Quiz yourself by thinking what should be in each of the black spaces below before clicking on it to display the answer.
        Help!  

Question
Answer
Which condition is needed to produce an x ray?   1. a source of electrons 2. a vacuum 3. a large potential difference  
🗑
The function of the cathode is to produce   a thermionic cloud  
🗑
Each of the following is a part of the cathode addembly except the   stator  
🗑
Each of the following is a material of choice for the filament except   Copper  
🗑
Filament materials are chosen for their   high melting points  
🗑
Which of the following would affect the production of a thermionic cloud?   filament diameter, length, and temperature  
🗑
A filament may break due to   vaporization  
🗑
Driving the thermionic cloud from the cathode to the anode requires   a large potential difference  
🗑
The focusing cup has a/an ____ charge   negative  
🗑
X ray photon production occurs at the   anode  
🗑
Tungsten is the metal of choice for the source of x ray photons because of its   high atomic number  
🗑
Pitting of the anode commonly occurs   with extended use  
🗑
Melting of the anode commonly occurs   when the stator fails and the rotor ceases to turn  
🗑
The exact point in the x ray unit where photons are created is   focal spot  
🗑
The physical area of the focal tract affected by the electron beam is the   actual focal spot  
🗑
The area of the focal spot that is projected out of the tube toward the object being radiographed is the   effected focal spot  
🗑
The actual focal spot is controlled by the   length of the filament  
🗑
The only portion of the anode assembly placed outside the vacuum of the envelope is the   stator  
🗑
The function of the stator and rotor assembly is to   rotate the anode  
🗑
The rotor is located ___ the stator and ____ the envelope   inside, inside  
🗑
Insulation of the high voltage components of the tube and absorption of the heat produced by x ray production is achieved through the use of   dielectric oil  
🗑
Fluoroscopy is a ___ radiograph examination   dynamic, static, diagnostic  
🗑
Fluoroscopy is commonly used for observation of   the heart, GI tract, and surgical procedure  
🗑
The fluoroscopic xray tube and image receptor are mounted on a   C Arm  
🗑
Fluoroscopic mA range is commonly ___ mA   0.5 - 5.0  
🗑
During fluoroscopy, the SOD cannot be less than __ inches   20  
🗑
During fluoroscopic image intensification, the primary x ray beam exits the patient and strikes the ___ of the image intensifier   input screen  
🗑
The input screen absorbs __ and emits __   x ray photons, light photons  
🗑
The photocathode absords __ and emits __   light photons, electrons  
🗑
Electrostatic lenses are used to accelerate and focus   electrons  
🗑
The output screen absorbs __ and emits __   electrons, light photons  
🗑
The input screen is __ in shape   concave  
🗑
The shape of the input screen helps to control   distortion  
🗑
Photoemissive materials absorb __ and emit __   light photons, electrons  
🗑
Electrostatic lenses are actually   charge electrodes  
🗑
The greater the voltage supplied to the electrostatic lenses, the __ the acceleration and the __ the focal point to the input screen   greater, closer  
🗑
The edge distortion problem in image intensification tubes called   vignetting  
🗑
The most common solution for quantum mottle is to   increase the fluoro tube mA  
🗑
The standard size cassette for spot filming is   9x9  
🗑
Digitization of the fluoroscopic image permits   storage on a CD, transfer via internet, hard copy printing via laser disc  
🗑


   

Review the information in the table. When you are ready to quiz yourself you can hide individual columns or the entire table. Then you can click on the empty cells to reveal the answer. Try to recall what will be displayed before clicking the empty cell.
 
To hide a column, click on the column name.
 
To hide the entire table, click on the "Hide All" button.
 
You may also shuffle the rows of the table by clicking on the "Shuffle" button.
 
Or sort by any of the columns using the down arrow next to any column heading.
If you know all the data on any row, you can temporarily remove it by tapping the trash can to the right of the row.

 
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how
Created by: wloucks08
Popular Radiology sets