Save
Busy. Please wait.
Log in with Clever
or

show password
Forgot Password?

Don't have an account?  Sign up 
Sign up using Clever
or

Username is available taken
show password

Your email address is only used to allow you to reset your password. See our Privacy Policy and Terms of Service.


Already a StudyStack user? Log In

Reset Password
Enter the associated with your account, and we'll email you a link to reset your password.

Question

Why was AEC originally developed?
click to flip
focusNode
Didn't know it?
click below
 
Knew it?
click below
Don't know

Question

Which AED works by using a gas-filled chamber?
Remaining cards (94)
Know
0:00
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

PRI-II

AEC/APR

QuestionAnswer
Why was AEC originally developed? To provide consistency in exposure and images
Which AED works by using a gas-filled chamber? Ionization chamber system
Where is the sensor located in an AED with a gas filled chamber? Between the patient and the IR
Where is the sensor located in an AED that converts light to electrical signals? Behind the IR
AED that convert light to electrical signals are considered? Exit type
AED that use gas filled chambers are considered what type of AED? Entrance type
AEC controls what factor? Time only
Which AED works by converting photons to light and then electrical signals? Phototimers
Which AED is considered the first generation device? Phototimers
What would happen if film screen radiography exam required an exposure time faster than the minimum response time of the AEC sensors? Image would be overexposed, because it would continue to to accept photons until the MRT was reached
With film screen radiography which factor would be adjusted to manipulate the contrast? KVP
In digital radiography which factor would be adjusted to manipulate the contrast? None, KVP does provide change, but it is not consistent. It is mainly controlled by computer-processing
What is the appropriate setting for backup time/mAs? 150%-200% of the expected exposure
What is the purpose of backup time/mAs? Safety feature in the event the equipment malfunctions of the tech does not use the appropriate technical factors
When using film screen, what happens when an outside detector is selected for an AP thoracic spine? The system shuts off prematurely and the image appears underexposed
When doing a film screen study using AEC, what will happen to the density in the area of interest when changing from -1 to +1 density? Density is increased by 50%
The typical minimum response time on modern x-ray equipment: 1 ms
What is the most common cause for repeating images done with film screen and AEC: Patient centering
What is the optimal KVP starting point for small extremities? 60 kVp
How does increased kVp affect density (film screen) or exposure (digital imaging)? There is no change to density
What are the basic functions of AEC: -Radiation is transmitted thru pt. -Exit radiation is converted to electrical signal -Once predetermined amt of radiation is reached exposure is terminated
The predetermined amount of radiation reaching the detector is known as being _________ and who sets it? Calibrated Service personnel
Who determines the standards for calibration of machine? Department
The two basic types of AEC systems: -Phototimers -Ionization chamber
Most common type of AEC system used today? Ionization chamber
Most reliable type of AEC system is ________: Ionization chamber
Type of AEC system least prone to failure: Ionization chamber
Phototimer AEC systems use ________? PMT or PD
Which AEC system uses a fluorescent screen and a device that converts the light to electricity? Phototimers
Phototimers are also known as __________ type devices: Exit
_______ detectors are light paddles coated in fluorescent material: Phototimers or Exit-type devices
The fluorescent material used in phototimer systems is comprised of what two materials? Lucite with phosphors
What component of the phototimer (Exit type) system converts light to electrical signals: Photomultiplier or photodiode
Basic function of phototimer systems: -Radiation interacts with paddles & produces light -Light is transmitted to remote PMT or PD -Light is converted to electric current & then amplified to electrical charge -Timer is tripped and exposure stops -
The electrical charges are __________ to the radiation received by light paddles: Proportional
All AEC systems use ________, they are also called what other names: Detectors Sensors Cells
The AEC devices that measure the amount of radiation transmitted: Sensors or cells
Basic function of ionization chamber system: -Ionization chamber is exposed to exit radiation -Air in chamber is ionized creating an electrical charge -Charge travels along wire to timer circuit -Timer is tripped and exposure stops when calibrated limit is reached
Electrical charge is proportional to the radiation received in the ______ _______: Ion chamber
AEC affects what factor reaching the IR: Quantity
kVp controls what three factors: -Contrast -Penetration -Patient Exposure
A decrease mA can be used for what two reasons? -Increase time -Change FSS
When kVp is decreased what will happen to time, mAs & patient exposure? All increase
When kVp increases all other factors ________? Decrease
Adequate kVp must be selected to ______ & _______? -Penetrate the anatomical part -Produce adequate contrast
Increased mA does what to time? Decreases
Increased time ________ mA? decreases
The shortest exposure time a system can produce is known as: Minimum response time
Minimum response time is used for: -Pediatrics -Motion control
If mA is set too high to reach the minimum exposure time how will this affect film & digital? -Film, will be too dark -Digital, EI #s out of range and pt. dose increased
The maximum exposure time allowed during an AEC examination is known as _______: Backup time
Phillips uses density controls of (-) or (+) 1-4, each increment is roughly? 10%-15%
If density controls are being used routinely to achieve adequate exposure, what needs to happen? Equipment needs to be recalibrated
If a machine is calibrated for a 400 speed and a 200 screen speed is used, how will it affect the image and why? It will take longer for the sensors to fill so the image will be overexposed
If more than one sensor/cell is used the measure of radiation and electrical signal is ________: Averaged
Improper patient centering can result in: Underexposure or overexposure
The detector size is set by ______: Manufacture
Over collimation can result in _______ _______: Prolonged exposure
If AEC is calibrated for a slow IR and a fast IR is used the result will be __________: Overexposure
If AEC is calibrated for a fast IR and a slow IR is used the result will be _________: Underexposure
Are APR and AEC the same? No
APR stands for: Anatomically programmed radiography
APR can include both _____ and _________ techniques: AEC & Manual
APR can or can not be alter once selected: Can be adjusted by tech once a selection has been made
_____ is a preprogramed set of factors: APR
APR operates the _____ on all equipment: Same
_____ is controlled by integrated circuit or computer chip that has been programmed with exposure factors for different exposures and projections: APR
________ are established guidelines used by the radiographer to select standardized manual or AEC exposure factors for each type of radiographic examination: Exposure technique chart
Exposure technique charts are used to provide what two purposes? Consistent quality images & reduce repeats
Factors on the expsoure technique chart must include: -kVp -mAs (possibly seperated) -Type of IR -Grid -SID -Possibly FSS
Standardization of technique charts must be done within _________: The dept.
For technique charts to work efficiently, what two things must occur? -Equipment must be functioning properly -Proper measurements must be taken
Measurements should be taken from which portion of the anatomical structure being examined? -The midpoint or -The thickest portion
What are the two types of technique charts? -Variable kVp/Fixed mAs -Fixed kVp/Variable mAs
Baseline kVp should be increased by ____ for every _____: -2 -1 cm
What are the disadvantages of a variable kVp/fixed mAs technique chart? -Changing kVp also affect contrast -Less accuracy at extreme ends -Part penetration
Variable kVp/Fixed mAs charts are effective with which two types of examinations? -Pedis -Small extremities
At what kVp range is more effective to change kVp rather than mAs? Lower level
Advantages of a fixed kVp/variable mAs technique chart: -Charts can be completed in patient groups vs individual 1 cm adjustments -every 4-5 cm increase in thickness mAs in adjusted by a factor of two
Optimal kVp ranges for the following anatomical parts; Chest, Neck, Abdomen, Sm. extremities, Lg. extremities, shoulder girdle, Hip/pelvis: Chest=100-130, Neck=75-80, Abdomen=70-80, Small ext.=50-60,Large ext.=65-75,Shoulder girdle=70-80, Hip/pelvis=75-85
_____ ____ are not standardized, they are meant to provide a common range of values for different anatomical areas: Optimal kVp ranges
Advantages to using fixed kVp/variable mAs charts: -Easier to use -More consistency -Part penetration is mor accurate -More uniform contrast -Increased accuracy at the extreme measurement ends
Optimal kVp ranges for cervical, thoracic, lumbar/sacrum, coccyx, sternum/SC joints, Skull/headwork, ribs: C-spine=70-80, T-spine=75-85, L-spine/Sacrum:AP/Obl=75-80, Lat=85-95, Spot=95-100, Coccyx=75-85, Sternum/SC joints: Obl=60-70, Lat=70-75, Skull/headwork=70-80, Ribs: Above=65-70, below=70-80
Conventional radiography demonstrates ______? Superimposition of structures
Conventional tomography demonstrates _________? Unsuperimposed structure, by blurring structures above and below a specific plane
Main uses of tomography? Nephrotomography study to visualize kidney stones
How does linear tomography work? The IR and tube move in opposite directions to blur structure above and below the specific plane of interest
The principal advantage of tomography is ______? Improved contrast resolution
A larger tomographic angle will result in _______? More blurring of the objects above and below the fulcrum point
A smaller tomographic angle will result in ________? Less blurring of the objects above and below the fulcrum point
The larger the tomo angle the ________ the tomographic slice: Thinner
The tomo angle and the tomo slice have a ________ relationship: Inverse
The smaller the tomo angle the ________ the tomo slice: Thicker
What is zonography? When tomo angle of less than 10 degrees is used, it is utilized when subject contrst of the object is so low that a thin slice would result in poor image
Zonography is used what what type of examinations? Chest and renal examinations
Created by: akhernandez19
 

 



Voices

Use these flashcards to help memorize information. Look at the large card and try to recall what is on the other side. Then click the card to flip it. If you knew the answer, click the green Know box. Otherwise, click the red Don't know box.

When you've placed seven or more cards in the Don't know box, click "retry" to try those cards again.

If you've accidentally put the card in the wrong box, just click on the card to take it out of the box.

You can also use your keyboard to move the cards as follows:

If you are logged in to your account, this website will remember which cards you know and don't know so that they are in the same box the next time you log in.

When you need a break, try one of the other activities listed below the flashcards like Matching, Snowman, or Hungry Bug. Although it may feel like you're playing a game, your brain is still making more connections with the information to help you out.

To see how well you know the information, try the Quiz or Test activity.

Pass complete!
"Know" box contains:
Time elapsed:
Retries:
restart all cards