Using AEC, a 50 speed is mistakenly used instead of a 400 speed IR. What will occur?   insuffient optical density  
the use of an impulse or synchronous timer in a single phase will permit exposure times as short as   17 ms (1/60)  
timer circuits are located in the _________ circuits for safety reasons   primary  
what is the likely cause of an excessively long AEC exposure?   failure to properly center the x-ray tube  
thsi terminates an AEC exposure   Thyristor  
most common type of device in a modern 3 phase timer circuit is a silicon controlled rectifier. what is the smallest exposure that it is capble of?   1 millisecond (.001)  
where are the ioniztion type sensors located?   between the table and the grid in the bucky tray  
For AEC, falure to utilize the proper factors will result in   a radiograph possessing excessive or insufficient optical density  
a minus 2 setting in the density control setting is the equivalent of how much of a reduction in the mAs value?   50%  


   

 
 

 
 

 

 