse all components in clean water to remove all disinfectant solution. | |
| 32. Chronic Obstructive Pulmonary Disease could limit or prevent the use of the BioPak 240 Revolution. | |
| 33. Place freeze forms onto a level surface in a freezer for a minimum of 8-hour period at a temperature of 10 degrees F or less. | |
| 34. If the optional magnetic wiper is utilized soak both chamois surfaces of the wiper pieces with water. | |
| 35. The manual(s) are the minimum recommended procedures for maintaining the BioPak 240R. | |
| 36. Failure to follow the minimum procedures presented in the manual(s) may violate government or agency approvals as well as void the manufacturer’s warranty. | |
| 37. To prevent mold growth, remove the moisture control foam pads after each use. | |
| 38. Do not pre-pack any BioPak that will be stored at temperatures at or below 32 degrees F. | |
| 39. Pre-packed carbon dioxide scrubbers may only be stored in the apparatus for a maximum period of 1-year. | |
| 40. Record the Carbon Dioxide scrubber serial number and use by date on to the maintenance tag or affix the scrubber label to back of tag. | |
| 41. Moisture control sponges must be installed dry when pre-packing the BioPak. | |
| 42. Apparatus that are pre-packed with the carbon dioxide scrubber shall be stored within the specific storage temperature and humidity levels and must be sealed air-tight in the apparatus. | |
| 43. Failure to install the moisture pad will result in scrubber flooding and cause elevated carbon dioxide levels in the inhalation gas. | |
| 44. Users are not permitted to mix versions of the Orbsorb within a BioPak. | |
| 45. Install two carbon dioxide scrubber canisters into the breathing chamber making sure that they are properly aligned and fully seated. | |
| 46. The alarm system battery shall be replaced after 200-hours of use, every 6-months or after the alarm system low battery alarm flashes with corresponding horn sounding. | |
| 47. The oxygen cylinder must be fully charged to above 1500 psi to perform a high-pressure leak test. | |
| 48. Use caution when installing the flow restrictor to ensure that the O-ring does not roll out of its gland. | |
| 49. The Cylinder should be inspected regularly for signs of damage to the outer wrapping. | |
| 50. The Turn-Around Maintenance Tag should be attached to the apparatus in a prominent location to show completion of all maintenance steps. | |
| 51. In addition to normal Turn-Around Maintenance, the SCBA shall be visually inspected, and pressure tested on a monthly basis if the apparatus is being used at least once a month or is placed in long term storage. | |
| 52. BioPak’s that have been placed in long term storage should have the Long-Term Maintenance Procedure conducted every-6 months. | |
| 53. The LED indication will cease when the pressure gauge reads less than 25 psi. | |
| 54. Never pry an O-ring from its glands with a screwdriver. Remove O-rings by hand or with the pick tool provided in the service kit. | |
| 55. Cristo-Lube and Dow-111 are the only lubricants approved for use in the apparatus. | |
| 56. Never lubricate the outlet tube O-ring or the seal that sits between the oxygen cylinder and the pressure regulator. | |
| 57. Cylinders that have been hydro-static tested shall be cleaned for high-pressure oxygen service per national standards. | |
| 58. Cylinders are to be retired from service 15-years after the date of manufacture. | |
| 59. The Alarm module will require replacement if any damage to the housing is discovered. | |
| 60. If the flow does not meet the requirements of the table in the Bench Manual the flow restrictor will need replacement. | |
| 61. A good facemask seal is important to achieving full protection and duration. | |
| 62. Personnel who intend to use protective breathing equipment in a dangerous atmosphere must have the proper training, temperament, and experience. | |
| 63. A clean-shaven user will significantly increase the chances of achieving an adequate face seal. | |
| 64. The ongoing effectiveness and reliability of any protective breathing equipment is dependent upon the user’s standard of care in maintaining the equipment. | |
| 65. The BioPak has been tested for intrinsic safety in methane-air mixtures only. | |
| 66. The battery is to be changed in fresh air only. Do not change in hazardous areas. | |
| 67. The BioPak is a Self-Contained Closed Circuit Pressure Demand type certified as Entry and Escape with a 4-hour duration. | |
| 68. The constant Add is 1.8 liters average. | |
| 69. The Demand Add flow is a minimum 80 liters per minute. | |
| 70. The oxygen cylinder has a volume of 440 liters at 3000 psi. | |
| 71. The BioPak breathing chamber has a Tidal Volume greater than 6 liters. | |
| 72. For extreme temperature ranges, the BioPak should be configured with ice coolers. When ambient temperatures are greater than 140 degrees F the recommended duration is no more than 15 minutes and is limited by human endurance. | |
| 73. The emergency Add has a minimum flow rate of 80 liters per minute. | |
| 74. The BioPak operational conditions as it relates to relative humidity is 0 to 100%. | |
| 75. BioPak weight fully charged is 34 pounds. | |
| 76 .Do not re-use CO2 scrubber chemical. | |
| 77. The flow test results at 0-5280 feet at a cylinder pressure of 1500-2000 shall be 1.8-2.4 liters per minute. | |
| 78. During an alarm test the LED indication should turn to a flashing red with a horn sounding when the pressure gauge reads between 650-1000 psi. | |
| 79. Do not allow oil, grease, or other foreign materials to come in contact with cylinder, cylinder valve or cylinder pressure regulator to prevent possible ignition. | |
| 80. The end of service life or low Oxygen alarm is a flashing red light and horn sounding. | |
| 81. The flashing blue light indicates an Ice Reminder. | |
| 82. The pressure gauge is protected against sudden loss of oxygen in the event of a gauge line severing by a manual disconnect located at the gauge pass through point of the housing. | |
| 83. A Pacemaker or other Cardiac Condition could limit or prevent the use of the BioPak 240Revolution. | |
| 84. Breathing difficulties could limit or prevent the use of the BioPak 240 Revolution. | |
| 85. Claustrophobia or anxiety when wearing a SCBA could limit or prevent the use of the BIOPak 240 Revolution. | |
| 86. The instructions provided by the USER and BENCHMAN manuals cannot replace accredited training provided by qualified instructors in the proper and safe use of Biomarine breathing apparatus. | |
| 87. X-Ray evidence of Pneumonia could limit or prevent the use of the BioPak 240 Revolution. | |
| 88. Epilepsy-Grand Mal or Petit Mal could limit or prevent the use of the BioPak 240 Revolution. | |
| 89. Use the ¼ inch hex driver from the service Kit to remove the flow restrictor. | |
| Stack created by Jim Dean, WVU Mining Extension @ James.Dean@mail.wvu.edu |