Question | Answer |
Which condition is needed to produce an x ray? | 1. a source of electrons 2. a vacuum 3. a large potential difference |
The function of the cathode is to produce | a thermionic cloud |
Each of the following is a part of the cathode addembly except the | stator |
Each of the following is a material of choice for the filament except | Copper |
Filament materials are chosen for their | high melting points |
Which of the following would affect the production of a thermionic cloud? | filament diameter, length, and temperature |
A filament may break due to | vaporization |
Driving the thermionic cloud from the cathode to the anode requires | a large potential difference |
The focusing cup has a/an ____ charge | negative |
X ray photon production occurs at the | anode |
Tungsten is the metal of choice for the source of x ray photons because of its | high atomic number |
Pitting of the anode commonly occurs | with extended use |
Melting of the anode commonly occurs | when the stator fails and the rotor ceases to turn |
The exact point in the x ray unit where photons are created is | focal spot |
The physical area of the focal tract affected by the electron beam is the | actual focal spot |
The area of the focal spot that is projected out of the tube toward the object being radiographed is the | effected focal spot |
The actual focal spot is controlled by the | length of the filament |
The only portion of the anode assembly placed outside the vacuum of the envelope is the | stator |
The function of the stator and rotor assembly is to | rotate the anode |
The rotor is located ___ the stator and ____ the envelope | inside, inside |
Insulation of the high voltage components of the tube and absorption of the heat produced by x ray production is achieved through the use of | dielectric oil |
Fluoroscopy is a ___ radiograph examination | dynamic, static, diagnostic |
Fluoroscopy is commonly used for observation of | the heart, GI tract, and surgical procedure |
The fluoroscopic xray tube and image receptor are mounted on a | C Arm |
Fluoroscopic mA range is commonly ___ mA | 0.5 - 5.0 |
During fluoroscopy, the SOD cannot be less than __ inches | 20 |
During fluoroscopic image intensification, the primary x ray beam exits the patient and strikes the ___ of the image intensifier | input screen |
The input screen absorbs __ and emits __ | x ray photons, light photons |
The photocathode absords __ and emits __ | light photons, electrons |
Electrostatic lenses are used to accelerate and focus | electrons |
The output screen absorbs __ and emits __ | electrons, light photons |
The input screen is __ in shape | concave |
The shape of the input screen helps to control | distortion |
Photoemissive materials absorb __ and emit __ | light photons, electrons |
Electrostatic lenses are actually | charge electrodes |
The greater the voltage supplied to the electrostatic lenses, the __ the acceleration and the __ the focal point to the input screen | greater, closer |
The edge distortion problem in image intensification tubes called | vignetting |
The most common solution for quantum mottle is to | increase the fluoro tube mA |
The standard size cassette for spot filming is | 9x9 |
Digitization of the fluoroscopic image permits | storage on a CD, transfer via internet, hard copy printing via laser disc |