Save
Upgrade to remove ads
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.
Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't Know
Remaining cards (0)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

171 Final

QuestionAnswer
Is the distortion of image-intensified fluoroscopic image that appears as unequal magnification: Pincushion appearance
Releasing the pressure applied to the______will terminate the radiation exposure during fluoroscopy: Deadman switch
Is coated with light-emitting crystals to absorb electrons and emit light: Output phosphor
Absorbs light and converts light to electrons: Photocathode
The gain related to the decrease in size from the input to the output phosphor that increases the light intensities is the: Minification gain
In reference to image intensification, the function that adjusts and maintains the overall image brightness and contrast during fluoroscopic procedure is: ABC
Magnification of the fluoroscopic image results in improved: Spatial resolution
Is essentially the sum total of air kerma over the exposed area of the patient’s body surface: Dose area product
Types of ionizing radiation that produce virtually the same biologic effect for equal absorbed doses in body tissue? X-rays, beta particles, and gamma rays
Units are not SI units: Roentgens, rads and rems
Scatter radiation is primarily the result of: Compton interaction
The probability of a Compton interaction occurring is related to: The energy of the x-ray photon
The two major factors that affect the amount of scatter radiation production produced and exiting the patient are: Volume of tissue irradiated and kVp
The volume of tissue irradiated is determined by: Area of collimation, patient thickness, field size
The most effective and often used radiographic beam restriction device is the: Collimator
The field shape produced by a Collimator is: Rectangular
The purpose of automatic collimation is: Limiting patient exposure
A focus grid lead strips are angled to match: The angle of the x-rays in the primary beam
The recommended SID that can be used with focused grid is the: The focal range
To compensate for the reduction of radiation reaching the IR when grid is used___ is typically increased: mAs
Absorption of transmitted photons by a grid caused by misalignment is: Grid cutoff
A grids provides excellent scatter clean-up but requires close to perfect alignment to avoid cutoff: 16:1 ratio
Air-gap technique reduces the amount of scatter radiation reaching the IR because of: More scatter misses the IR
T/F When using automatic collimation, collimating to the anatomy of interest is an unnecessary waste of time; the beam is already collimated: False
T/F grids absorbed almost all the scatter radiation and allow all of the transmitted photons to pass through to the IR: False
T/F adding a grid or increasing grid ratio results in an increase in patient dose: True
Absorbs x-rays and emits light: Input phosphor
Is bonded directly to the input phosphor: Photocathode
Photoemission occurs at the: Photocathode
Is found along the entire length of the image intensifier: Electrostatic focusing lenses
When operated in magnification mode, the electrostatic lenses have ___of a negative charge: More
The disadvantage of using magnification mode during fluoroscopy is: Patient receives additional dose
When the electrons beam is sweeping the anode, if the electron beam and light from the output phosphor are incident on the same place at the same time, ______through the target to the signal plate: Electron transmitted
With digital systems, this is important to monitor to avoid dose creep: Exposure index number
A histogram: is a technique chart, a graph of all data points, extraction of the latent image, values of interest? A graph of all data points
With digital systems, the two detector types are: CR and DR
Are not additive disease: Bowel obstruction
Device that measure part of thickness: Calipers
Back-up time for AEC exposure should be set to: 150-200% of expected exposure time
AEC determines exposure by controlling the? Length of exposure
When using mag mode, voltage to the focusing lenses is: Increased
The increase brightness in fluoroscopy? Image intensifier
When using the air-gap technique, the ___ must also be increased: SID
The number of lead lines per unit length is the: Grid frequency
Which is the most common beam restricting device: Collimators
Are beam restricting devices? Cylinders, Collimator, aperture diaphragm
The SI equivalent for REM is? Sievert
The SI equivalent of RAD is: Gray
T/F the latent image from the PSP plate remains in place for hours after exposure: True
T/F digital systems are more sensitive to scatter radiation: True
Part of the direct capture DR system, the____absorbs x-rays and produces an electronic signal: Photoconductor
T/F histogram analysis is the basis for determining exposure index: True
The Bucky factor(Grid conversion factor) for a grid ratio of 8:1 is? 4
Part of PSP plate that includes a random distribution of crystals is the: Turbid phosphor layer
The most important layer for creating the image; the layer of film that is sensitive to light or radiation is the: Emulsion
What is the role of the CCD in indirect DR? Converts light to electrons
Component of the CR reader converts light to an electronic signal? Photodetector
What part of the CR reader digitizes the signal? ADC
Describes the extra step, and is therefore a limitation of indirect-capture methods? X-ray are converted to light and then to electrons
T/F grids are used with digital systems in the same way as film/screen? True
Statement is concerning computed radiography(CR) system: Use Storage phosphor and require the storage phosphor
Sequences of events for indirect capture systems in order: Absorbed, transmitted, converts, signal
Correct order for automatic film processing: Developing, fixing, washing, drying
The scintillator for the indirect capture DR system that utilizes a CCD is: Cesium iodide
Part of DR system with indirect capture, the light-sensitive device that is commonly in digital cameras is the: CCD
Intensifying screens are used to: Reduce patient exposure, convert x-ray photons
Once the image is required, what is the role of PACS? Display and storage
T/F with digital systems, a higher kVp can be used than with film/screen: True
Part of the CR reader detects light released from the PSP plate: Photo detector
Fluorescence refers to: The ability of phosphors to emit visible lights only while exposed to x-rays
Digital imaging control focuses on: Display and viewing
T/F the DAP is used for DR digital systems and is a measure of the dose to the patient: True
When the PSP plate is exposed to the laser of the reader, the energy is released and converted to a digital signal, becoming a: Manifest image
What is the primary difference of direct DR over indirect DR? Lacks the conversion of x-rays to light
The DR image capture method that uses an amorphous silicon photodetector and TFT represents of the: Indirect
Delivery of hot air to the surface of the film is the function of: Drying
The type of phosphor found in the PSP plate for computed radiography is: Barium fluorohalide
What part to the flat-panel DR converts x-rays to light? Scintiliator
T/F DR image receptors do not use histograms: False
Europium is the_____of the photostimulable phosphor: Activator
T/F after the digital data leaves the ADC, the process of digital image formation is very different for the three types of DR systems: False
Histogram analysis compares the histogram with pre established histogram specific to: The anatomic part being imaged
DICOM serves what purpose in PACS? Allow different system to communicate
T/F direct DR does not use a scintillator: True
Is the standard format for digital medical Image: DICOM
The part of a PSP plate that gives some rigidity to the plate is the: Support layer
Created by: Faithshay
 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards