Formulas for PRI

Quiz yourself by thinking what should be in each of the black spaces below before clicking on it to display the answer.
        Help!  

Question
Answer
Calculating mAs   mAs = mA x seconds  
🗑
Calculating Heat Units   HU = mA x time x kVp x generator factor  
🗑
Adjusting Milliamperage or Exposure Time   100 mA @ 0.10s = 10mAs. To increase the mAs to 20, you could use: 200 mA @ 0.10 = 20mA; 100mA @ 0.20s = 20mAs  
🗑
Adjusting Milliamperage and Exposure Time to Maintain mAs   100mA @ 100 ms(0.10s) = 10mAs. To maintain the mAs, you could use: 200mA @ 50 ms(0.5s) = 10mAs; 50mA @ 200 ms(0.20s) = 10mAs  
🗑
Using the 15% Rule: To increase density   Multiply the kVp by 1.15 (original kVp + 15%0  
🗑
Using the 15% Rule: To decrease density   Multiply the kVp by 0.85 (original kVp - 15%)  
🗑
Using the 15% Rule: To maintain density   When increasing kVp by 15% (kVp x 1.15), divide the original mAs by 2. When decreasing the kVp by 15% (kVp x 0.85), multiply the mAs by 2  
🗑
Inverse Square Law Formula   I1/I2 = (D1)^2/(D1)^2  
🗑
Density Maintenance Formula   mAs1/mAs2 = (SID1)^2/(SID2)^2  
🗑
Adjusting mAs for Changes in Grid   mAs1/mAs2 = Grid conversion factor1/Grid conversion factor2  
🗑
Adjusting mAs for Changes in Film-Screen System Speed   mAs1/mAs2 = RS1/RS2  
🗑
Adjustiong mAs for Changes in Part Thickness   An optimal radiograph was obtained using 40 mAs on an naatomic part that measured 18cm.The same anatomic part is radiographed in another patient and measures 22cm.What ne mAs is needed to maintain density?Because  
🗑
  the part thickness was increased by 4cm the original mAs is multiplied by 2 yielding 80 mAs.If another patient for the same part measures 26 cm what new mAs is needed to maintain density?Because the part thickness increased by another 4cm the mAs is  
🗑
  by another 4cm the mAs is multiplied by 2 yielding 160mAs.This mAs is 4 times greater that for the original patient who measured 8cm less.  
🗑
Calculating Geometric Unsharpness   Geometric unsharpness = Focal spot size x OID/SOD where Focal spot size = Dimensions of the nominal focal spot in millimeters(mm)  
🗑
  OID=the distance between the object(area of interest) and the image receptor;SOD=the distance between the focal spot(source) and object (area of interest); SOD=SID - OID. When calculating the amount of geometric unsharpness the radiographer must keep the  
🗑
  radiographer must keep the distance measurement in the same units(i.e. inches or centimeters)  
🗑
The Magnification Factor   SOD=SID - OID; MF=SID/SOD  
🗑
Determining Object Siz   Object size = Image size/MF  
🗑


   

Review the information in the table. When you are ready to quiz yourself you can hide individual columns or the entire table. Then you can click on the empty cells to reveal the answer. Try to recall what will be displayed before clicking the empty cell.
 
To hide a column, click on the column name.
 
To hide the entire table, click on the "Hide All" button.
 
You may also shuffle the rows of the table by clicking on the "Shuffle" button.
 
Or sort by any of the columns using the down arrow next to any column heading.
If you know all the data on any row, you can temporarily remove it by tapping the trash can to the right of the row.

 
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how
Created by: dawn2000
Popular Miscellaneous sets