7 Volume imaging
Review
| Question | Answer |
|---|---|
| Conventional spin echo pulse sequences are rarely used for acquiring the anatomical volume data used for NPR because | Excessive time required with long TR of spin echo sequences are used in volume imaging. 2 Gaps from crosstalk in 2D the spin echo imaging. 3. Low SNR ratios result when acquiring thin slices in 2D the spin echo imaging |
| The formula for calculating scan time for a 3-D acquisition is: | Measurement time= (TR) * (PHASE LINES IN MATRIX) * (#SLICES) * (NSA) / Echo train length |
| In a 3-D volume pulse sequence with a 256 x 256 matrix and 16 transverse slices,_________ unique phases must be generated by the Z gradient, assuming one acquisition: | As many fees and coding steps are required along the slice direction as the number of slices: 16 |
| In a 3D volume pulse sequence with a 256 x 256 matrix, 16 transverse slices, and the Y-gradient responsible for in plain phase encoding,_____unique phases must be generated by the Y-gradient assuming one acquisition and no RFOV or parallel imaging: | The in-plane special encoding of the traverse slices are accomplished by the Y-and-X gradients. The X- gradient is used as the slice select gradient. 256 |
| In a 3-D volume pulse sequence with a 256 x 256 matrix, 16 transverse slices, and the Y-gradient responsible for in-plain phase encoding,_____ unique phases must be generated by the X-gradient, assuming 1 acquisition: | NO. In-plane special encoding of the transverse slices are accomplish by the Y-and-X gradients. Therefore, the X-gradient is used for frequency encoding and it's not used to generate phase information in the sequence. |
| In a 3D spoiled gradient echo sequence,_________contributes to the image contrast: Spin spin interactions magnetic susceptibility effects magnetic inhomogeneities | ALL |
| Which of the following is TRUE of the specific absorption rate of a 3D spoiled gradient echo sequence with 16 slices VS a 2D spoiled gradient echo sequence with 16 slices, assuming all other parameters are identical: | The SAR is the same on both sequences. Total number of RF pulses = (phase matrix) * ( number of signal averages) * (number of slices) |
| Scan Time of a 3D pulse sequence with 205 phase steps in the image matrix, a TR of 40 ms, a FOV of 28 cm, 20 slices, 2 signal averages, and echo train length of 1 and no parallel imaging is_______ minutes: | Therefore, the scan time = (TR) * (phase matrix) * ( NSA) * (#of slices) * (echo train length). =0.04sec * 205 * 2 * 20 / 1 = 328sec =5.5 minutes |
| In a set of volume images, emotion artifact may be seen along the: 1) in-plane phase direction. 2) Slice direction, from one image to another. 3) frequency direction. | 1&2 |