click below
click below
Normal Size Small Size show me how
171 Final
| Question | Answer |
|---|---|
| Is the distortion of image-intensified fluoroscopic image that appears as unequal magnification: | Pincushion appearance |
| Releasing the pressure applied to the______will terminate the radiation exposure during fluoroscopy: | Deadman switch |
| Is coated with light-emitting crystals to absorb electrons and emit light: | Output phosphor |
| Absorbs light and converts light to electrons: | Photocathode |
| The gain related to the decrease in size from the input to the output phosphor that increases the light intensities is the: | Minification gain |
| In reference to image intensification, the function that adjusts and maintains the overall image brightness and contrast during fluoroscopic procedure is: | ABC |
| Magnification of the fluoroscopic image results in improved: | Spatial resolution |
| Is essentially the sum total of air kerma over the exposed area of the patient’s body surface: | Dose area product |
| Types of ionizing radiation that produce virtually the same biologic effect for equal absorbed doses in body tissue? | X-rays, beta particles, and gamma rays |
| Units are not SI units: | Roentgens, rads and rems |
| Scatter radiation is primarily the result of: | Compton interaction |
| The probability of a Compton interaction occurring is related to: | The energy of the x-ray photon |
| The two major factors that affect the amount of scatter radiation production produced and exiting the patient are: | Volume of tissue irradiated and kVp |
| The volume of tissue irradiated is determined by: | Area of collimation, patient thickness, field size |
| The most effective and often used radiographic beam restriction device is the: | Collimator |
| The field shape produced by a Collimator is: | Rectangular |
| The purpose of automatic collimation is: | Limiting patient exposure |
| A focus grid lead strips are angled to match: | The angle of the x-rays in the primary beam |
| The recommended SID that can be used with focused grid is the: | The focal range |
| To compensate for the reduction of radiation reaching the IR when grid is used___ is typically increased: | mAs |
| Absorption of transmitted photons by a grid caused by misalignment is: | Grid cutoff |
| A grids provides excellent scatter clean-up but requires close to perfect alignment to avoid cutoff: | 16:1 ratio |
| Air-gap technique reduces the amount of scatter radiation reaching the IR because of: | More scatter misses the IR |
| T/F When using automatic collimation, collimating to the anatomy of interest is an unnecessary waste of time; the beam is already collimated: | False |
| T/F grids absorbed almost all the scatter radiation and allow all of the transmitted photons to pass through to the IR: | False |
| T/F adding a grid or increasing grid ratio results in an increase in patient dose: | True |
| Absorbs x-rays and emits light: | Input phosphor |
| Is bonded directly to the input phosphor: | Photocathode |
| Photoemission occurs at the: | Photocathode |
| Is found along the entire length of the image intensifier: | Electrostatic focusing lenses |
| When operated in magnification mode, the electrostatic lenses have ___of a negative charge: | More |
| The disadvantage of using magnification mode during fluoroscopy is: | Patient receives additional dose |
| When the electrons beam is sweeping the anode, if the electron beam and light from the output phosphor are incident on the same place at the same time, ______through the target to the signal plate: | Electron transmitted |
| With digital systems, this is important to monitor to avoid dose creep: | Exposure index number |
| A histogram: is a technique chart, a graph of all data points, extraction of the latent image, values of interest? | A graph of all data points |
| With digital systems, the two detector types are: | CR and DR |
| Are not additive disease: | Bowel obstruction |
| Device that measure part of thickness: | Calipers |
| Back-up time for AEC exposure should be set to: | 150-200% of expected exposure time |
| AEC determines exposure by controlling the? | Length of exposure |
| When using mag mode, voltage to the focusing lenses is: | Increased |
| The increase brightness in fluoroscopy? | Image intensifier |
| When using the air-gap technique, the ___ must also be increased: | SID |
| The number of lead lines per unit length is the: | Grid frequency |
| Which is the most common beam restricting device: | Collimators |
| Are beam restricting devices? | Cylinders, Collimator, aperture diaphragm |
| The SI equivalent for REM is? | Sievert |
| The SI equivalent of RAD is: | Gray |
| T/F the latent image from the PSP plate remains in place for hours after exposure: | True |
| T/F digital systems are more sensitive to scatter radiation: | True |
| Part of the direct capture DR system, the____absorbs x-rays and produces an electronic signal: | Photoconductor |
| T/F histogram analysis is the basis for determining exposure index: | True |
| The Bucky factor(Grid conversion factor) for a grid ratio of 8:1 is? | 4 |
| Part of PSP plate that includes a random distribution of crystals is the: | Turbid phosphor layer |
| The most important layer for creating the image; the layer of film that is sensitive to light or radiation is the: | Emulsion |
| What is the role of the CCD in indirect DR? | Converts light to electrons |
| Component of the CR reader converts light to an electronic signal? | Photodetector |
| What part of the CR reader digitizes the signal? | ADC |
| Describes the extra step, and is therefore a limitation of indirect-capture methods? | X-ray are converted to light and then to electrons |
| T/F grids are used with digital systems in the same way as film/screen? | True |
| Statement is concerning computed radiography(CR) system: | Use Storage phosphor and require the storage phosphor |
| Sequences of events for indirect capture systems in order: | Absorbed, transmitted, converts, signal |
| Correct order for automatic film processing: | Developing, fixing, washing, drying |
| The scintillator for the indirect capture DR system that utilizes a CCD is: | Cesium iodide |
| Part of DR system with indirect capture, the light-sensitive device that is commonly in digital cameras is the: | CCD |
| Intensifying screens are used to: | Reduce patient exposure, convert x-ray photons |
| Once the image is required, what is the role of PACS? | Display and storage |
| T/F with digital systems, a higher kVp can be used than with film/screen: | True |
| Part of the CR reader detects light released from the PSP plate: | Photo detector |
| Fluorescence refers to: | The ability of phosphors to emit visible lights only while exposed to x-rays |
| Digital imaging control focuses on: | Display and viewing |
| T/F the DAP is used for DR digital systems and is a measure of the dose to the patient: | True |
| When the PSP plate is exposed to the laser of the reader, the energy is released and converted to a digital signal, becoming a: | Manifest image |
| What is the primary difference of direct DR over indirect DR? | Lacks the conversion of x-rays to light |
| The DR image capture method that uses an amorphous silicon photodetector and TFT represents of the: | Indirect |
| Delivery of hot air to the surface of the film is the function of: | Drying |
| The type of phosphor found in the PSP plate for computed radiography is: | Barium fluorohalide |
| What part to the flat-panel DR converts x-rays to light? | Scintiliator |
| T/F DR image receptors do not use histograms: | False |
| Europium is the_____of the photostimulable phosphor: | Activator |
| T/F after the digital data leaves the ADC, the process of digital image formation is very different for the three types of DR systems: | False |
| Histogram analysis compares the histogram with pre established histogram specific to: | The anatomic part being imaged |
| DICOM serves what purpose in PACS? | Allow different system to communicate |
| T/F direct DR does not use a scintillator: | True |
| Is the standard format for digital medical Image: | DICOM |
| The part of a PSP plate that gives some rigidity to the plate is the: | Support layer |