CT Image production
| Question | Answer |
|---|---|
| _____________ refers to the number of xray photons produced in the beam | Intensity |
| the _____ is the target of the electron stream and the actual source of xray photon creation | anode |
| the ________ is the dedicated electronic circuit that allows the computer system to operate at high speeds, especially during data reconstruction | Array processor |
| the ______ is responsible for converting the electrical signal from the detectors into a digital signal prepared for processing | Data acquisition system (DAS) |
| _______ are responsible for converting remnant beam to an electrical signal | Detectors |
| Information leaving the data acquisition system(DAS) is known as _____, which is a digital matrix of numbers | Raw data |
| Which component of the CT scanner measures the radiation reading produced by the detector array? | Data acquisition system (DAS) |
| Reconstruction of raw data into image data occurs inside the ________ | Computer system |
| the ______er scanner can collect more anatomic information with each rotation of the tube, which results in faster scan time | larger |
| The number of data acquisition channels (or slices) is `__________with change in image noise, spatial resolution, and contrast resolution | not directly associated |
| In CT the x axis extends ___ | Horizontally (right and left) |
| in ct the y axis extends _______ | Vertically (anterior to posterior) |
| in ct the z axis extends ______ | Longitudinally , (superior to inferior) |
| the amount of data collected per tube rotation in axial scanning is dependent on the number of ________ | Activated detector rows |
| As aquisition thickness increases, patient exposure ________ | decreases |
| combining detector rows _________ aquisition thickness | Increases |
| as aquisition thickness increases, spatial resolution __________ | Decreases |
| as pitch increases, spatial resolution __________ | decreases |
| as pitch increases, patient dose __________ | Decreases |
| as pitch decreases, spatial resolution _______ | increases |
| helical scanning may slightly _______ spatial resolution | decrease |
| as slice thickness increases, spatial resolution ______ | decreases |
| what is the max size of scan field of view (SFOV) in most GE scanners? | 50cm |
| GE scanners have a small SFOV of __cm and large SFOV of __cm | 25 cm 50 cm |
| During axial CT, the patient table moves_____ | foward between each sequential exposure |
| _____ is the ratio of table movement to the beam width | pitch |
| pitch applies to _______ scanning only | Helical |
| Pitch = | table movement / beam width |