Busy. Please wait.

Forgot Password?

Don't have an account?  Sign up 

show password


Make sure to remember your password. If you forget it there is no way for StudyStack to send you a reset link. You would need to create a new account.

By signing up, I agree to StudyStack's Terms of Service and Privacy Policy.

Already a StudyStack user? Log In

Reset Password
Enter the email address associated with your account, and we'll email you a link to reset your password.

Remove ads
Don't know (0)
Know (0)
remaining cards (0)
To flip the current card, click it or press the Spacebar key.  To move the current card to one of the three colored boxes, click on the box.  You may also press the UP ARROW key to move the card to the "Know" box, the DOWN ARROW key to move the card to the "Don't know" box, or the RIGHT ARROW key to move the card to the Remaining box.  You may also click on the card displayed in any of the three boxes to bring that card back to the center.

Pass complete!

"Know" box contains:
Time elapsed:
restart all cards

Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how

Anesth Machines 1

Clinical - Supply & disposal

The current anesthesia gas machine standard is: ASTM F1850 (American Standard for Testing & Materials)
SPDD stands for: Supply, processing, delivery, disposal
Supply: fresh gases get to the anesthesia machine by _____ and _____ pipeline, cylinders
Supply: power gets to the anesthesia machine by _____ and _____ electricity, battery backup
Processing: 6 ways the machine prepares gases before delivery to the patient are: valves, flowmeters, oxygen flush, low pressure alarms, vaporizers, common gas outlet
Delivery: 3 ways that delivered gases are controlled & monitored: breathing circuits, ventilators, integral monitors
How are waste gases disposed? scavenge systems
Cylinder pressure = ____ pressure system Pipeline pressure = ____ pressure system cylinder = high pressure system pipeline = intermediate pressure system
Low pressure system includes: flowmeter tubes, vaporizers, check valves, common gas outlet
Cylinder pressure is measured by: Bourdon gauges
What type of regulator is on a cylinder? What type of regulator is on a pipeline? Cylinder = 1st stage regulator Pipeline = oxygen 2nd stage regulator
1st stage regulator drops cylinder pressure to: 40 - 48 psi (intermediate pressure)
The volume of oxygen & air in E cylinder is: 625 - 700 L
The volume of nitrous oxide in E cylinder is: 1590 L
Service pressure (PSI) of oxygen & air is: 1900
Service pressure (PSI) of nitrous oxide is: 745
What safety system is on cylinders? How is the safety disabled? PISS (pin index safety system). Disabled if pins are missing or if more than 1 washer is used
Valve stem on cylinder is fragile & contains one or more of the 3 safety features: frangible disc (bursts w/ high pressure to release cylinder contents); safety relief (opens w/ high pressure to release cylinder content); fusible plug (alloy melts at 200 degrees F to release cylinder contents)
Service pressure is: the pressure in the cylinder
For cylinder supply the hanger yoke does 3 things, which are: orients the cylinder, gas tight seal, ensures unidirectional flow
Which gas law pertains to volume & pressure? Boyle's Law
Minimum of 1 E cylinder of O2 is: 1000 psi (about 1/2 tank)
What is the formula to calculate oxygen cylinder supply duration? O2 volume (L) = capacity in L * gauge pressure (psi) divided by service pressure (psi). Divide this answer by O2 flow rate
What is the formula to estimate oxygen cylinder supply duration? Time remaining (hrs) = PSIG divided by (200 * flow rate)
Nitrous oxide is manufactured by heating _________ ammonium nitrate
Nitrous oxide has a service pressure of ____ until ___% of the tank is used 745 psi, 75%
The only reliable way to determine the remaining volume of nitrous oxide is to: weigh the tank
The energy consumed in conversion of liquid to gas is called: latent heat of vaporization
Pipeline pressure is _____ psi & _______ (low, intermediate, or high) pressure 45 - 55 psi, plus or minus 5; intermediate pressure
______ prevents pipeline supply from going to the O2 tank instead of to the patient (prevent retrograde flow) one way check valve
Pipeline pressure is monitored by ______ bourdon gauge
With pipeline supply what is the safety mechanism for low pipeline oxygen pressure? Fail-safe valve stops delivery of all other gases to prevent hypoxic gas delivery
If there is a loss of pipeline oxygen pressure the 3 things that are to be done are: open E cylinder fully, disconnect pipeline O2 supply, consider low fresh gas flows & manual ventilation (b/c bellows are driven by O2 pressure)
What is the function of the scavenging system? collects anesthetic waste gases from the breathing circuit & ventilator & remove it from the operating room
Per OSHA standards the exposure to halogenated agents is less than ____; halogentated agents w/ nitrous is less than ___; nitrous oxide is less than _____ halogenated: 2 ppm halogenated agents w/ N2O: 0.5 ppm N2O: 25 ppm
Room ventilation requires a minimum of ___ air exchanges per hour w/ at least ___ fresh/outdoor air exchanges per hour 15 air exchanges; 3 fresh/outdoor air exchanges
5 ways to limit waste gas exposure are: use low fresh gas flows, check machine for leaks, disconnect N2O at wall outlet at end of day, use TIVA, or avoid use of nitrous oxide
The 2 types of disposal interfaces are _____, which includes ___ & _____; and _____, which includes ____ Closed interface includes active & passive scavenge. Open interface includes active scavenge
Closed interface communicates with ________ & _____ (does / does not) hiss communicates w/ atmosphere only through valves; does not hiss
Open interface communicates with ________ & _____ (does / does not) hiss communicates directly with atmosphere; does hiss
In the closed interface system, what is the difference between passive & active scavenge? Passive: needs positive pressure relief valve Active: vacuum. Has positive & negative pressure relief valves
Improper use of the open interface = OR contamination
_____ & _____ systems rely on proper vacuum control Active closed interface & open interface scavenge systems
Hissing with closed interface means: positive pressure relief valve is releasing waste gases into the operating room
Hissing with open interface means: proper operation & waste gas scavenge