click below
click below
Normal Size Small Size show me how
Bushong Ch38
Designing for Radiation Protection
| Question | Answer |
|---|---|
| Protective Tube Housing | Must be less than 100 mR/hr (1mGy/hr) at a distance of 1 m from protective housing. |
| Control Panel | X-ray beam on must be positively and clearly indicated to the rad tech. |
| Source-to-Image Distance Indicator | Tape measure or lasers; must be accurate w/in 2% of indicated SID. |
| Collimation | X-ray beam and light beam must coincide w/in 2% of SID |
| Positive-Beam Limitation | Automatic, light-localizer; must be accurate w/in 2% of SID. |
| Beam Alignment | Each tube s/be provided with a mechanism to ensure proper beam alignment. |
| Filtration | Total filtration (inherent plus added) must be at least 2.5 mm Al when operated above 70 kVp. |
| Reproducibility | Output radiation intensity s/be constant from one exposure to another. The variation in x-ray intensity s/not exceed 5%. |
| Linearity | The maximum acceptable variation in linearity is 10% from on mA station to an adjacent mA station. |
| Mobile X-ray Imaging System | System must allow the operator to remain at least 2 m from the x-ray tube during exposure. Useful beam must be directed away from rad tech. |
| Fluoroscopic Source-to-Skin Distance | SSD must be no less than 38 cm on stationary fluoroscopes and not less than 30 cm on mobile fluoroscopes. |
| Fluoroscopic Primary Protective Barrier | Fluoroscopic IR serves as a primary protective barrier and must be 2 mm Pb equivalent. |
| Fluoroscopic Filtration | Filtration must be at least 2.5 mm Al equivalent. |
| Fluoroscopic Collimation | Must be adjusted so unexposed border is visible on the image monitor when the input phosphor is positioned 35 cm above tabletop and collimators are fully open. |
| Fluoroscopic Exposure Control | Should be the dead man type |
| Fluoroscopic Bucky Slot Cover | Bucky tray is moved to end of exam table, leaving an opening in the side of the table approx 5 cm wide at gonadal level. S/have at least .25 mm Pb equivalent. |
| Fluoroscopic Protective Curtain | Panel must be at least .25 mm Pb equivalent between fluoroscopist and patient. |
| Fluoroscopic Cumulative Timer | Timer that produces an audible signal after 5 mins. To make radiologist aware. |
| Fluoroscopic Dose Area Product | Intensity of x-ray beam at the tabletop s/not exceed 2.1 R/min for each mA of operation at 80 kVp. DAP is a quantity that reflects not only the dose but also the volume of tissue irradiated. This is expressed in R-cm2 |
| Primary Protective Barrier | Any wall to which the useful beam can be directed |
| Scatter Radiation | Secondary radiation when useful beam intercepts any object such as a patient. |