Clinical - Supply & disposal

Quiz yourself by thinking what should be in each of the black spaces below before clicking on it to display the answer.
        Help!  

Question
Answer
The current anesthesia gas machine standard is:   ASTM F1850 (American Standard for Testing & Materials)  
🗑
SPDD stands for:   Supply, processing, delivery, disposal  
🗑
Supply: fresh gases get to the anesthesia machine by _____ and _____   pipeline, cylinders  
🗑
Supply: power gets to the anesthesia machine by _____ and _____   electricity, battery backup  
🗑
Processing: 6 ways the machine prepares gases before delivery to the patient are:   valves, flowmeters, oxygen flush, low pressure alarms, vaporizers, common gas outlet  
🗑
Delivery: 3 ways that delivered gases are controlled & monitored:   breathing circuits, ventilators, integral monitors  
🗑
How are waste gases disposed?   scavenge systems  
🗑
Cylinder pressure = ____ pressure system Pipeline pressure = ____ pressure system   cylinder = high pressure system pipeline = intermediate pressure system  
🗑
Low pressure system includes:   flowmeter tubes, vaporizers, check valves, common gas outlet  
🗑
Cylinder pressure is measured by:   Bourdon gauges  
🗑
What type of regulator is on a cylinder? What type of regulator is on a pipeline?   Cylinder = 1st stage regulator Pipeline = oxygen 2nd stage regulator  
🗑
1st stage regulator drops cylinder pressure to:   40 - 48 psi (intermediate pressure)  
🗑
The volume of oxygen & air in E cylinder is:   625 - 700 L  
🗑
The volume of nitrous oxide in E cylinder is:   1590 L  
🗑
Service pressure (PSI) of oxygen & air is:   1900  
🗑
Service pressure (PSI) of nitrous oxide is:   745  
🗑
What safety system is on cylinders? How is the safety disabled?   PISS (pin index safety system). Disabled if pins are missing or if more than 1 washer is used  
🗑
Valve stem on cylinder is fragile & contains one or more of the 3 safety features:   frangible disc (bursts w/ high pressure to release cylinder contents); safety relief (opens w/ high pressure to release cylinder content); fusible plug (alloy melts at 200 degrees F to release cylinder contents)  
🗑
Service pressure is:   the pressure in the cylinder  
🗑
For cylinder supply the hanger yoke does 3 things, which are:   orients the cylinder, gas tight seal, ensures unidirectional flow  
🗑
Which gas law pertains to volume & pressure?   Boyle's Law  
🗑
Minimum of 1 E cylinder of O2 is:   1000 psi (about 1/2 tank)  
🗑
What is the formula to calculate oxygen cylinder supply duration?   O2 volume (L) = capacity in L * gauge pressure (psi) divided by service pressure (psi). Divide this answer by O2 flow rate  
🗑
What is the formula to estimate oxygen cylinder supply duration?   Time remaining (hrs) = PSIG divided by (200 * flow rate)  
🗑
Nitrous oxide is manufactured by heating _________   ammonium nitrate  
🗑
Nitrous oxide has a service pressure of ____ until ___% of the tank is used   745 psi, 75%  
🗑
The only reliable way to determine the remaining volume of nitrous oxide is to:   weigh the tank  
🗑
The energy consumed in conversion of liquid to gas is called:   latent heat of vaporization  
🗑
Pipeline pressure is _____ psi & _______ (low, intermediate, or high) pressure   45 - 55 psi, plus or minus 5; intermediate pressure  
🗑
______ prevents pipeline supply from going to the O2 tank instead of to the patient (prevent retrograde flow)   one way check valve  
🗑
Pipeline pressure is monitored by ______   bourdon gauge  
🗑
With pipeline supply what is the safety mechanism for low pipeline oxygen pressure?   Fail-safe valve stops delivery of all other gases to prevent hypoxic gas delivery  
🗑
If there is a loss of pipeline oxygen pressure the 3 things that are to be done are:   open E cylinder fully, disconnect pipeline O2 supply, consider low fresh gas flows & manual ventilation (b/c bellows are driven by O2 pressure)  
🗑
What is the function of the scavenging system?   collects anesthetic waste gases from the breathing circuit & ventilator & remove it from the operating room  
🗑
Per OSHA standards the exposure to halogenated agents is less than ____; halogentated agents w/ nitrous is less than ___; nitrous oxide is less than _____   halogenated: 2 ppm halogenated agents w/ N2O: 0.5 ppm N2O: 25 ppm  
🗑
Room ventilation requires a minimum of ___ air exchanges per hour w/ at least ___ fresh/outdoor air exchanges per hour   15 air exchanges; 3 fresh/outdoor air exchanges  
🗑
5 ways to limit waste gas exposure are:   use low fresh gas flows, check machine for leaks, disconnect N2O at wall outlet at end of day, use TIVA, or avoid use of nitrous oxide  
🗑
The 2 types of disposal interfaces are _____, which includes ___ & _____; and _____, which includes ____   Closed interface includes active & passive scavenge. Open interface includes active scavenge  
🗑
Closed interface communicates with ________ & _____ (does / does not) hiss   communicates w/ atmosphere only through valves; does not hiss  
🗑
Open interface communicates with ________ & _____ (does / does not) hiss   communicates directly with atmosphere; does hiss  
🗑
In the closed interface system, what is the difference between passive & active scavenge?   Passive: needs positive pressure relief valve Active: vacuum. Has positive & negative pressure relief valves  
🗑
Improper use of the open interface =   OR contamination  
🗑
_____ & _____ systems rely on proper vacuum control   Active closed interface & open interface scavenge systems  
🗑
Hissing with closed interface means:   positive pressure relief valve is releasing waste gases into the operating room  
🗑
Hissing with open interface means:   proper operation & waste gas scavenge  
🗑


   

Review the information in the table. When you are ready to quiz yourself you can hide individual columns or the entire table. Then you can click on the empty cells to reveal the answer. Try to recall what will be displayed before clicking the empty cell.
 
To hide a column, click on the column name.
 
To hide the entire table, click on the "Hide All" button.
 
You may also shuffle the rows of the table by clicking on the "Shuffle" button.
 
Or sort by any of the columns using the down arrow next to any column heading.
If you know all the data on any row, you can temporarily remove it by tapping the trash can to the right of the row.

 
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how
Created by: Thommy413
Popular Science sets