Digital Radiography.
Bushong ch 26
| Question | Answer |
|---|---|
| DR is More efficient in 3 | time, space and personnel than screen-film radiography |
| CR uses | photostimulable phosphor |
| SPR uses | photodiode (not used) |
| Indirect DR uses 2 | TFT and charge couple devices |
| Direct DR uses | TFT |
| The 3 c's in IDing DR Systems | Capture element Coupling element Collection element |
| In CR, image is captured where | Photostimulable phosphor |
| In DR, image is captured where, 3 | Cesium iodide (CsI) Gadolinium oxysulfide (GdOS) Amorphous selenium (a-Se) |
| Coupling element does what and with what3 | Couples x-ray signal to the collector Lens Fiber optic assembly Contact layer a-Se |
| Collection element | collects electrons |
| 3 things that collect electrons | Photodiode, CCD, TFT |
| Photodiode and CCD is | a light sensative device |
| TFT does what | charge sensative device collects electrons |
| Advantages of CCD 3 | Sensativity, Dynamic Range, Size |
| Sensativity 2 | Can detect and respond to very low levels of light Ideal for low patient dose |
| Dynamic range | Responds to a wide range of light |
| Size | Very small and highly adaptable to DR Measures approx. 1-2cm and the pixel size is 100 x 100µm |
| CCD/CsI 2 | Used in indirect DR Converts x-rays to light then to electric signals |
| CsI works how 2 | Scintillation light from the CsI phosphor is efficiently transmitted through fiber optic bundles to the CCD array Results in high x-ray capture efficiency and good spatial resolultion – up to 5 lp/mm |
| What kind of element is used in indirect DR | Cesium Iodide/Amorphous Silicon Collection |
| Cesium Iodide/Amorphous Silicon Collection Element | Captures x-ray and transmits the collected light to element (CsI a-Se) CsI a-Se is a liquid that is painted onto a support surface |
| Low pt dose b/c of | High photoelectric capture |
| Fill factor | Due to the other components on the face of the pixel, the fill factor is approx. 80% |
| As a result of the small pixel size | spatial resolution is increased but patient dose is increased |
| Whats Similar to CeI a-Si | Gadolinium Oxysulfide Collection Element |
| Unlike screen film.... | increasing the amount (thickness) of Gd OS increases the speed of the system but without a loss of spatial resolution. |
| Whats used for direct DR | Amorphous Selenium; No scintillation phosphor |
| The imaging forming x-ray beam interacts | directly w/ the a-Se, producing a charged pair |
| The a-Se is 2 | both the capture and coupling elements |
| The IR for a direct dr is | an active matrix array of TFT |