click below
click below
Normal Size Small Size show me how
Varkey Final
2dn Year 1st semester
| Question | Answer |
|---|---|
| 1. What is the state-required filtration for radiographic x-ray tubes? | • 2.5 mm Al |
| 2. What is the state requirement for the accuracy of the collimation? How often do you need to calibrate it? | • Misalignment should not exceed 2% OF SID. • The distance indicator must be accurate to 2% of SID. • Centering indicator must be accurate to 1% of SID. • Must be checked semiannually |
| 3. Name the three methods used to evaluate focal spot size. How often calibrated? | • Pinhole camara • Star pattern • Slit camara • Checked annually or when new equipment or replacement x-ray tube is installed |
| 4. What is the most recent method used for focal spot size evaluation? | • Line pair test tool. |
| 5. ___________% change in kVp is required for a visible change in OD. | • 4% |
| 6. ___________% change in mAs is required for a visible change in OD. | • 30% |
| 7. Measured kVp should be within _______________% of the indicated kVp. | • 10% |
| 8. Accuracy of the exposure timer must be ______________% for exposure time less than 10 ms. | • 20% |
| 9. Accuracy of the exposure timer must be ____________% for exposure time greater than 10 ms. | • 5% |
| 10. ___________Provide constant OD regardless of tissue thickness, composition and exposure time. | • Automatic Exposure Control (AEC) |
| 11. A ____________________ will terminate the exposure at 6s or 600 mAs whichever occurs first. | • Backup timer |
| 12. Exposure linearity must be within __________________% of the adjacent mA stations. | • 10% |
| 13. mR/mAs value must be within ___________________% between adjacent mA stations. | • 10% |
| 15. A ____________is used to measure light intensity at various locations of the illuminator and the values should not vary more than ______% | • Output Radiation • 5% |
| 16. Incomplete erasure of a previous image on a CR Imaging plate might cause _____________ | • Ghost image |
| 17. Failure of electronic preprocessing can cause uninterpretable images due to ________________ | • One dead pixel. • Rows or columns of dead pixels |
| 18. Too rapid exposures can cause ____________________________artifacts. | • Electronic memory artifacts |
| 19. What is interpolation? | • It’s a correctional algorithm. Interpolation is a mathematical process of assigning a digital value to a dead pixel based on recorded values of adjacent pixels. |
| 20. Incorrectly oriented grid, with grid lines parallel to plate reader’s scan lines will cause: _________________ artifact. | • Moiré |
| 21. Frequency harmonics can cause_______________artifacts. | • Moiré |
| 22. Noise in the plate reader electronics can cause ____________________ patterns. | • Extraneous line patterns. |
| 23. A black line on the digital image is caused by _______________________________ | • Backscatter |
| 24. If the x-ray exposure field is not properly collimated, sized, and positioned, ___________errors may occur. | • Exposure field recognition. |
| 25. Failure of electronic preprocessing can cause uninterpretable images due to the presence of _________on the image matrix | • Software artifacts: o One dead pixel. o Rows or columns of dead pixels |
| 26. The software correction that equalizes the response of each pixel to a uniform exposure of x-rays is called ______________________________ | • Flat fielding |
| 27. Name the tools that can be used to standardize the performance of digital display devices. | • SMPTE • NEMA – DICOM • DIN 2001 • VESA • AAPM TG 18 |
| 28. Study SMPTE Pattern and AAPM TG 18 | • SMPTE: o Used for the measurement of the resolution of display system. o Help to point out any gross deviation in luminance adjustment. o (5% and 95% Luminance Patches) |
| 28. Study SMPTE Pattern and AAPM TG 18 2nd part | • AAPM TG 18: o Developed a set of test patterns and outlined related procedures in Task Group Report 18 o Emphasis is placed on details of associated patterns that can be used by the technologist to perform checks to ensure system performance |
| 29. What is SMPTE stands for? | • Society of Motion Pictures and Television Engineers. |
| 30. SMPTE can be used for evaluating 1)___________________. 2)_______________ | • Measurement of the resolution of display system. • Help to point out any gross deviation in luminance adjustment. |
| 31. What is AAPM TG 18 stand for? | • American Association of Physicists in Medicine. |
| 32. What is Geometric distortion? Give two examples. | • Displayed imagen appears geometrically different from the original image. • Examples: pincushion and barrel. |
| 33. Which test pattern can be used to evaluate pincushion and barrel distortions? | • TG 18 QC Test |
| 34. Identify pincushion and barrel distortion from an image | Barrel Pincushion |
| 35. What is the acceptable level of distortions? | • The acceptable level of distortion in primary class devices is 2%. • The acceptable level of distortion in secondary class devices is 5%. |
| 36. What is the difference between specular reflection and diffuse reflection? | • Specular reflection: the viewer sees a reflection at only one point. • Diffuse reflection: light from all the points on the surface reaches the viewer. |
| 37. How can you minimize reflections from a display monitor? | • Anti-reflective coating can minimize reflections. |
| 38. What test pattern can be used to evaluate diffuse reflection? | • TG 18 AD |
| 39. What is luminance response? | • Refers to the relationship between displayed luminance and input values of a standardized display system |
| 40. What test pattern can be used for luminance non-uniformity? | • TG 18 UN 10 and 80 test patterns |
| 41. What is the acceptable value of non-uniformity? | • Maximum non-uniformity of a display device should be less than 30%. |
| 42. What test pattern can be used to evaluate display resolution? | • TG 18 CX and TG 18 QC |
| 43. What test pattern can be used to evaluate resolution uniformity? | • TG 18 PX |
| 1. In reference to body planes, name four different types of images that can be obtained in CT Scan | • Axial • Coronal • Sagittal • Oblique |
| 2. Name the plane that divide the body into superior and inferior sections. | • Horizontal |
| 3. Name the plane that divide the body into right and left sections | • Sagittal (axial) |
| 4. Name the plane that divide the body into anterior and posterior sections. | • Coronal |
| 5. Name the plane that divide the body into equal halves of anterior and posterior sections. | • Midcoronal |
| 6. Who invented clinical computed tomography? | • Dr. Godfrey Hounsfield |
| 7. Write four features of First-Generation CT scanner. | • Pencil-thin x-ray beam • One detector • Translate rotate scanning motion • 4.5 minutes for one image |
| 8. Write the name of a type of filter used in CT scan. | • High Pass |
| 9. Write four features of Second-Generation CT scanner. | • Fan beam • 30 or more detectors • Rotate translate scanning motion • 15 seconds for one image |
| 10. Write four features of Third-Generation CT scanner. | • Fan beam • 960 detectors (curved detector array) • Rotate continuously around the patient for 360 degrees • Less than one second per image |
| 11. Write four features of Fourth-Generation CT scanner. | • Developed during 1980’s • Fixed ring of 4800 detectors • Rotate continuously around the patient for 360 degrees • Less than one second per image. • Not more advanced than third generation |
| 12. Which invention in CT scan replaced the high-tension cables and made it possible for the x-ray tube and detector to rotate around the patient continuously without the need for stop and rewind? | • Slip Ring |
| 13. Write three functions of the slip ring technology. | • Transfers electrical power, scanning instructions, and detected signals between stationary components and rotating gantry components. |
| 13. Write three functions of the slip ring technology... | • Eliminate the need to stop the rotation of the gantry assembly after each scan since there is no cables to wind up. • Led to the development of helical/spiral scanning (volume scanning) |
| 14. Explain the difference between a single slice CT and a Helical CT. | • Helical CT: X-ray tube rotate continuously around the patient as the patient/table increment through the gantry at selected speed and a volume of data is collected. In helical CT the data is obtained from a volume of tissue. |
| 14. Explain the difference between a single slice CT and a Helical CT...2nd part | These information need to be divided into individual cross sectional slices. The difficulty in this task is that the data needed to define a given slice was measured through projections which do not correspond to a plane. |
| 14. Explain the difference between a single slice CT and a Helical CT...3rd part | • Single Slice CT: The x-ray tube rotate around the patient 360 degree at a selected location to create an image. The patient/table moves to a new location in the gantry to create the next image. Imaging only one slice at a time. |
| 15. What are the advantages of volume scanning (Spiral CT) | • Multiplanar reconstruction • Shorter scan time • Reduced motion artifacts • Less amount of IV Contrast injection |
| 17. List four advantages of multi-slice scanners. | • Short acquisition time • Decreased amount of contrast media. • Improved spatial resolution. • Improved image quality. |
| 18. List three major components of a CT Scanner. | • Gantry • Computer • Operator console |
| 21. Explain the structure and function of detector array in CT scan. | • Solid state detectors. • Composed of photodiode coupled with scintillation crystal material. • Convert transmitted x-ray into light. |
| 21. Explain the structure and function of detector array in CT scan... 2nd part | • Light is converted into electrical energy and then into digital signal. • Affects patient dose and the efficiency of the CT unit |
| 22. What is the function of a pre-patient collimator? | • Pre-patient collimator determines the slice thickness. • Reduce patient dose and improves image quality. |
| 23. What is the function of a post-patient collimator? | • Post-patient collimator further defines the slice thickness |
| 25. Which collimator reduces patient dose in the CT scanner? | • Pre-patient collimation. |
| 26. How many mathematical calculations per slice will be processed by the computer when using a 256 X 256 matrix size? | • 65536 |
| 27. What are the components in the operator console? | • Keyboard • Input device (mouse or trackball) • Monitor |
| 28. What are the functions of the operator console? | • Allows the tech to control the parameters of the examination (Protocol) • View and manipulate the image. |
| 29. List the various image archiving devices used in CT scan | • Magnetic Tape • Hard Disks • Optical disks • PACS |
| 30. PIXEL stands for _________________ | • Picture Element |
| 31. VOXEL stands for ________________ | • Volume element or volume pixel |
| 32. The number of rows and columns in a digital CT image is known as _______________ | • Image Matrix |
| 33. A three-dimensional tissue volume in a CT image is represented by _____________(Pixel/Voxel) | • Voxel |
| 34. A Pixel in a CT image represent_________of a volume of tissue at that position in the slice. | • Number |
| 35. The depth of a VOXEL represents ___________in a CT image | • The slice thickness |
| 36. The degree of attenuation is known as ___________________ | • Attenuation coefficient |
| 37. A CT Number represent ________________ | • The degree of attenuation values of each voxel |
| 38. CT Numbers were also known as ___________________ | • Hounsfield Units |
| 39. The baseline for CT number is _______, which is assigned a CT number of ___________ | • Water; 0 |
| 40. A Positive CT number indicate_____________anatomy (Radiopaque/Radiolucent) | • Radiopaque |
| 41. Iodinated Contrast media filled structures appears ____________on a CT image. | • White |
| 42. Air appears ________________on a CT image | • Black |
| 43. Fat Muscle, and organs appear ___________on a CT image | • Dark gray to black (varying shades of gray) |
| 44. Image contrast can be manipulated by ____________ (WW/ WL) | • Window width (WW) |
| 45. A wide window width indicates ______________ (low contrast/ High contrast) | • Low contrast (which is long scale) |
| 46. A narrow window width indicates _____________ (short-scale contrast/ Long scale contrast) | • Short-scale contrast (high contrast) |
| 47. Window level controls _______________ (image brightness/ image contrast) | • Image brightness |
| 48. The CT number that is at the center of the window width is represented by __________ | • Window level |
| 49. What is a Pitch means in CT scan? | • The amount of anatomy covered during a particular scan. |
| 50. Learn the formula and math application of a pitch. | • Pitch=length of table travel per rotation/slice thickness. |
| 52. What is MPR stands for in CT Scan? | • Multiplanar reconstruction |
| 53. The preliminary image during a CT procedure is called____________ (3 different names) | • Scanogram • Scout • Topogram |
| 54. What is meant by scan FOV? | • Determines the area over which projection data is recollected • Determines the number of detectors collecting data for a particular scan. |
| 54. What is meant by scan FOV? 2nd part | • Should be large enough to cover all of the patient anatomy • If any part of the pt lies outside the scan FOV, out-of-field artifacts will occur |
| 55. What is meant by reconstruction FOV? | • Is the region of interest illustrated in the image and is reconstructed from the completer set of raw data. • Always equal to or smaller than the scan FOV |
| 56. Reconstruction FOV is also known as __________________ | • Display FOV |
| 57. Reconstruction FOV is always equal to or ____________ than the scan FOV | • smaller |
| 58. To magnify an image a ___________FOV must be used (Small / Large) | • Small |
| 59. To minimize noise in a CT image a _______FOV must be used (small/ Large) | • Large |
| 60. To achieve increased spatial resolution a ________FOV must be used. | • Small |
| FOV and Magnification: | • Small FOV – more magnification • Large FOV-less magnification (shows the image smaller) |
| FOV and Spatial Resolution | • Large FOV – reduces spatial resolution (because if you maintain the matrix size and increase the FOV means that you have bigger pixels, and the smaller the pixel the better the spatial resolution) • Small FOV – increases spatial resolution |
| Attenuation: | • Also known as differential absorption • Each voxel in the tissue slice is assigned a number by the computer that is proportional to the degree of x-ray attenuation of that tissue volume |
| Attenuation coefficient | • The degree of attenuation |
| ADC: | • Analog to digital converter. • Digitizing processing assigns an integer to each amplified electrical signal |
| Two Types of collimators used in CT and their functions: | • Pre-patient collimator: determines the slice thickness. Reduces pt dose and improves image quality • Post-patient collimator: further defines the slice thickness |
| Multislice CT Scanners: | • Image many slices simultaneously per tube rotation • Scanners are available to image 4 slices to 256 slices per x-ray tube rotation |
| Multislice CT Scanners Advantages | • Advantages: o Short acquisition time o Decreased amount of contrast media o Improved spatial resolution o Improved image quality |
| CT System Components: | • Gantry • Computer • Operator Console |
| Gantry Components | X-ray tube Detector array Collimator |
| X-ray tube: | Similar to radiographic x-ray tube. Higher heat capacity to accommodate increased exposure time |
| Detector array: | Solid state detectors. Composed of photodiode coupled with scintillation crystal material. Convert x-ray into light. Light is converted into electrical energy and then into digital signal. Affects patient dose and the efficiency of the CT unit. |
| Collimator | Pre-patient Post patient. Pre-patient collimator determines the slice thickness. Reduce patient dose and improves image quality. Post-patient collimator further defines the slice thickness. |
| Computer | Requires two types of highly sophisticated software. 1. For the operating system. 2. For the application. Operating system manages the hardware. |
| Computer 2nd part | Application software manages preprocessing, image reconstruction, and a wide variety of post processing operations. CT computer must have staggering speed and memory capacity. 262,144 mathematical calculations per slice for a 512 X512 matrix size. |
| Operator Console | o Operator console is the key point of interaction between the system operator and the imaging system. 1. Keyboard 2. Input device (mouse or trackball) 3. Monitor |
| Operator console functions | o Allows the technologists to control the parameters of the examination o View and manipulate the image. |
| Pitch: | o The amount of anatomy covered during a particular scan is determined by the pitch. o Pitch is a ratio reflecting the relationship between table speed and slice thickness. o Pitch = length of table travel per rotation/ slice thickness |
| Chapter 25 | |
| 1. Who invented Fluoroscopy | • MR. Thomas A. Edison in 1896. |
| 2. Unit for illumination level | • Lamberts and millilamberts |
| 3. Function of rods and cones | • They are responsible for the sensation of vision. When the light arrives at the retina, it is detected by the rods and cones. • The rods are sensitive to low light levels. • The cones are sensitive to intense light levels. |
| 4. Photopic vision | • The cones are used for daylight vision, called photopic vision. • Cones have better visual acuity and contrast perception than rods. |
| 5. Scotopic vision | • Rods are used for night vision, called scotopic vision. |
| 6. What is visual acuity? | • Ability to perceive small objects better. |
| 7. Factors that affect the brightness of fluoroscopic images | • Anatomy being examined • kVp • mA |
| 8. What is the mA used in fluoroscopy? | • Less than 5 mA |
| 9. Compare the patient dose of Fluoroscopy Vs Conventional x-ray | • Fluoroscopy: o Uses small kVp, more radiation to the pt, and exposure time is in seconds. • Conventional x-ray: o Use small kVp, less radiation to the pt, and exposure time is in milliseconds. |
| 10. Learn the components of an Image intensifier. (Very important) | • Metal housing • Vacuumed glass or metal enveloped • Input phosphor • Photocathode • Electrostatic focusing lenses • Anode • Output phosphor |
| 11. Learn the materials/chemicals used in each component (very important) | • Input phosphor: o Tightly packed cesium iodine crystals. • Photocathode: o Composed of cesium and antimony compounds. • Anode o The anode is the positive electrode. • Output phosphor: o Made of Zinc Cadmium Sulfide. |
| 12. Learn the functions of each component of the Image intensifier | • Input phosphor: o Converts the x-ray energy into visible light. • Photocathode: o Emits electrons when stimulated by the light from the input phosphor (photoemission)/ Converts light into electrons |
| 12. Learn the functions of each component of the Image intensifier . (Cont) | • PMT (photomultiplier): o Multiply the electrons with the help of dynode. • Electrostatic focusing lenses: o Focuses the electron stream on the output phosphor. |
| 12. Learn the functions of each component of the Image intensifier . (Cont) | • Anode o Attracts electrons emitted by the negative photocathode (a potential difference of 25 kV across the tube) o Allows the electrons to pass through to the output phosphor. |
| 12. Learn the functions of each component of the Image intensifier . (Cont) | • Output phosphor: o Interaction with the electron stream produces light photons of much greater intensity than was produced at the input phosphor. / Converts electrons into visible light. |
| 12. Learn the functions of each component of the Image intensifier . (Cont) | • Focusing lens: converts light into CCD • CCD (Monitor Camera): to capture the real-time x-rays and display on the monitor screen. |
| Flux gain: | Is the ration of the number of light photons at the output phosphor to the number of x-rays at the input phosphor. Flux gain=(number of output lightphotons)/(number of input x-ray photons) |
| • Minification Gain: | Is the ratio of the square of the diameter of the input phosphor to the square of the diameter of the output phosphor. Minification gain=(D1/( D2))2 D1: diameter of the input phosphor D2: diameter of the output phosphor |
| • Flux Gain: Formula | o Flux gain= number of output light photons/ number of input x-ray photon |
| • Total Brightness Gain: Formula | o Brightness Gain= minification gain X flux gain |
| • Minification gain: Formula | o D12/D22 |
| 15. What is veiling glare? | • Scattered radiation in the form of x-rays, electrons, and particularly light can reduce the contrast of image intensifier tubes through a process called veiling glare. |
| 15. What is veiling glare? (Cont) | • A veiling glare signal is produced behind a lead disc positioned on the input phosphor. |
| 16. What is vignetting? | • Vignetting is the reduction in brightness at the periphery of the image. • The portion of any image resulting from the periphery of the input phosphor is unfocused and causes vignetting. |
| 17. What is a multi-field image intensifier? | • Is equipped with up to three different input phosphor sizes. • 25cm/17cm/12cm |
| 18. Which is the magnification mode (27 or 17 or12)? | o 12 (less field but more magnification) |
| 19. Advantages of magnification mode | • Better spatial resolution • Better contrast resolution |
| 20. Which mode gives you more coverage of the anatomy? | • The bigger number covers more field |
| 21. What is focal point? | |
| 22. What happens to the focal point as you change the mode from big(25) to small(12)? | • THE ELECTRON FOCAL POINT MOVES AWAY FROM THE OUTPUT PHOSPHOR • VOLTAGE ON THE ELECTROSTATIC FOCUSING LENSES INCREASES • INCREASE MAGNIFICATION |
| 23. How can you calculate the magnification factor of a multi-field image intensifier? | • The magnification factors is directly proportional to the ratio of the diameters. • MF = Initial diameter of the input phosphor New diameter of the input phosphor |
| 24. What is the disadvantage of the magnification mode? | • Higher pt dose |
| 25. How the light from the output phosphor is converted to a visible image on the monitor? | • The TV camera tube converts the light image from the output phosphor into an electrical signal |
| 26. What is the function of television camera tube? Name the two common models. | • Converts the light image from the output phosphor into an electrical signal. • 2 models: o Vidicon tube o Plumbicon tube |
| 27. Learn the components of the television camera tube | • Cathode • Electron gun • Control grids • Electrostatic grids • Steering coils • Target assembly: o Face plate (window) o Signal plate o Target plate |
| 28. Learn how each component of the television camera tube work | • Electron gun: o Consists of heated filament that supplies constant electron current by thermionic emission. • Control grids: o Converts the electrons into electron beam • Electrostatic grids: |
| 28. Learn how each component of the television camera tube work (Cont) | o Accelerate and focus the electron beam • Steering coils: o Controlling the size and position of the electron beam. |
| 29. Learn the three layers of the target assembly and their functions | • Face plate (window) o Thin part of the glass envelope • Signal plate o A thin layer of metal or graphite coated on the inside of the window. It conducts the video signal out of the tube into the external video circuit. |
| 29. Learn the three layers of the target assembly and their functions (Cont) | • Target plate o A photoconductive layer of antimony trisulfide applied to the inside of the signal plate. |
| 30. What are the chemicals used in signal plate and target plate. | • Signal plate o A thin layer of metal or graphite • Target plate o antimony trisulfide. |
| 31. Which is the photoconductive layer of the television camera tube? | • Target plate |
| 32. What are the two systems used to connect the television camera tube to the image intensifier? | • Fiber optic system • Lens system |
| 33. Learn the advantage and disadvantages of each system | • Fiber optic system: o Advantages: • Use a bundle of fiber optics (thousands of thin glass fiber) • The simplest method • Compact • Easy to move the image intensifier tower • Can withstand rough handling. |
| 33. Learn the advantage and disadvantages of each system (Cont) | o Disadvantages • Cannot accommodate auxiliary imaging devices such as cine or photo spot cameras. • Lens system o Advantages • Allows the fluoroscopist to view the image while it is being recorded. |
| 33. Learn the advantage and disadvantages of each system (Cont) | o Disadvantages • Results in a much larger assembly • To be handled with care • Lens and mirror should remain precisely adjusted • Malposition results in blurred images |
| 34. How many lines per frames are used in CRT monitors used in regular fluoroscopy? | • Standard broadcast and closed circuit television have 525 lines per frame. |
| 35. Learn the function of a photo spot camera. What is the advantage of using it? | • SIMILAR TO MOVIE CAMARA • RECEIVES IMAGES FROM THE OUTPUTPHOSPHOR • USES FILM SIZE OF 70 AND 105 MM • LARGE FILM FORMAT RESULTS BETTER IMAGE QUALITY BUT INCREASED PATIENT DOSE |
| 35. Learn the function of a photo spot camera. What is the advantage of using it? (Cont) | • PROVIDE ADEQUATE IMAGE QUALITY WITH OUT THE INTERRUPTION OF FLUOROSCOPY • PROVIDE UP TO 12 IMAGES PER SECOND • REQUIRES LESS PATIENT EXPOSURE THAN CASSETTE LOADED SYSTEM |
| Chapter 26 |