Save
Upgrade to remove ads
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.
Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't Know
Remaining cards (0)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

12 ARTIFACTS

REVIEW

QuestionAnswer
A pulse sequence that demonstrates the least amount of magnetic susceptibility artifact: Both spin echo and fast spin echo sequences. We do not see magnetic susceptibility artifact in images which are collected using a 180° RF pulse to form the echo.
Magnetic susceptibility, is a measure of the tendency of a material to become magnetized. The artifact results in tissues with differing magnetic susceptibilities having different resident frequencies. For this reason, it is a dephasing mechanism
In which of the following directions may Aliasing occur with a 3-D volume acquisition: Slice select direction, read out direction, phase encoding direction. Occurs in 3D and 2D imaging when anatomy extends past the selected FOV. In 3D imaging, it can also occur along the volume in the slice direction. Signal maybe missmapped from the outer edges
Scan time with no frequency wrap option applied is the same as the scan time if the no frequency wrap option is not applied Sampling additional frequencies does not affect scan time. It is a "free" way of eliminating wraparound in the frequency encoding direction.
We only correct for wraparound in the phase direction when necessary: Additional phases increases scan time. If a no phase wrap option is available, it may make other changes to the scan protocol (halving NSA) so that scan time is not effected.
Spoiled gradient echo pulse sequences display effects of magnetic susceptibility Any pulse sequence which does not utilize the 180° RF pulse to rephase the magnetization in the traverse plane may demonstrate the effects of magnetic susceptibility
How can fat suppression act as a motion reduction technique? Fat suppression reduces the signal from fatty tissues, and with a darker signal, the tissues will generate less motion artifact.
Fat suppression is a technique which uses a spectrally selective saturation region to eliminate the signal from fat: These fatty tissues, now rendered low in signal intensity, will generate less of a motion artifact.
Metal artifacts in MR images appear as a local distortion of the signal in the image, for nonferrous metals. Or is it severe distortion of the signal in the image in the case of Ferris metals This can occur if the metal is within the selected FOV, or even if the metal is very near the selected FOV
Transverse image through kidneys example: The kidneys are a water-based tissue with little fat. They are however, surrounded by peri-renal fat. Presenting an ideal location in the body for chemical shift artifact to occur. There is shift of the kidneys relative to the peri-renal fat, along the frequency encoding direction of the image
Chemical shift artifact results from a shift in the frequencies of fat-based protons relative to water-based protons. Can be minimized by increasing the receiver bandwidth such that each pixel corresponds to a wider range of frequencies; In this way, a given chemical shift will translate to fewer pixels of shift. Note, reducing the matrix will not reduce chemical shift artifact. In fact, it will increase pixel size, and shift translated.
Several ways to reduce the severity of flow related artifacts: Utilizing gradient moment nulling to properly rephase the signal from flowing protons. Saturation pulses to saturate the protons of inflowing blood. Physiological triggering and gating schemes. Note, the use of a narrow receiver bandwidth will improve the SNR of the image and can make flow artifacts more pronounce.
The dephasing effects of magnetic susceptibility are re-phased with a 180° RF pulse. Therefore we do not see magnetic susceptibility artifacts in images which are collected using a 180° RF pulse to form the echo It describes the tendency of a material to become magnetized
Chemical shift artifact appear as a shift, or miss mapping, of signal along the frequency direction of the image. Fat-based tissue shifts away from the water-based tissue, there is a signal void, thus, a dark pixel intensity Where the fat-based tissues shift towards the water-based tissue, there is an addition of signal from overlapping tissue, and a bright pixel intensity results.
The dark pixel intensity that results from a chemical shift artifact arises from: A signal void which is caused by a signal miss-mapping along the frequency encoding direction of the image
Signal fall off artifact System related, occurs when using surface coils
Aliasing artifact on volume images: Some systems have automatic features to eliminate this artifact in the slice direction that automatically collects more slices the needed and then discard the extra slices containing the artifacts. Ex: volume sagittal slice image, applied both no phase & the no frequency wrap option. slice. Wrap still occurs in volume/slice direction. both no frequency/no phase wrap option selected: no wraparound at top, bottom, left or right side to the image
Applying a no frequency wrap option will eliminate wraparound artifact from the top and bottom of a coronal knee image. Application of no frequency wrap option will effectively eliminate the wraparound from the top and bottom of a coronal on the image In coronal and sagittal imaging, increasing FOV is not an affective option for eliminating wraparound because there'll always be anatomy extending outside the FOV, along the long axis of the body
We can eliminate aliasing artifact in the frequency encoding direction by: Using the no frequency wrap option with no scan time penalty
RF artifacts appear as a line orientated perpendicular to the frequency encoding direction of the image. Aliasing artifact appears as miss mapping of signal from one side of the image to the opposite side. Truncation artifact appears as a ringing of high contrast Borders in the image. Metal artifact includes local distortion of the image and areas of signal void with associated bright signals
Metal artifacts cannot be removed from the image without removing the offending metal from the patient. There are no other ways to eliminate them, they however do increase in severity with gradient echo pulse sequences Because gradient echo sequences do not utilize the 180° pulse to rephase the echo. Therefore they're more susceptible to anything that causes local distortion in the magnetic field
Three ways to eliminate aliasing artifact: Increasing FOV, Saturation pulse, and utilizing no wrap option No wrap option: sampling more frequencies or phases allows us to determine those signals emanating from outside FOV for elimination.
Chemical shift artifact: Results from how the resident frequencies form fat base protons and water-based protons differ. Translating into positional differences in the image along the frequency encoding direction We can minimize this shift in the image by utilizing a wider receiver bandwidth. This way each pixel will represent a wider range of frequencies, and the chemical shift will translate into fewer pixels of shift
Chemical shift artifact occurs at the interface of two tissues who's protons experience different resident frequencies Magnetic susceptibility artifact occurs at the interface of two tissues that experience different magnetic susceptibilities
Result of reducing slice thickness to reduce partial Voluming effects: Decreased SNR, increased spatial resolution, decreased anatomical coverage with the same number of slices. A thinner slice yields improve spatial resolution along a slice spacial direction. Thinner slices will decrease the SNR, since fewer protons are contributing to the signal on each voxel.
True regarding Truncation artifacts: And apparent when a high contrast border falls within a single pixel, reduce by applying a raw data filter, reduced by using a high value for the image matrix: Utilizing a raw data filter to minimize the chopping off of the side lobes of the echo. Utilizing a larger matrix will reduce the likelihood of the sharp contrast border falling within one pixel, since the pixels are now smaller.
Chemical shift artifact will appear in the: Frequency encoding direction
Created by: tomvilla21
 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards