Save
Upgrade to remove ads
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.
Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't Know
Remaining cards (0)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

Rad QA Test 1

QuestionAnswer
Who is responsible for QA? Radiologist
Who is responsble for QC? phycisist
Quality assurance deals with ____ people
image interpretation that involves reconciling the patient's ultimate disease condition with the radiologist's diagnosis outcome analysisi
10 Step Quality Assurance Program1. Assign _______2. Delineate _____ of care3. Identify _____ of care4. Identify _______ that affect the aspects of care.5. Establish _____ of the scope of ______.6. Collect and organize ____. 1. responsiblity2. scope3. aspects4. outcomes5. limits, assessment6. data
7. Evaluate care when ______ are reached.8. Take action to _____ care.9. Assess and document ____.10. Communicate information to organization-wide ____ ___ ____ 7. outcomes8. improve9. actions10. quality assurance programs
All of these QA activites require attention from the ___ ___, but they are principally the responsibility of the _____. imaging team, radiologist
A program of QC is designed to ensure that the radiologist is provided with an optimal _____ produced through good _____ performance and resulting in minimal patient radiation _____. image, equipment, exposure
QControl begins the xray ___ ___ used to produce the image and continues evaluation of the image processing _____. QC concludes with a dedicated analysis of each image to identify _____ and _____ (along with their cause) and to minimize _______. imaging systems, facitlities.deficiencies and artifacts, reexamination
Each new peice of radiologic equipment should be ____ tested before it is applied clinically. acceptance
Each acceptance test must be done by someone other than the ___ ____ because it is designed to show that the equipment is performing within the _______ _______ and is producing and acceptable patient radiation ___. manufacturer;s respresentative.manufacturer's specifications, dose
On most systems, _____ monitoring is satisfactory, uless a major compoenent such as an ___ ___ has been replaced annual, xray tube
An acceptable QC programs consists of 3 steps:_____ testingroutine ______ monitoring______ acceptance performance maintenance
The most important protection characteristic of a radiographic imaging system is ______ of the xray beam. State statutes require that general purpose radiographic units have a minimum total filtration of _________. filtration.2.5 mm Al
The xray field must conincide with the ____ ____ of the variable-aperture light-localizing collimator light field
Misalignment must not exceed ____ of the source-to-image receptor distance. (SID) 2%
With PBL collimator, the xray beam must not be large that the ___ ___, except in the ____ mode. image receptor,override
The spatial resolution of a imaging system is determined principally by the __ __ __ of the xray tube. when new equipment is installed it must be measured. focal spot size
3 tools for measurement of a focal spot size:1. ___ camera2. the ___ pattern3. ___ camera 1. pinhole cameral2. the star pattern3. slit camera
The standard for measurement of the effective focal spot size is the ___ camera. slit
The kVp should be evaluated annually or whenever __ ___ ___ components have changed significantly high voltage generator
A variation in kVp of approximately ___ is necessary to affect image optical density and radiographic contrast. 4%
The measured kVp should be within ___of the indicated kVp. 10%
Exposure timer accuracy whould be within ___ of the indicated time for exposure times greater than ___. 5%, 10ms
Accuracy of __ is acceptable for exposure times of 10ms or less 20%
AEC systems are evaluated by exposing an image recptor through various thicknesses of ____ or ____ aluminum or acrylic
regardless of the material thickness and the abosolute exposure time, the ___ ____ of the processed image should be constant. optical density
insertion of a ___ ___ allows one to adequately assess the functining of the ____ ____. If the phototimer fails, the backup timer should terminate the exposure at __ or ___, whichever occurs first. 6s or 600mAs
Many ____ of mA and exposure time produce the same mAs value. combinations
the ability of a radiographic unit to produce a constant radiation output for vaious combinations of mA and exposure time exposure linearity
exposure linearity must be within __ for adjacent mA stations 10%
sequential radiation exposures should be repoducible to within __. +5%
Screen film contact should be evaluated once or twice a year. Done by the __ __ tool wire mesh
viewbox illumination is checked with a ____, which meausre light ntensity at several areas of the illuminator. Instensity should be at least ______ and should not vary by more than ___ over. 1500 cd/m2, +10%
The entrance skin exposure for an adult averages __________ . For interventional procedures, a dose is ______. 3 of 5 R/min, 100 rad
federal law requires that ESE rates do not exeed ___. Interventiional up to _____. Unlimited for recorded fluoroscopy. 10 R/min20 R/min
Measurements are made with a ____. Lucite aliminum, copper and lead filters are used todetermine the adequacy of any ____ system. ABS
An ESE of appox ____ may be assumed for a cassette spot film. 200 mR
Spot films uses _____ images. They are recorded from the ___ ____ of an image intensifier tube. An ESE of ___ may be assumed for a photofluorospot photofluorospotoutput phosphur100 mR
Patient exposure should be measured for the most ___ type of tomographic examination. frequent
With incrementing from one tomographic section to the next, the section level should be accurate to within ___. Constancy of ___ from one QC evaluation to the next should be achieved. +2mm+1mm
refers to the routine and special procedures developed to ensure that the final product is of consistently high quality. QC
double capacity processors can handle up to ____ films per hour 500
A _______ strip should be passed through the processor, and fog, speed, and contrast should be measured and recorded properly. sensitometric
For any radiographic technique, the output radiation ___ should be contant from one exposure to another. This is check by making repeated radiation expostures through through the same technique and observing the average ____ in radiation intensity. The intensityvariation
The variation in xray intensity should not exceed __ 5%
a process that monitors and evaluates the quality of the care and services provided to patients within a health care facility. quality management
terms used to encompass all activites related to quality management, including:utililization and risk management, infection control, surgical case review, medicatin usage eval, healt record review, blood usage review, pharmacy and therapeutic rev, casem quality assurance and qualaity assessment
JCAHO's 10 step process was created to help organizations make the transition from quality assurance to ___ ___ ___ continuous quality improvement
Have our activities acheived desired outcomes? efficacy
is the activity relavant to our clinical needs? appropriatenes
is intervention available and accessible availability
do care and resources acheive the desired outcome effictiveness
how are activites or interventions coordinated among providers and over time? continuity
what is the degree to which associated risks are minimized safety
how do we maximize the relationsip btween outcomes and resources? efficiency
what is the degree of patient involvement in care decisions and the degree of sensitivity shown for individual needs respect and caring
the data for quality management should be reported to the ___ __ and the hospitals __ _ ___. medical staff, board of trustees
JCAHO uses evaluation or clinical outcomes as part of the ____ process accreditation
Created by: pspillman17
 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards