Save
Upgrade to remove ads
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.
Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't Know
Remaining cards (0)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

RAD 122 Unit 3 Test

QuestionAnswer
Why does an underexposure and overexposure still yield images? - wide dynamic range, - automatic rescaling, normalization (displays values of interest to specific lvls, produces imgs w/ consistent brightness over a wide exposure range) -LUT (displays img w/ appropriate contrast)
Wide range of exposure from as low as ___ mR up to ____ mR .01 mR to 100 mR
____ exposure to receptor produces most acceptable images(approx. 200 speed class) in s# 1mR (0.001 R)
advantages of digital over film-screen are? preprocessing (ability to manipulate img before display) post-processing (ability to manipulate img after display)
Carestream, Every change of ____ EI represents a change in actual exposure by a factor of? 300 EI, by a factor of 2
Agfa, center of scale is ___ LgM. below ___ is considered underexposed while above ____ is overexposed. center at 2.2 LgM. below 1.9 is underexposed and above 2.5 is overexposed.
Inverse proportional scales. Uses S#. Fuji and Konica
detector elements in an active matrix array stop functioning, which means that no data from that area are included in the final img. appear as white specks mimicing pathologies, can be fixed w/ flat fielding. rough handling can cause this damage Dropped pixels
processing error that can happen when switching from fluoro to regular exam on FPD. The old latent image is still on IR. image lag
wireless DR IR's suffers from ____ _____ from devices that draw lrg amt of electronic power or from cell phones, laptops, and the like. Electronic interference
digital system uses a preprocessing manipulation known as _____, which is a software correction to equalize the response of each pixel to make the image more uniform. flat fielding
corrects defects by giving an average of surrouding pixel to correct defective pixels signal interpolation
corrects defects by neutralizing any residue charges offset voltage
ranges of values/receptor exposure intensities that can be displayed by the digital detector / imaging system dynamic range
ellimates signals outside collimated margins. if margins not detected, extreneous data not included histogram automatic exposure field edge detection
uses shape of image signal distribution to determine relevant information within the exposure field and exposure indicator value histogram analysis
mapping grayscale "values of interest" to achieve specific display levels. provides images that have uniform display brightness over wide exposure range automatic rescaling
system needs to see a defined collimated border of a minimum of 3 sides. this type of error is called collimation error
each detector element has a certain amount of _________ associated with it electronic noise
caused by the collection element not sufficiently sending all data to the computer for processing, resulting in incomplete image formation. incomplete charge transfer
exposure are taken in too rapid a sequence, resulting in not enough time for each previous exposure to transfer the entire signal, creating an artifact known as "ghosting" electronic memory artifact
in lrg detector arrays, several amplifiers are used that can vary in gain from one to the next. to correct in gain btwen detector elements (tiling), a process known as flat fielding is used. Uneven amplification
happens through the back of the cassette that can either create noise in the image or may create an image of the internal circuitry found in the back of the cassette Back scatter
Gross overexposure is when the mAs is ____% over the # suppose to use. this creates a lost of contrast too many photons. 400%
increase in window width will ____ contrast increase contrast
changes to window level will cause a change in ____ brightness
we have no control over which source type of radiation? Natural and internal
this source of radiation is emitted by the sun and stars cosmic radiation
this source of radiation comes from naturally occurring radioactive materials that are located throughout the earth's crust. terrestrial radiation
this source of radiation refers to the small amts of radioactive minerals normally found in the body, ingested from the small amount present in food and inhaled as air-borne particles internal radiation
CR image plates contain a ___ and will therefore "store" a charge produced by background radiation. storage phosphor
normal background radiation is anywhere from ____uR/day to 80uR/day or _____ to 3.5 mSv per year but can be more than 50 mSv/yr 40uR/day to 80uR/day, 1.5mSv to 3.5mSv
CR imaging plates unused for more than ___ hours should be erased 48 hours
are constantly refreshed and do not require erasure between long period without use. They do not storage phosphor like CR FPD
_____ radiation is more intense than background radiation. this should be controlled for optimum image quality. scatter radiation
T/F lead strips is like a shield and cause a histogram analysis error. True
T/F using a grid decreases the amount of scatter and decreases patient dose. False a grid has no affect on patient dose
T/F does not matter how big a part is needed between multiple field exposures. False. the larger the part, the more spaces needed between multiple-field exposures. collimated border is needed for proper histogram analysis
over exposure causes more or less scatter production? thus increasing unwanted _____ and change of improper _______ more scatter, increasing unwanted receptor exposure and chance of improper exposure field detection
is a double set of grid lines caused by the placement of a grid cassette in a bucky. this artifact also can occur in CR in the range of 85 to 100 lines per inch is used, and the rid lines are parallel to the CR reader scan lines. Moire effect or zebra pattern artifact
_______ = approximately 1/2 the sampling frequency Nyquist frequency. avoid using grid frequencies near nyquist to avoid moire/alaising
can occur if not enough samples are taken (less than twice per cycle), and the rep of the orginal signal will not be accurate after computer processing. Also can cause moire pattern artifact Aliasing
T/F the number of conversions in CR creates a loss of information True
T/F no light is lost during the light-to-digital conversion bc of the spreading out of light photons False
bc there is a small distance between the _____ surface and the _______ of the photomultiplier, some light spreads out there as well, resulting in loss of information. phosphor plate surface and the photosensitive diode of the photomultiplier
T/F the longer the electrons are stored, the more energy they lose True
T/F if enough energy was lost, some lower-energy electrons are not stimulated enough to escape and information is lost True
When laser stimulates electrons, some lower-energy electrons escape the ____ layer active
T/F Indirect and direct radiography lose more signal to light spread than conventional radiography. False they lose less signal to light spread than conventional radiography
intro in 1942, eliminated the need for the tech to set mAs bc this determines the exposure time and therefore total mAs. Tech still has to set kVp and correct chamber and density AEC
With AEC, the only way to increase or decrease receptor exposure is to change the ______ density selection
With AEC, if the body part is not centered over the selected chanbers then the image could be under/over exposed, receptor exposure will be affected resulting in a repeat
With AEC, receptor exposure is affected by - centering of the part - detector selection - density control setting
one of two types of AEC, where it measures quantity of radiation coming out of patient photo timer behind receptor
one of two types of AEC, where it terminates exposure when IR has received required intensity ionizing chambers
XR exit patient and interacts w/ flourescent screen which en converts xr to light, light photons enter photocathode where they are converted to electrons w/ same intensity as photons that created them photodiode akaphototimer, or photomultiplier tube
T/F in AEC using ionization chambers, no phototube is needed bc it is replaced by a parallel plate ion chamber btwen the patient and the IR. True
is the shortest exposure time the system can produce. hence why we dont use AEC w/ small anatomy Minimum response time
is the maximum length of time the exposure continues using an AEC. backup timer occurs at 5-6 seconds or at 600 mAs
adjustments allow the radiographer to adjust the amount of preset radiation detection sensitivity density control
The standard configuration of detectors is a triad of rectangular shaped demarcated lines on the tabletop over vertical bucky. detector configuration
T/F the presence of orthopedic metal prothesis, which cause radiopaque artifacts can cause an underexposed image false it causes an overexposed image
In digital imaging display, which is the weakest link the monitor bc the room is worth millions vs a $50 dollar screen
total intensity from the source. measured in lumens luminous flux
intensity of light incident on a surface. unit is 1 foot candle, or 1 lumen per sq/m illuminance
1 foot candle = 10.8 lux
property of source of light measured in lumens or candela luminance intensity
similar to luminance intensity / another measure of brightness. measured in candela per sq/m or nit luminance
the study of light photometry
measures light photometer
Primary display monitors should have a luminance greater than ____ nit (50 foot-lamberts) 170 nit
Secondary display monitors should have a luminance greater than ____ nit (30 foot-lamberts) 100 nit
light-emitting, curved face, scanned electron beam, veilling glare, spot pixels(phosphors) CRT
light-modulating, flat face, active matrix address (TFT), pixel cross talk distortion, square pixel LCD
refers to images displayed on computer monitor soft copy
ration of width of display to the height of the display aspect ratio
how many times each second the monitor rewrites the image on display refresh rate
is an electronic vacuum tube used for displaying electrical data in the form of images or videos on the screen. it is used in electronic devices such as tvs, computer monitors, radar, and oscilloscope tubes. Cathode ray tube (CRT)
a component of CRT, is used for generating, controlling and focusing the beam of electrons enclosed in a vacuum tube. internally consists of five components, heater, cathode, control grid, accelerating anode, and focusng anode electron gun
a component of CRT, the horizontal deflection plates deflects the electrons horizontally horizontal deflection plates
a component of CRT, deflects electrons vertically vertical deflection plates
T/F both the horizontal and vertical deflection plates control the path of electron beam so that it can be directed towards a specified position on the phosphor coated screen TRue
a component of CRT, is a transparent screen coated on one side with a phosphor that grows brightly when exposed to cathode rays fluorescent screen
T/F to offset high noise levels, the CRT monitor must have a low SNR False must have a high SNR
type of glare that occurs when ambient or room lighting reflects off of the surface of the monitor or display screen reflective glare
CRT monitor screen have an internal glare called _____. electron backscatter, lateral leakage of light from the phosphor crystals within the phosphor layer. veiling glare
T/F a disadvantage of LCD is that it becomes defocused over time bc electron beam is used to create the image False, an advantage of LCD is that it does not defocus over time bc no electron beam is used to create the image.
T/F LCD has better grayscale definition than CRT True
T/F LCD increases the effect of ambient light on image contrast (less contrast resolution) False, LCD better reduce the effect of ambient light on image contrast (greater contrast resolution)
T/F LCD produces a higher lvl of luminance than CRT's (up to 700 nit) and less noise TRue
T/F Disadvantages of LCD include greater viewing angle False, usually 40 degrees either side of a line that is perpendicular to the center of the faceplate.
T/F an advantage of LCD is that it is not subjected to dead pixels False, it is subjected to dead pixels.
a viewing condition, primary workstations are used for official interpretation of medical images and must follow established guidelines to include maximum lum lvls of at least ____ candelas per meter squared. 171 candelas per meter squared
_____ workstations are used by technologist to view a digital image. workstation is often in a brightly lit work area while the radiologist high res monitors are in a dimly lit reading room. secondary workstations
universal standard for exchaning medical imgs among modality, workstations and archives. contains pt name, iden info, date of procedure, ordering MD and location digital imaging and communications in medicine DICOM
RIS is under Epic
data base that handles the workflow of the system by moving images to storage for viewing at workstation then back to the archive PACS image manager
T/F PACS image manager populates information to the electronic medical record (EMR) TRue
Government standard where it uses 7 layers of encryption Health Level 7 (HL7)
in a clinical setting EMR is Epic
The displayed value on the processed digital image that indicates the level of exposure received on the IR. exposure indicator value
The evidence of motion on a radiographic image is seen as: Image blur
Increase grid ratio, AOFC, has what effect on the img? a. increase RE and increase img contrast b. increase RE and decrease img contrast c. decrease RE and increase img contrast d. decrease RE and decrease img contrast C. decrease Receptor exposure and increase image contrast
Foreshortening occurs when The part is angled in relationship to the IR
Elongation occurs when: a. The part is angled in relationship with the IR b. Tube is off-centered or angled c. IR is angled and not parallel with part d. Both b and c e. All of the above D. both b and c
Defines differences in pixel brightness values between various areas of the image: image contrast
Increasing tube filtration results in an x-ray beam with ____________ quantity and ______________ energy. Less; higher
The anode heel effect: due to x-ray absorption at the surface of the anode
All other factors remain constant, an increase in exposure time has what effect on receptor exposure? increase receptor exposure
An advantage of PACS is that it allows images to be: sent to several workstations and monitors simultaneously
Increasing the atomic number of tissue (Z#) has what effect on x-ray photon transmission? decrease
on the flash light looking diagram, it points to the the end of the flighlight where light comes out. Fluorescent screen
The set of computer standards used to permit a wide range of digital imaging systems to ensure that images taken at one facility are able to be viewed correctly at another facility is termed: DICOM
Increasing the size of the pixel has what effect on spatial resolution? decrease
This post processing feature allows the technologist to adjust the radiographic image brightness on the processed images. window level
If the distance from the source is decreased by a factor of 2 (one half), the mAs must be ______to maintain the same receptor exposure. decreased by a factor of 4
When comparing viewing monitors, increased signal-to-noise ratios results in: sharper images
The configuration formed by a scanning electron gun in a CRT is termed a/n: raster pattern
In order to eliminate reflective glare when using an image viewing device, the technologist or radiologist should: view the image at 90 degrees, rather than from an angle
T/F The technologist should manually adjust all radiographic images (window level/window center) at the QC station prior to sending to the radiologist for viewing. False
The exposure rate of the x-ray beam is BEST controlled by: milliamperage
The four steps of creating an image with computed radiography are: exposure, stimulate, read, and erase
This grid error results when the central ray is not aligned from side to side with the center of the focused grid. Off-center
The type of display monitor that contains an electron gun and accelerating grids: cathode ray tube
Primary display monitors should have a luminance greater than: 170 nit
Created by: Btran573
 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards